Assessing the information content of phosphor produced medical images:: application to Zn2SiO4:Mn phosphor

被引:12
作者
Cavouras, D
Kandarakis, I
Nomicos, CD
Panayiotakis, GS
Fezoulidis, I
机构
[1] Inst Educ Technol, Dept Med Instrumentat Technol, Athens 12210, Greece
[2] Inst Educ Technol, Dept Elect, Athens 12210, Greece
[3] Univ Patras, Sch Med, Dept Med Phys, Patras 26500, Greece
[4] Univ Thessalia, Sch Med, Dept Radiol, Larisa 41222, Greece
关键词
medical phosphors; phosphor screens; MTF; signal-to-noise ratio;
D O I
10.1016/S0969-8043(99)00116-5
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In this study a method to assess the information content of medical images produced by phosphors is described. The optical signal emitted by the phosphor after X-ray excitation, the detective quantum efficiency (DQE), expressing the signal-re-noise ratio (SNR) transfer efficiency, and the information;capacity were experimentally determined. The method was based on light flux and modulation transfer function (MTF) measurements and was used to assess the imaging performance of the Zn2SiO4:Mn phosphor. The latter was employed in the form of laboratory prepared phosphor layers (test screens). Results showed that high values for optical signal emission and DQE were obtained for medium thickness phosphor layers (56 and 89 mg/cm(2)) at 20 kVp X-ray tube voltage. The information capacity was found to decrease continuously with phosphor coating weight. (C) 2000 Elsevier Science Ltd, All rights reserved.
引用
收藏
页码:119 / 126
页数:8
相关论文
共 20 条
[1]  
[Anonymous], 1986, 41 ICRU
[2]  
Barnes G. T., 1979, PHYSICS MED IMAGING, P138
[3]   ANALYSIS OF THE DETECTIVE QUANTUM EFFICIENCY OF A RADIOGRAPHIC SCREEN FILM COMBINATION [J].
BUNCH, PC ;
HUFF, KE ;
VANMETTER, R .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05) :902-909
[4]   An evaluation of the Y2O3:Eu3+ scintillator for application in medical x-ray detectors and image receptors [J].
Cavouras, D ;
Kandarakis, I ;
Panayiotakis, GS ;
Evangelou, EK ;
Nomicos, CD .
MEDICAL PHYSICS, 1996, 23 (12) :1965-1975
[5]   Signal-to-noise-ratio (SNR) of X-ray imaging scintillators determined by luminescence measurements [J].
Cavouras, D ;
Kandarakis, I ;
Kanellopoulos, M ;
Nomicos, CD ;
Panayiotakis, GS .
APPLIED RADIATION AND ISOTOPES, 1999, 51 (01) :59-68
[6]   An investigation of the imaging characteristics of the Y2O2S:Eu3+ phosphor for application in X-ray detectors of digital mammography [J].
Cavouras, D ;
Kandarakis, I ;
Panayiotakis, GS ;
Kanellopoulos, E ;
Triantis, D ;
Nomicos, CD .
APPLIED RADIATION AND ISOTOPES, 1998, 49 (08) :931-937
[7]  
DAINTY JC, 1974, IMAGE SCI, P152
[8]  
GREENING JR, 1985, MED PHYSICS HDB
[9]  
HENDEE WR, 1970, MED RAD PHYSICS, P145
[10]   IMAGING PERFORMANCE OF A DIGITAL STORAGE PHOSPHOR SYSTEM [J].
HILLEN, W ;
SCHIEBEL, U ;
ZAENGEL, T .
MEDICAL PHYSICS, 1987, 14 (05) :744-751