Finite element analysis of stresses developed in the blood sac of a left ventricular assist device

被引:6
作者
Donahue, T. L. Haut [1 ]
Dehlin, W. [1 ]
Gillespie, J. [1 ]
Weiss, W. J. [2 ]
Rosenberg, G. [2 ]
机构
[1] Michigan Technol Univ, Dept Mech Engn, Houghton, MI 49331 USA
[2] Milton S Hershey Med Ctr, Penn State Coll Med, Div Artificial Organs, Hershey, PA USA
关键词
Finite element analysis; LVAD; SPEUU; Artificial heart; MECHANICAL-BEHAVIOR; STRAIN BEHAVIOR; POLYURETHANES; EXPERIENCE; SYSTEM; HEART;
D O I
10.1016/j.medengphy.2008.11.011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The goal of this research is to develop a 3D finite element (FE) model of a left ventricular assist device (LVAD) to predict stresses in the blood sac. The hyperelastic stress-strain curves for the segmented poly(ether polyurethane urea) (SPEUU) blood sac were determined in both tension and compression using a servo-hydraulic testing system at various strain rates. Over the range of strain rates studied, the sac was not strain rate sensitive, however the material response was different for tension versus compression. The experimental tension and compression properties were used in a FE model that consisted of the pusher plate, blood sac and pump case. A quasi-static analysis was used to allow for nonlinearities due to contact and material deformation. The 3D FE model showed that blood sac stresses are not adversely affected by the location of the inlet and outlet ports of the device and that over the systolic ejection phase of the simulation the prediction of blood sac stresses from the full 3D model and an axisymmetric model are the same. Minimizing stresses in the blood sac will increase the longevity of the blood sac in vivo. (C) 2008 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:454 / 460
页数:7
相关论文
共 23 条
[1]  
Abraham GA, 1997, J APPL POLYM SCI, V65, P1193, DOI 10.1002/(SICI)1097-4628(19970808)65:6<1193::AID-APP15>3.0.CO
[2]  
2-V
[3]   Biostability of thoralon left ventricular assist device blood pumping sacs after long-term clinical use [J].
Babu, R ;
Gill, R ;
Farrar, D .
ASAIO JOURNAL, 2004, 50 (05) :479-484
[4]  
Donahue Tammy L Haut, 2003, Comput Methods Biomech Biomed Engin, V6, P7, DOI 10.1080/1025584031000078943
[5]   Initial experience with the AbioCor Implantable Replacement Heart System [J].
Dowling, RD ;
Gray, LA ;
Etoch, SW ;
Laks, H ;
Marelli, D ;
Samuels, L ;
Entwistle, J ;
Couper, G ;
Vlahakes, GJ ;
Frazier, OH .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2004, 127 (01) :131-141
[6]  
HAYASHI K, 1994, AM SOC TEST MATER, V1173, P9, DOI 10.1520/STP18090S
[7]   Fluid dynamic analysis of the 50 cc Penn State artificial heart under physiological operating conditions using particle image velocimetry [J].
Hochareon, P ;
Manning, KB ;
Fontaine, AA ;
Tarbell, JM ;
Deutsch, S .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2004, 126 (05) :585-593
[8]   In vivo and in vitro stability of modified poly(urethaneurea) blood sacs [J].
Liu, Q ;
Runt, J ;
Felder, G ;
Rosenberg, G ;
Snyder, AJ ;
Weiss, WJ ;
Lewis, J ;
Werley, T .
JOURNAL OF BIOMATERIALS APPLICATIONS, 2000, 14 (04) :349-366
[9]  
MCCARTHY PM, 1991, J THORAC CARDIOV SUR, V102, P578
[10]  
MCCARTHY PM, 1991, J THOR CARDIOVASC SU, V102, P586