Estimation of boron isotope ratios using high resolution continuum source atomic absorption spectrometry

被引:17
作者
Wiltsche, Helmar [1 ]
Prattes, Karl [2 ]
Zischka, Michael [1 ]
Knapp, Guenter [1 ]
机构
[1] Graz Univ Technol, Inst Analyt Chem & Radiochem, A-8010 Graz, Austria
[2] Bohler Edelstahl GmbH, Kapfenberg, Austria
关键词
Boron; Isotope ratio; Steel; Continuum source AAS; INDUCTIVELY-COUPLED PLASMA; MASS-SPECTROMETRY; SPECTRA; SAMPLES; SHIFT;
D O I
10.1016/j.sab.2009.03.008
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In the production of (10)B enriched steels. the production-recycling process needs to be closely monitored for inadvertent Mix-up of materials with different B isotope levels. A quick and simple method for the estimation of boron isotope ratios in high alloyed steels using high resolution continuum source flame AAS (HR-CS-FAAS) was developed. On the 208.9 nm B line the wavelength of the peak absorption of (10)B and (11)B differs by 2.5 pm. The wavelength of the peak absorption of boron was determined by fitting a Gauss function through spectra simultaneously recorded by HR-CS-FAAS. It was shown that a linear correlation between the wavelength of the peak absorption and the isotope ratio exists and that this correlation is independent of the total boron concentration. Internal spectroscopic standards were used to compensate for monochromator drift and monochromator resolution changes. Accuracy and precision of the analyzed samples were thereby increased by a factor of up to 1.3. Three steel reference materials and one boric acid CRM, each certified for the boron isotope ratio were used to validate the procedure. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:341 / 346
页数:6
相关论文
共 20 条