Coaxial technique: Approach to breast core biopsies

被引:31
作者
Helbich, TH
Mayr, W
Schick, S
Youssefzadeh, S
Rudas, M
Taucher, S
Wagner, T
Kelkar, P
Wolf, G
Thurnher, M
Mostbeck, GH
机构
[1] UNIV VIENNA,DEPT RADIOL,VIENNA,AUSTRIA
[2] UNIV VIENNA,DEPT BIOMED ENGN & PHYS,VIENNA,AUSTRIA
[3] UNIV VIENNA,DEPT CLIN PATHOL,VIENNA,AUSTRIA
[4] UNIV VIENNA,DEPT SPECIAL GYNECOL,VIENNA,AUSTRIA
关键词
biopsies; technology; breast; biopsy; breast neoplasms; diagnosis; stereotaxis; ultrasound; (US); guidance;
D O I
10.1148/radiology.203.3.9169689
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To assess the feasibility of the coaxial core breast biopsy technique performed under stereotactic and ultrasound (US) guidance in vitro and in vivo. MATERIALS AND METHODS: Biopsies were performed in vitro and in vivo with a coaxial technique. In vitro, the true needle-tip deviation was measured with a breast phantom on a stereotactic device with alteration of x and y axes. In vitro US studies were performed to evaluate the optimal technique for harvesting sufficient material for histologic work-up. In 205 patients, coaxial biopsy was performed in 210 suspicious lesions under US (61 lesions) or stereotactic (patient in the sitting position, n = 67; patient in the prone position, n = 82) guidance. In addition, the coaxial system was used for preoperative localization. Surgery and histologic work-up were performed in all eases. RESULTS: In vitro, the true needle-tip deviation was found to be less than indicated on the stereotactic device. A factor was introduced to correct this error. For US guidance, angulation or rotation of the coaxial needle within the lesion proved to be the best technique to increase the size of histologic specimen. Of the 210 lesions, 112 were benign and 98 were malignant. Agreement between biopsy results and final postsurgical histologic analysis was found in 205 cases (98%). CONCLUSION: The coaxial breast biopsy technique is an accurate, simple, and time-saving method performed under stereotactic or US guidance.
引用
收藏
页码:684 / 690
页数:7
相关论文
共 31 条
[1]  
Altman DG, 1990, PRACTICAL STAT MED R
[2]   When is core breast biopsy or fine-needle aspiration not enough [J].
Berg, WA .
RADIOLOGY, 1996, 198 (02) :313-315
[3]   STEREOTAXIC NEEDLE CORE BIOPSY OF BREAST-LESIONS USING A REGULAR MAMMOGRAPHIC TABLE WITH AN ADAPTABLE STEREOTAXIC DEVICE [J].
CAINES, JS ;
MCPHEE, MD ;
KONOK, GP ;
WRIGHT, BA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 163 (02) :317-321
[4]   FINE-NEEDLE ASPIRATION CYTOLOGY OF NONPALPABLE BREAST-LESIONS - ULTRASOUND VERSUS STEREOTAXIC GUIDANCE [J].
CIATTO, S ;
CATARZI, S ;
MORRONE, D ;
DELTURCO, MR .
RADIOLOGY, 1993, 188 (01) :195-198
[5]   NONPALPABLE BREAST-LESIONS - FINDINGS OF STEREOTAXIC NEEDLE-CORE BIOPSY AND FINE-NEEDLE ASPIRATION CYTOLOGY [J].
DOWLATSHAHI, K ;
YAREMKO, ML ;
KLUSKENS, LF ;
JOKICH, PM .
RADIOLOGY, 1991, 181 (03) :745-750
[6]   STEREOTAXIC CORE BIOPSY OF BREAST-LESIONS [J].
DRONKERS, DJ .
RADIOLOGY, 1992, 183 (03) :631-634
[7]   COAXIAL NEEDLES FOR REPEATED BIOPSY SAMPLING [J].
FREDERICK, PR ;
MILLER, MH ;
BAHR, AL ;
LONGSON, FW .
RADIOLOGY, 1989, 170 (01) :273-274
[8]   BREAST BIOPSY - A COMPARATIVE-STUDY OF STEREOTAXICALLY GUIDED CORE AND EXCISIONAL TECHNIQUES [J].
GISVOLD, JJ ;
GOELLNER, JR ;
GRANT, CS ;
DONOHUE, JH ;
SYKES, MW ;
KARSELL, PR ;
COFFEY, SL ;
JUNG, SH .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 162 (04) :815-820
[9]   SOLID BREAST-LESIONS - DIAGNOSIS WITH US-GUIDED FINE-NEEDLE ASPIRATION BIOPSY [J].
GORDON, PB ;
GOLDENBERG, SL ;
CHAN, NHL .
RADIOLOGY, 1993, 189 (02) :573-580
[10]   MALIGNANT SEEDING OF THE NEEDLE TRACK DURING STEREOTAXIC CORE NEEDLE BREAST BIOPSY [J].
HARTER, LP ;
CURTIS, JS ;
PONTO, G ;
CRAIG, PH .
RADIOLOGY, 1992, 185 (03) :713-714