Physiological stress monitoring using sodium ion potentiometric microsensors for sweat analysis

被引:60
作者
Cazale, A. [1 ,2 ,3 ]
Sant, W. [1 ]
Ginot, F. [1 ]
Launay, J. -C. [4 ]
Savourey, G. [4 ]
Revol-Cavalier, F. [5 ]
Lagarde, J. M. [6 ]
Heinry, D. [7 ]
Launay, J. [2 ,3 ]
Temple-Boyer, P. [2 ,3 ]
机构
[1] HEMODIA SA, F-31682 Labege, France
[2] CNRS, LAAS, F-31400 Toulouse, France
[3] Univ Toulouse, UPS, LAAS, F-31400 Toulouse, France
[4] IRBA CRSSA, F-38702 La Tronche, France
[5] CEA, LETI, Technol Healthcare & Biol Div, F-38054 Grenoble, France
[6] Ctr Rech Peau, Lab Pierre Fabre Dermocosmet, F-31025 Toulouse, France
[7] PAUL BOYE TECHNOL, F-31860 Labarthe Sur Leze, France
来源
SENSORS AND ACTUATORS B-CHEMICAL | 2016年 / 225卷
关键词
Potentiometric sensor; ISE; ISFET; Na+ ion detection; Sweat analysis; Physiological stress; FIELD-EFFECT TRANSISTORS; SENSORS; NA+; EXERCISE; LACTATE; UREA;
D O I
10.1016/j.snb.2015.10.114
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In the frame of sweat analysis, two technologies, based on either ISE or ISFET devices, were developed for the implementation of pNa potentiometric microsensors. Both of them demonstrated good sodium ion Na+ detection properties with a global sensitivity of around 110 mV/pNa in NaCl-based solutions due to the use of an integrated "Ag/AgCl ink" pseudo-reference electrode. Then, in order to deal with in vivo analysis of sweat natremia, a physiological sweatband prototype was developed, consisting of pNa-ISE and pNa-ISFET electronic detection modules as well as a textile-based sweat pump. Finally, sweating process was studied during series of experiments on twenty-five healthy consenting subjects. The sodium ion concentration [Na+] was successfully monitored in sweat during various heat exposures, demonstrating a global increase with exercise trial duration. Furthermore, a strong correlation was found between the sweat [Na+] concentration and the subject's internal temperature 0, allowing monitoring the subject's heat stress state. All in all, the relevance of the Na+ ion analysis was demonstrated for the physiological stress monitoring and pNa potentiometric microsensors were shown to be very promising for the development of smart sweatbands. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 51 条
  • [1] Photocurable Polymers for Ion Selective Field Effect Transistors. 20 Years of Applications
    Abramova, Natalia
    Bratov, Andrei
    [J]. SENSORS, 2009, 9 (09) : 7097 - 7110
  • [2] [Anonymous], 1947, PHYSL MAN DESERT
  • [3] Sodium microsensors based on ISFET/REFET prepared through an ion-implantation process fully compatible with a standard silicon technology
    Baccar, ZM
    JaffrezicRenault, N
    Martelet, C
    Jaffrezic, H
    Marest, G
    Plantier, A
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1996, 32 (02) : 101 - 105
  • [4] Non-invasive wearable electrochemical sensors: a review
    Bandodkar, Amay J.
    Wang, Joseph
    [J]. TRENDS IN BIOTECHNOLOGY, 2014, 32 (07) : 363 - 371
  • [5] Epidermal tattoo potentiometric sodium sensors with wireless signal transduction for continuous non-invasive sweat monitoring
    Bandodkar, Amay J.
    Molinnus, Denise
    Mirza, Omar
    Guinovart, Tomas
    Windmiller, Joshua R.
    Valdes-Ramirez, Gabriela
    Andrade, Francisco J.
    Schoening, Michael J.
    Wang, Joseph
    [J]. BIOSENSORS & BIOELECTRONICS, 2014, 54 : 603 - 609
  • [6] Tattoo-based potentiometric ion-selective sensors for epidermal pH monitoring
    Bandodkar, Amay J.
    Hung, Vinci W. S.
    Jia, Wenzhao
    Valdes-Ramirez, Gabriela
    Windmiller, Joshua R.
    Martinez, Alexandra G.
    Ramirez, Julian
    Chan, Garrett
    Kerman, Kagan
    Wang, Joseph
    [J]. ANALYST, 2013, 138 (01) : 123 - 128
  • [7] Brisson G.R., J APPL PHYSL, V66, P552
  • [8] A SIMPLE AND DISPOSABLE SWEAT COLLECTOR
    BRISSON, GR
    BOISVERT, P
    PERONNET, F
    PERRAULT, H
    BOISVERT, D
    LAFOND, JS
    [J]. EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY AND OCCUPATIONAL PHYSIOLOGY, 1991, 63 (3-4): : 269 - 272
  • [9] THE DESIGN OF DURABLE NA+-SELECTIVE CHEMFETS BASED ON POLYSILOXANE MEMBRANES
    BRUNINK, JAJ
    LUGTENBERG, RJW
    BRZOZKA, Z
    ENGBERSEN, JFJ
    REINHOUDT, DN
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1994, 378 (1-2): : 185 - 200
  • [10] CHEMICALLY MODIFIED FIELD-EFFECT TRANSISTORS - A SODIUM-ION SELECTIVE SENSOR BASED ON CALIX[4]ARENE RECEPTOR MOLECULES
    BRUNINK, JAJ
    HAAK, JR
    BOMER, JG
    REINHOUDT, DN
    MCKERVEY, MA
    HARRIS, SJ
    [J]. ANALYTICA CHIMICA ACTA, 1991, 254 (1-2) : 75 - 80