A separated primary and scatter model for independent dose calculation of intensity modulated radiotherapy

被引:7
作者
Baker, Colin R.
Clements, Richard
Gately, Alan
Budgell, Geoff J.
机构
[1] Univ Liverpool, Sch Hlth Sci, Div Med Imaging & Radiotherapy, Liverpool L69 3GB, Merseyside, England
[2] Clatterbridge Ctr Oncol, Dept Phys, Wirral, Merseyside, England
[3] Christie Hosp NHS Trust, N Western Med Phys, Manchester M20 4BX, Lancs, England
关键词
IMRT; dose calculation; verification; MONITOR UNIT CALCULATION; RADIATION-THERAPY COMMITTEE; PROSTATE-CANCER; CLINICAL IMPLEMENTATION; MULTILEAF COLLIMATOR; QUALITY-ASSURANCE; HEAD SCATTER; IMRT; VERIFICATION; CHECKING;
D O I
10.1016/j.radonc.2006.08.011
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Independent checking of beam monitor units is an essential step in the preparation of any radiotherapy plan. The present work describes a simple model for calculating a point dose from an arbitrary number of irregular, asymmetric fields, typical of beam segments used for IMRT. Materials and methods: Primary and scatter dose contributions were separated using a two-parameter exponential fit to measured beam data, from which differential scatter is determined. A total of 60 IMRT patient plans for a five-field prostate class solution were investigated to validate the model. Results: The average difference between the model's prediction and direct measurement of reference dose was found to be -0.6% (ranging from -2.9% to +1.6%), with a standard deviation of 1.0%. This compares well with the observed average difference between treatment planning system prediction and direct measurement of +0.8% (SD 0.6%). Conclusions: The model is shown to provide a reliable and accurate independent check of planning system monitor units for the prostate IMRT plans studied. Implementation of the model could significantly reduce the time needed for point dose verification of IMRT plans currently performed by direct measurement. (c) 2006 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:385 / 390
页数:6
相关论文
共 27 条
[1]   Prediction of the benefits from dose-escalated hypofractionated intensity-modulated radiotherapy for prostate cancer [J].
Amer, AM ;
Mott, J ;
Mackay, RI ;
Williams, PC ;
Livsey, J ;
Logue, JP ;
Hendry, JH .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01) :199-207
[2]  
[Anonymous], 2004, MAN CANC SERV
[3]  
Ayyangar K M, 2001, Med Dosim, V26, P29, DOI 10.1016/S0958-3947(00)00057-1
[4]  
BAKER CR, 2002, RADIOTHER ONCOL S1, V64, pS36
[5]   Theoretical considerations of monitor unit calculations for intensity modulated beam treatment planning [J].
Boyer, A ;
Xing, L ;
Ma, CM ;
Curran, B ;
Hill, R ;
Kania, A ;
Bleier, A .
MEDICAL PHYSICS, 1999, 26 (02) :187-195
[6]   Independent monitor unit calculation for intensity modulated radiotherapy using the MIMIC multileaf collimator [J].
Chen, Z ;
Xing, L ;
Nath, R .
MEDICAL PHYSICS, 2002, 29 (09) :2041-2051
[7]   Patient-specific point dose measurement for IMRT monitor unit verification [J].
Dong, L ;
Antolak, J ;
Salehpour, M ;
Forster, K ;
O'Neill, L ;
Kendall, R ;
Rosen, I .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (03) :867-877
[8]   Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee [J].
Ezzell, GA ;
Galvin, JM ;
Low, D ;
Palta, JR ;
Rosen, I ;
Sharpe, MB ;
Xia, P ;
Xiao, Y ;
Xing, L ;
Yu, CX .
MEDICAL PHYSICS, 2003, 30 (08) :2089-2115
[9]   Head scatter modelling for irregular field shaping and beam intensity modulation [J].
Hounsell, AR ;
Wilkinson, JM .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (09) :1737-1749
[10]   TISSUE STANDARD RATIOS FOR IRREGULARLY SHAPED RADIOTHERAPY FIELDS [J].
HOUNSELL, AR ;
WILKINSON, JM .
BRITISH JOURNAL OF RADIOLOGY, 1990, 63 (752) :629-634