An Accurate Multimodal 3-D Vessel Segmentation Method Based on Brightness Variations on OCT Layers and Curvelet Domain Fundus Image Analysis

被引:17
作者
Kafieh, Raheleh [1 ]
Rabbani, Hossein [1 ,2 ]
Hajizadeh, Fedra [3 ]
Ommani, Mohammadreza [1 ]
机构
[1] Isfahan Univ Med Sci, Dept Biomed Engn, Med Image & Signal Proc Res Ctr, Esfahan, Iran
[2] Univ Iowa, Iowa Inst Biomed Imaging, Iowa City, IA 52242 USA
[3] Noor Eye Hosp, Noor Ophthalmol Res Ctr, Tehran, Iran
关键词
Curvelet transform; fundus image; multimodal; optical coherence tomography (OCT); 3-D reconstruction; vessel segmentation; OPTICAL COHERENCE TOMOGRAPHY; RETINAL BLOOD-VESSELS; AUTOMATED SEGMENTATION; EXTRACTION; VASCULATURE; RETINOPATHY; ANGIOGRAMS; ALGORITHM; NERVE;
D O I
10.1109/TBME.2013.2263844
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper proposes a multimodal approach for vessel segmentation of macular optical coherence tomography (OCT) slices along with the fundus image. The method is comprised of two separate stages; the first step is 2-D segmentation of blood vessels in curvelet domain, enhanced by taking advantage of vessel information in crossing OCT slices (named feedback procedure), and improved by suppressing the false positives around the optic nerve head. The proposed method for vessel localization of OCT slices is also enhanced utilizing the fact that retinal nerve fiber layer becomes thicker in the presence of the blood vessels. The second stage of this method is axial localization of the vessels in OCT slices and 3-D reconstruction of the blood vessels. Twenty-four macular spectral 3-D OCT scans of 16 normal subjects were acquired using a Heidelberg HRA OCT scanner. Each dataset consisted of a scanning laser ophthalmoscopy (SLO) image and limited number of OCT scans with size of 496 x 512 (namely, for a data with 19 selected OCT slices, the whole data size was 496 x 512 x 19). The method is developed with least complicated algorithms and the results show considerable improvement in accuracy of vessel segmentation over similar methods to produce a local accuracy of 0.9632 in area of SLO, covered with OCT slices, and the overall accuracy of 0.9467 in the whole SLO image. The results are also demonstrative of a direct relation between the overall accuracy and percentage of SLO coverage by OCT slices.
引用
收藏
页码:2815 / 2823
页数:9
相关论文
共 45 条
[1]   Fast retinal vessel tree extraction:: A pixel parallel approach [J].
Alonso-Montes, C. ;
Vilarino, D. L. ;
Dudek, P. ;
Penedo, M. G. .
INTERNATIONAL JOURNAL OF CIRCUIT THEORY AND APPLICATIONS, 2008, 36 (5-6) :641-651
[2]   High speed detection of retinal blood vessels in fundus image using phase congruency [J].
Amin, M. Ashraful ;
Yan, Hong .
SOFT COMPUTING, 2011, 15 (06) :1217-1230
[3]  
[Anonymous], I MATLAB VERS 7 8
[4]   Different averages of a fuzzy set with an application to vessel segmentation [J].
Ayala, G ;
León, T ;
Zapater, V .
IEEE TRANSACTIONS ON FUZZY SYSTEMS, 2005, 13 (03) :384-393
[5]   DETECTION OF BLOOD-VESSELS IN RETINAL IMAGES USING TWO-DIMENSIONAL MATCHED-FILTERS [J].
CHAUDHURI, S ;
CHATTERJEE, S ;
KATZ, N ;
NELSON, M ;
GOLDBAUM, M .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1989, 8 (03) :263-269
[6]   Detection of blood vessels in ophthalmoscope images using MF/ant (matched filter/ant colony) algorithm [J].
Cinsdikici, Muhammed Goekhan ;
Aydin, Dogan .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2009, 96 (02) :85-95
[7]   Automatic model-based tracing algorithm for vessel segmentation and diameter estimation [J].
Delibasis, Konstantinos K. ;
Kechriniotis, Aristides I. ;
Tsonos, C. ;
Assimakis, Nicholas .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2010, 100 (02) :108-122
[8]   EXTRACTION OF RETINAL BLOOD VESSELS BY CURVELET TRANSFORM [J].
Esmaeili, Mahdad ;
Rabbani, Hossein ;
Mehri, Alireza ;
Dehghani, Alireza .
2009 16TH IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING, VOLS 1-6, 2009, :3353-+
[9]  
Eva D., 2009, C MED IM UND AN LOND
[10]  
Foracchia M., 2001, P 2 INT WORKSH COMP