Diffusion-weighted echo planar MR imaging of the neck at 3 T using integrated shimming: comparison of MR sequence techniques for reducing artifacts caused by magnetic-field inhomogeneities

被引:21
|
作者
Gatidis, Sergios [1 ,4 ]
Graf, Hansjoerg [2 ]
Weiss, Jakob [1 ]
Stemmer, Alto [2 ]
Kiefer, Berthold [2 ]
Nikolaou, Konstantin [1 ]
Notohamiprodjo, Mike [1 ]
Martirosian, Petros [3 ]
机构
[1] Univ Tubingen, Dept Radiol Diagnost & Intervent Radiol, Tubingen, Germany
[2] Siemens Healthcare GmbH, Erlangen, Germany
[3] Univ Tubingen, Dept Radiol, Expt Radiol, Tubingen, Germany
[4] Univ Tubingen Hosp, Dept Radiol, Hoppe Seyler Str 3, D-72076 Tubingen, Germany
关键词
MRI; Diffusion weighted; Echo-planar Imaging; Artifacts; Neck; BRAIN; LESIONS; HEAD; EPI;
D O I
10.1007/s10334-016-0582-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Our objective was to compare available techniques reducing artifacts in echo planar imaging (EPI)-based diffusion-weighed magnetic resonance imaging MRI (DWI) of the neck at 3 Tesla caused by B0-field inhomogeneities. A cylindrical fat-water phantom was equipped with a Maxwell coil allowing for additional linear B0-field variations in z-direction. The effect of increasing strength of this superimposed gradient on image quality was observed using a standard single-shot EPI-based DWI sequence (sEPI), a zoomed single-shot EPI sequence (zEPI), a readout-segmented EPI sequence (rsEPI), and an sEPI sequence with integrated dynamic shimming (intEPI) on a 3-Tesla system. Additionally, ten volunteers were examined over the neck region using these techniques. Image quality was assessed by two radiologists. Scan durations were recorded. With increasing strength of the external gradient, marked distortions, signal loss, and failure of fat suppression were observed using sEPI, zEPI, and rsEPI. These artifacts were markedly reduced using intEPI. Significantly better in vivo image quality was also observed using intEPI compared with the other techniques. Scan time of intEPI was similar to sEPI and zEPI and shorter than rsEPI. The use of integrated 2D shim and frequency adjustment for EPI-based DWI results in a significant improvement in image quality of the head/neck region at 3 Tesla. Combining integrated shimming with rsEPI or zEPI can be expected to provide additional improvements.
引用
收藏
页码:57 / 63
页数:7
相关论文
共 50 条
  • [21] Diffusion-weighted magnetic resonance imaging in rat kidney using two-dimensional navigated, interleaved echo-planar imaging at 7.0 T
    Liu, Qiang
    Xu, Zhongbiao
    Zhao, Kaixuan
    Hoge, W. Scott
    Zhang, Xinyuan
    Mei, Yingjie
    Lu, Qiqi
    Niendorf, Thoralf
    Feng, Yanqiu
    NMR IN BIOMEDICINE, 2022, 35 (05)
  • [22] Image quality evaluation of diffusion-weighted imaging in bladder cancer: a comparison between integrated slice-specific dynamic shimming and single-shot echo-planar imaging
    Wang, Yiqian
    Chen, Shaobin
    Zhang, Wei
    Xiao, Weixiong
    Su, Jiawei
    Zhang, Ruochen
    Wei, Yongbao
    Luo, Min
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2023, 13 (04) : 2526 - +
  • [23] Reduced Field-of-View Diffusion-Weighted Magnetic Resonance Imaging of the Prostate at 3 Tesla: Comparison With Standard Echo-Planar Imaging Technique for Image Quality and Tumor Assessment
    Tamada, Tsutomu
    Ream, Justin M.
    Doshi, Ankur M.
    Taneja, Samir S.
    Rosenkrantz, Andrew B.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2017, 41 (06) : 949 - 956
  • [24] High-Resolution Diffusion-Weighted Imaging of Neck Lymph Nodes Using 2D-Single-Shot Interleaved Multiple Inner Volume Imaging Diffusion-Weighted Echo-Planar Imaging at 3T
    Park, J. K.
    Kim, S. -E.
    Trieman, G. S.
    Parker, D.
    Jeong, E. -K.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2011, 32 (07) : 1173 - 1177
  • [25] Single-Shot Echo-Planar Diffusion-Weighted MR Imaging at 3T and 1.5T for Differentiation of Benign Vertebral Fracture Edema and Tumor Infiltration
    Park, Hee Jin
    Lee, So Yeon
    Rho, Myung Ho
    Chung, Eun Chul
    Kim, Mi Sung
    Kwon, Heon Ju
    Youn, In Young
    KOREAN JOURNAL OF RADIOLOGY, 2016, 17 (05) : 590 - 597
  • [26] Improve the image quality of orbital 3 T diffusion-weighted magnetic resonance imaging with readout-segmented echo-planar imaging
    Xu, Xiao-Quan
    Liu, Jun
    Hu, Hao
    Su, Guo-Yi
    Zhang, Yu-Dong
    Shi, Hai-Bin
    Wu, Fei-Yun
    CLINICAL IMAGING, 2016, 40 (04) : 793 - 796
  • [27] Diffusion-weighted echo-planar imaging of the head and neck using 3-T MRI: Investigation into the usefulness of liquid perfluorocarbon pads and choice of optimal fat suppression method
    Maehara, Minoru
    Ikeda, Koshi
    Kurokawa, Hiroaki
    Ohmura, Naoto
    Ikeda, Shigeki
    Hirokawa, Yuzo
    Maehara, Saori
    Utsunomiya, Keita
    Tanigawa, Noboru
    Sawada, Satoshi
    MAGNETIC RESONANCE IMAGING, 2014, 32 (05) : 440 - 445
  • [28] Endometrial cancer: preoperative staging using three-dimensional T2-weighted turbo spin-echo and diffusion-weighted MR imaging at 3.0 T: a prospective comparative study
    Hori, Masatoshi
    Kim, Tonsok
    Onishi, Hiromitsu
    Imaoka, Izumi
    Kagawa, Yuki
    Murakami, Takamichi
    Nakamoto, Atsushi
    Ueguchi, Takashi
    Tatsumi, Mitsuaki
    Enomoto, Takayuki
    Kimura, Tadashi
    Tomiyama, Noriyuki
    EUROPEAN RADIOLOGY, 2013, 23 (08) : 2296 - 2305
  • [29] Zoomed echo-planar imaging using parallel transmission: impact on image quality of diffusion-weighted imaging of the prostate at 3T
    Rosenkrantz, Andrew B.
    Chandarana, Hersh
    Pfeuffer, Josef
    Triolo, Michael J.
    Shaikh, Mohammed Bilal
    Mossa, David J.
    Geppert, Christian
    ABDOMINAL IMAGING, 2015, 40 (01): : 120 - 126
  • [30] Improved Readout-Segmented Echo-Planner Diffusion-Weighted Magnetic Resonance Imaging of Nasopharyngeal Carcinoma Using Simultaneous Multislice Acquisitions at 3 T
    Li, Qiao
    Jiang, TingTing
    Wang, TingTing
    Huang, Yan
    Hu, XiaoXin
    Zhang, Ling
    Liu, Wei
    Fu, CaiXia
    Gu, YaJia
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2022, 46 (05) : 815 - 822