Validation of OSLD and a treatment planning system for surface dose determination in IMRT treatments

被引:38
作者
Zhuang, Audrey H. [1 ]
Olch, Arthur J. [1 ,2 ]
机构
[1] Univ So Calif, Keck Sch Med, Dept Radiat Oncol, Los Angeles, CA 90033 USA
[2] Childrens Hosp Los Angeles, Dept Radiat Oncol, Los Angeles, CA 90027 USA
关键词
optically stimulated luminescent dosimeters; OSLD; in vivo dosimetry; eclipse; surface dose; STIMULATED LUMINESCENT DOSIMETERS; MEGAVOLTAGE PHOTON; ANGULAR-DEPENDENCE; BUILDUP REGION; BASAL-CELL; IONIZING-RADIATION; SKIN-CANCER; IMPACT; MOSFET; HEAD;
D O I
10.1118/1.4890795
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the accuracy of skin dose determination for composite multibeam 3D conformal radiation therapy (3DCRT) and intensity modulated radiation therapy (IMRT) treatments using optically stimulated luminescent dosimeters (OSLDs) and Eclipse treatment planning system. Methods: Surface doses measured by OSLDs in the buildup region for open field 6 MV beams, either perpendicular or oblique to the surface, were evaluated by comparing against dose measured by Markus Parallel Plate (PP) chamber, surface diodes, and calculated by Monte Carlo simulations. The accuracy of percent depth dose (PDD) calculation in the buildup region from the authors' Eclipse system (Version 10), which was precisely commissioned in the buildup region and was used with 1 mm calculation grid, was also evaluated by comparing to PP chamber measurements and Monte Carlo simulations. Finally, an anthropomorphic pelvic phantom was CT scanned with OSLDs in place at three locations. A planning target volume (PTV) was defined that extended close to the surface. Both an 8 beam 3DCRT and IMRT plan were generated in Eclipse. OSLDs were placed at the CT scanned reference locations to measure the skin doses and were compared to diode measurements and Eclipse calculations. Efforts were made to ensure that the dose comparison was done at the effective measurement points of each detector and corresponding locations in CT images. Results: The depth of the effective measurement point is 0 8 mm for OSLD when used in the buildup region in a 6 MV beam and is 0.7 mm for the authors' surface diode. OSLDs and Eclipse system both agree well with Monte Carlo and/or Markus PP ion chamber and/or diode in buildup regions in 6 MV beams with normal or oblique incidence and across different field sizes. For the multiple beam 3DCRT plan and IMRT plans, the differences between OSLDs and Eclipse calculations on the surface of the anthropomorphic phantom were within 3% and distance-to-agreement less than 0.3 mm. Conclusions: The authors' experiment showed that OSLD is an accurate dosimeter for skin dose measurements in complex 3DCRT or IMRT plans. It also showed that an Eclipse system with accurate commissioning of the data in the buildup region and 1 mm calculation grid can calculate surface doses with high accuracy and has a potential to replace in vivo measurements. 2014 American Association of Physicists in Medicine.
引用
收藏
页码:231 / 238
页数:8
相关论文
共 29 条
[1]   Validation of Monte Carlo calculated surface doses for megavoltage photon beams [J].
Abdel-Rahman, W ;
Seuntjens, JP ;
Verhaegen, F ;
Deblois, F ;
Podgorsak, EB .
MEDICAL PHYSICS, 2005, 32 (01) :286-298
[2]   A new radiotherapy surface dose detector: The MOSFET [J].
Butson, MJ ;
Rozenfeld, A ;
Mathur, JN ;
Carolan, M ;
Wong, TPY ;
Metcalfe, PE .
MEDICAL PHYSICS, 1996, 23 (05) :655-658
[3]   MOSFET detectors in quality assurance of tomotherapy treatments [J].
Cherpak, Amanda ;
Studinski, Ryan C. N. ;
Cygler, Joanna E. .
RADIOTHERAPY AND ONCOLOGY, 2008, 86 (02) :242-250
[4]   Evaluation of surface and build-up region dose for intensity-modulated radiation therapy in head and neck cancer [J].
Chung, HT ;
Jin, HS ;
Dempsey, JF ;
Liu, CH ;
Palta, J ;
Suh, TS ;
Kim, SY .
MEDICAL PHYSICS, 2005, 32 (08) :2682-2689
[5]   Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system [J].
Court, Laurence E. ;
Tishler, Roy B. ;
Allen, Aaron M. ;
Xiang, Hong ;
Makrigiorgos, Mike ;
Chin, Lee .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2008, 9 (01) :29-35
[6]   Accurate skin dose measurements using radiochromic film in clinical applications [J].
Devic, S ;
Seuntjens, J ;
Abdel-Rahman, W ;
Evans, M ;
Olivares, M ;
Podgorsak, EB ;
Vuong, T ;
Soares, CG .
MEDICAL PHYSICS, 2006, 33 (04) :1116-1124
[7]   MEASUREMENT OF DOSE IN THE BUILDUP REGION USING FIXED-SEPARATION PLANE-PARALLEL IONIZATION CHAMBERS [J].
GERBI, BJ ;
KHAN, FM .
MEDICAL PHYSICS, 1990, 17 (01) :17-26
[8]   Evaluation of surface and superficial dose for head and neck treatments using conventional or intensity-modulated techniques [J].
Higgins, P. D. ;
Han, E. Y. ;
Yuan, J. L. ;
Hui, S. ;
Lee, C. K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (04) :1135-1146
[9]   Dose discrepancies in the buildup region and their impact on dose calculations for IMRT fields [J].
Hsu, Shu-Hui ;
Moran, Jean M. ;
Chen, Yu ;
Kulasekere, Ravi ;
Roberson, Peter L. .
MEDICAL PHYSICS, 2010, 37 (05) :2043-2053
[10]  
International Commission on Radiation Units and Measurements, 1985, 39 ICRU