Dual-band Fourier domain optical coherence tomography with depth-related compensations

被引:0
作者
Zhang, Miao [1 ]
Ma, Lixin [2 ,3 ]
Yu, Ping [1 ]
机构
[1] Univ Missouri, Dept Phys & Astron, Columbia, MO 65211 USA
[2] Univ Missouri, Dept Radiol, Columbia, MO 65211 USA
[3] Univ Missouri, Harry S Truman Mem Vet Hosp, Columbia, MO 65211 USA
来源
BIOMEDICAL OPTICS EXPRESS | 2014年 / 5卷 / 01期
基金
美国国家科学基金会;
关键词
SPECTRAL-DOMAIN; IN-VIVO; FULL-RANGE; SWEPT-SOURCE; SPECKLE REDUCTION; HIGH-SPEED; RESOLUTION; ATTENUATION; IMAGES; TISSUE;
D O I
10.1364/BOE.5.000167
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Dual-band Fourier domain optical coherence tomography (FDOCT) provides depth-resolved spectroscopic imaging that enhances tissue contrast and reduces image speckle. However, previous dual-band FD-OCT systems could not correctly give the tissue spectroscopic contrast due to depth-related discrepancy in the imaging method and attenuation in biological tissue samples. We designed a new dual-band full-range FDOCT imaging system and developed an algorithm to compensate depth-related fall-off and light attenuation. In our imaging system, the images from two wavelength bands were intrinsically overlapped and their intensities were balanced. The processing time of dual-band OCT image reconstruction and depth-related compensations were minimized by using multiple threads that execute in parallel. Using the newly developed system, we studied tissue phantoms and human cancer xenografts and muscle tissues dissected from severely compromised immune deficient mice. Improved spectroscopic contrast and sensitivity were achieved, benefiting from the depth-related compensations. (C) 2013 Optical Society of America
引用
收藏
页码:167 / 182
页数:16
相关论文
共 50 条
  • [1] Extended measuring depth dual-wavelength Fourier domain optical coherence tomography
    Al-Mohamedi, Haroun
    Kelly-Perez, Ismael
    Oltrup, Theo
    Cayless, Alan
    Bende, Thomas
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2021, 66 (06): : 557 - 562
  • [2] Fourier domain optical coherence tomography with ultralong depth range
    Ding Z.
    Shen Y.
    Bao W.
    Li P.
    Frontiers of Optoelectronics, 2015, 8 (2) : 163 - 169
  • [3] Removal of Mirror Image Using Reference Beam Phase Modulation in Dual-Band Fourier-Domain Optical Coherence Tomography
    Zhang, Miao
    Yu, Ping
    Ma, Lixin
    2012 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO), 2012,
  • [4] Simultaneous dual-band optical coherence tomography for endoscopic applications
    Xu, Jianbing
    Yu, Luoqin
    Wei, Xiaoming
    Wang, Xie
    Chui, Po Ching
    Chan, Kin Tak
    Lam, Edmund Y.
    Lee, Nikki P.
    Wong, Kenneth K. Y.
    JOURNAL OF BIOMEDICAL OPTICS, 2014, 19 (12)
  • [5] Simultaneous dual-band optical coherence tomography in the spectral domain for high resolution in vivo imaging
    Cimalla, Peter
    Walther, Julia
    Mehner, Mirko
    Cuevas, Maximiliano
    Koch, Edmund
    OPTICS EXPRESS, 2009, 17 (22): : 19486 - 19500
  • [6] Three-Dimensional Full-Range Dual-Band Fourier Domain Optical Coherence Tomography Accelerated by Graphic Processing Unit
    Zhang, Miao
    Ma, Lixin
    Yu, Ping
    IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 2019, 25 (01)
  • [7] A depth resolution enhancement technique in fourier domain optical coherence tomography
    Guo, Xin
    Wang, Xiangzhao
    Nan, Nan
    Li, Zhongliang
    Guangxue Xuebao/Acta Optica Sinica, 2015, 35 (03):
  • [8] Dual-band infrared optical coherence tomography using a single supercontinuum source
    Zorin, Ivan
    Gattinger, Paul
    Brandstetter, Markus
    Heise, Bettina
    OPTICS EXPRESS, 2020, 28 (06): : 7858 - 7874
  • [9] Effects of scattering on spectral shape and depth resolution in Fourier domain optical coherence tomography
    Wang, X. (wxz26267@siom.ac.cn), 1600, Chinese Optical Society (34):
  • [10] Fourier-Domain Optical Coherence Tomography Signal Analysis and Numerical Modeling
    Kalkman, J.
    INTERNATIONAL JOURNAL OF OPTICS, 2017, 2017