Titanium orthodontic brackets: structure, composition, hardness and ionic release

被引:33
作者
Gioka, C
Bourauel, C
Zinelis, S
Eliades, T
Silikas, N
Eliades, G
机构
[1] Univ Athens, Sch Dent, Biomat Lab, Athens, Greece
[2] Univ Bonn, Sch Dent, D-5300 Bonn, Germany
[3] Univ Manchester, Sch Dent, Biomat Sci Unit, Manchester, Lancs, England
关键词
titanium brackets; composition; microstructure; hardness; ionic release;
D O I
10.1016/j.dental.2004.02.008
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The aim of the present study was to investigate the composition, morphology, bulk structure and ionic release of two brands of titanium orthodontic brackets: Orthos2 (Ormco, USA) and Rematitan (Dentaurum, Germany). Methods: Five specimens of each group were examined with computerized X-ray microtomography, to reveal the morphology and structure of brackets, whilst resin-embedded and metallographically polished specimens were subjected to SEM/EDS analysis and Vickers microhardness measurements. Brackets were also maintained in 0.9% saline for 2 months and the ionic release in the immersion medium was determined with Inductively Coupled Plasma Atomic Emission Spectroscopy. The results of the hardness and ionic release measurements were statistically analyzed with two-way ANOVA and Tukey's test (alpha = 0.05). Results: Orthos2 brackets consisted of two parts, the base (commercially pure Ti grade II) and the wing (Ti-6Al-4V alloy), joined together by laser welding, producing large gaps along the base-wing interface. The base was of lower hardness (H = 145), than the wing (H = 392) and incorporated a standard foil base-mesh pad. Rematitan brackets consisted of commercially pure Ti grade IV, with a single-piece manufacturing pattern of virtually identical hardness (p > 0.05) at the base and wings, featuring a laser-etched base-mesh pad. The hardness of the Rematitan brackets was significantly lower than the hardness of the Orthos2 wings, but double them hardness of the Orthos2 base. Released Ti levels were below the threshold level (1 ng/ml) of analysis for both materials, whilst traces of Al (3 ppm) and V (2 ppm) were found in the immersion media for Ti-6Al-4V alloy. Significance: The structural and hardness differences found may influence the torque transfer characteristics from activated archwires to the brackets and the crevice corrosion potential at the base-wing interface (Orthos2). The detection of Al and V in the immersion medium (Orthos2) may imply a differrent biological response from the two types of Ti brackets. (C) 2004 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:693 / 700
页数:8
相关论文
共 31 条
[1]   BIODEGRADATION OF ORTHODONTIC APPLIANCES .2. CHANGES IN THE BLOOD LEVEL OF NICKEL [J].
BISHARA, SE ;
BARRETT, RD ;
SELIM, MI .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1993, 103 (02) :115-119
[2]  
BLACK J, 1999, BIOL PERFORMANCE MAT, P28
[3]  
BOYER R, 1994, TITANIUM ALLOYS, P226
[4]  
BRANTLEY WA, 2001, ORTHODONTIC MAT SCI, P95
[5]  
DAVIS RM, 1996, ASTM STP, V1272, P17
[6]  
*DENT, 2001, ORTH CAT, P57
[7]  
Eliades T, 1997, INT J ORAL MAX IMPL, V12, P621
[8]   Salivary metal levels of orthodontic patients: a novel methodological and analytical approach [J].
Eliades, T ;
Trapalis, C ;
Eliades, G ;
Katsavrias, E .
EUROPEAN JOURNAL OF ORTHODONTICS, 2003, 25 (01) :103-106
[9]   Nickel content of as-received, retrieved, and recycled stainless steel brackets [J].
Eliades, T ;
Zinelis, S ;
Eliades, G ;
Athanasiou, AE .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2002, 122 (02) :217-220
[10]  
Eliades T, 2002, ANGLE ORTHOD, V72, P222