Performance of a helium circulation system for a MEG

被引:7
作者
Takeda, Tsunehiro [1 ]
Okamoto, Masayoshi [1 ]
Atsuda, Kazuhiro [1 ]
Katagiri, Keishi [1 ]
机构
[1] Univ Tokyo, Grad Sch Frontier Sci, Dept Complex Sci & Engn, Chiba 2778561, Japan
关键词
Liquid helium; SQUIDS; Gifford-McMahon; Cryostats; Reuse;
D O I
10.1016/j.cryogenics.2008.12.001
中图分类号
O414.1 [热力学];
学科分类号
摘要
We report the performance of a helium circulation system (HCS) for a magnetoencephalography (MEG) that re-liquefies all the evaporating helium gas using two 1.5 W GM cryocoolers operating at 4.2 K. The MEG with the HCS was used to measure human brain responses for over one and a half years without any noise problems. The noise level is below 10 fr/Hz(1/2) for 2-40 Hz, below 30 fr/Hz(1/2) at 1 Hz, and 200 IT/Hz(1/2) at 50 Hz, which is the power supply frequency. As the amount of liquid helium used decreases less than one percent, the maintenance cost of the MEG becomes less than one-tenth of the previous cost. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:144 / 150
页数:7
相关论文
共 5 条
[1]  
KANG Y, 1998, J JPN BIOMAGNET BIOE, V11, P15
[2]  
TAKEDA T, 2008, BIOMAGNETISM, P36
[3]  
TAKEDA T, J PHYS C SE IN PRESS
[4]  
TAKEDA T, 2004, BIOMAG, P640
[5]   An efficient helium circulation system with small GM cryocoolers [J].
Takeda, Tsunehiro ;
Okamoto, Masayoshi ;
Atsuda, Kazuhiro ;
Kobayashi, Aklhiro ;
Owaki, Takashi ;
Katagiri, Keishi .
CRYOGENICS, 2008, 48 (1-2) :6-11