Comparison of four near-infrared spectroscopy devices shows that they are only suitable for monitoring cerebral oxygenation trends in preterm infants

被引:68
作者
Schneider, Anna [1 ]
Minnich, Bernd [2 ]
Hofstaetter, Edda [1 ]
Weisser, Christof [1 ]
Hattinger-Juergenssen, Erna [1 ]
Wald, Martin [1 ]
机构
[1] Paracelsus Med Univ, Univ Hosp Pediat, Div Neonatol, Salzburg, Austria
[2] Salzburg Univ, Dept Cell Biol, Div Anim Struct & Funct, A-5020 Salzburg, Austria
关键词
Brain monitoring; Extraction coefficient; Near-infrared spectroscopy; Preterm infants; Regional oxygen saturation; TISSUE OXYGENATION; SATURATION; INDEX; EXTRACTION; PERFUSION; OXIMETER;
D O I
10.1111/apa.12698
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Aim: Measuring cerebral oxygenation using near-infrared spectroscopy (NIRS) has taken on an increasingly important role in the field of neonatology. Several companies have already developed commercial devices, and more publications are reporting absolute boundary values or percentiles for neonates. We compared four commercially used devices to discover whether they provided consistent results in the same patients. Methods: We recruited nine preterm infants and tested them for 2 h, using sensors from two different devices. The measurements were carried out six times on each child, so that all four devices were compared with each other. A total of 54 measurements were conducted. The following devices were compared: the NIRO 200 (Hamamatsu Photonics K.K), the INVOS 5100c (Somanetics), the Fore-Sight (CAS Med.) and the SenSmart X-100 (NONIN). Results: The cerebral tissue oxygenation data yielded by the individual devices differed significantly from each other, ranging from a minimum difference of 2.93% to a maximum difference of 12.66%. Conclusion: The commercially available NIRS devices showed highly significant differences in local cerebral tissue oxygenation levels, to the extent that the industry cannot agree on uniform and reproducible standards. Therefore, NIRS should only be used for trend measurements in preterm infants.
引用
收藏
页码:934 / 938
页数:5
相关论文
共 18 条
[1]   Factors Affecting the Performance of 5 Cerebral Oximeters During Hypoxia in Healthy Volunteers [J].
Bickler, Philip E. ;
Feiner, John R. ;
Rollins, Mark D. .
ANESTHESIA AND ANALGESIA, 2013, 117 (04) :813-823
[2]   Impact of Extracranial Contamination on Regional Cerebral Oxygen Saturation A Comparison of Three Cerebral Oximetty Technologies [J].
Davie, Sophie N. ;
Grocott, Hilary P. .
ANESTHESIOLOGY, 2012, 116 (04) :834-840
[3]   Comparing near-infrared spectroscopy devices and their sensors for monitoring regional cerebral oxygen saturation in the neonate [J].
Dix, Laura M. L. ;
van Bel, Frank ;
Baerts, Wim ;
Lemmers, Petra M. A. .
PEDIATRIC RESEARCH, 2013, 74 (05) :557-563
[4]   Dynamic evaluation of near-infrared peripheral oximetry in healthy volunteers: A comparison between INVOS and EQUANOX [J].
Fellahi, Jean-Luc ;
Butin, Geraldine ;
Fischer, Marc-Olivier ;
Zamparini, Guillaume ;
Gerard, Jean-Louis ;
Hanouz, Jean-Luc .
JOURNAL OF CRITICAL CARE, 2013, 28 (05) :881.e1-881.e6
[5]   Comparison of two spatially resolved NIRS oxygenation indices [J].
Gagnon R.E. ;
Macnab A.J. ;
Gagnon F.A. ;
Blackstock D. ;
LeBlanc J.G. .
Journal of Clinical Monitoring and Computing, 2002, 17 (7-8) :385-391
[6]  
Hessel TW, 2014, ACTA PAEDIAT
[7]  
Hyttel-Sorensen S, 2013, DAN MED J, V60
[8]   Evaluation of Pediatric Near-Infrared Cerebral Oximeter for Cardiac Disease [J].
Kreeger, Renee N. ;
Ramamoorthy, Chandra ;
Nicolson, Susan C. ;
Ames, Warwick A. ;
Hirsch, Russel ;
Peng, Lynn F. ;
Glatz, Andrew C. ;
Hill, Kevin D. ;
Hoffman, Joan ;
Tomasson, Jon ;
Kurth, C. Dean .
ANNALS OF THORACIC SURGERY, 2012, 94 (05) :1527-1533
[9]   Does the resting state connectivity have hemispheric asymmetry? A near-infrared spectroscopy study [J].
Medvedev, Andrei V. .
NEUROIMAGE, 2014, 85 :400-407
[10]   Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children [J].
Nagdyman, N ;
Fleck, T ;
Schubert, S ;
Ewert, P ;
Peters, B ;
Lange, PE ;
Abdul-Khaliq, H .
INTENSIVE CARE MEDICINE, 2005, 31 (06) :846-850