The Treatment of Intertrochanteric Fractures: Results Using an Intramedullary Nail With Integrated Cephalocervical Screws and Linear Compression

被引:114
作者
Ruecker, Andreas H. [1 ]
Rupprecht, M. [1 ]
Gruber, M. [1 ]
Gebauer, M. [1 ]
Barvencik, F. [1 ]
Briem, D. [1 ]
Rueger, J. M. [1 ]
机构
[1] Univ Hamburg Hosp, Dept Trauma Hand & Reconstruct Surg, D-20246 Hamburg, Germany
关键词
proximal femoral fracture; per-/subtrochanteric fracture; intramedullary nail; functional outcome; DYNAMIC HIP SCREW; TROCHANTERIC GAMMA-NAIL; DISPLACED SUBCAPITAL FRACTURES; PROXIMAL FEMORAL FRACTURE; INTERNAL-FIXATION; SUBTROCHANTERIC FRACTURES; PERTROCHANTERIC FRACTURES; FINITE-ELEMENT; LAG SCREW; PLATE;
D O I
10.1097/BOT.0b013e31819211b2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: A new device for the treatment of intertrochanteric fractures that uses 2 cephalocervical screws in an integrated mechanism allowing linear intraoperative compression and rotational stability of the head/neck fragment has been developed. The aim of this study was to describe the results using this device for the treatment of stable and unstable intertrochanteric fractures. Design: Prospective, consecutive. Setting: Academic Trauma Center. Methods: Between March 1, 2005, and July 31, 2006, 100 consecutive patients with an intertrochanteric fracture were treated with a new trochanteric antegrade nail (InterTan; Smith-Nephew, Memphis, TN). All living patients were followed up for a minimum of 1 year postoperatively (range 12-27 months). Clinical and radiographic examinations were performed until healing and at the 1-year anniversary of the index procedure. Healing, pain with ambulation, return to activities of daily living, the modified Harris hip score, and Barthel Index were used to evaluate outcomes. Results: The mean age of the patients was 81.2 (+/-11.3) years. Thirty-seven patients died, 12 were too infirmed for follow-up, and 3 could not be located, leaving 48 patients available for final evaluation. The average surgical time was 41 minutes (13-95 minutes). This rose significantly with the complexity of the fracture (OTA/AO classification: A1 versus A3, P = 0.016). All fractures healed within 16 weeks (range 10-16 weeks). Radiographic analysis at healing revealed no loss of reduction,, no uncontrolled collapse of the neck, no nonunions, no femoral shaft fractures, and no implant failures. Two cases in the series were poorly reduced and settled into varus malalignment. There was no varus malposition seen in the remaining 46 fractures. The mean prefracture Harris hip score (75.1 +/- 13.4) was significantly reduced at the time of follow-up (70.3 +/- 14.5, P = 0.003); 58% of the patients recovered their prefracture status. No significant difference was seen for the Barthel Index. Conclusions: The InterTan device appears to be a reliable implant for the treatment of intertrochanteric femoral fractures. Its design provides for stability against rotation and minimizes neck malunions (shortening) through linear intraoperative compression of the head/neck segment to the shaft. As a result of the negligible complication rate and improved clinical outcomes, this implant is now the standard treatment for all intertrochanteric fractures at our institution.
引用
收藏
页码:22 / 30
页数:9
相关论文
共 64 条
  • [1] Prospective randomized controlled trial of an intramedullary nail versus dynamic screw and plate for intertrochanteric fractures of the femur
    Adams, CI
    Robinson, CM
    Court-Brown, CM
    McQueen, MM
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2001, 15 (06) : 394 - 400
  • [2] AHRENGART L, 2002, CLIN ORTHOP RELAT R, V401, P209
  • [3] Alvarez JR, 1998, CLIN ORTHOP RELAT R, P62
  • [4] [Anonymous], CLIN ORTHOP
  • [5] [Anonymous], 1997, AM J MED
  • [6] GAMMA-NAIL VS COMPRESSION SCREW FOR TROCHANTERIC FEMORAL FRACTURES - 15 REOPERATIONS IN A PROSPECTIVE, RANDOMIZED STUDY OF 378 PATIENTS
    AUNE, AK
    EKELAND, A
    ODEGAARD, B
    GROGAARD, B
    ALHO, A
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1994, 65 (02): : 127 - 130
  • [7] Bess R J, 1997, J South Orthop Assoc, V6, P173
  • [8] A new intramedullary nail device for the treatment of intertrochanteric hip fractures: Perioperative experience
    Bienkowski, Peter
    Reindl, Rudolf
    Berry, Gregory K.
    Iakoub, Elena
    Harvey, Edward J.
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2006, 61 (06): : 1458 - 1462
  • [9] BOHNEL P, 1987, UNFALLCHIRURG, V90, P292
  • [10] FIXATION OF INTERTROCHANTERIC FRACTURES OF THE FEMUR - A RANDOMIZED PROSPECTIVE COMPARISON OF THE GAMMA-NAIL AND THE DYNAMIC HIP SCREW
    BRIDLE, SH
    PATEL, AD
    BIRCHER, M
    CALVERT, PT
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (02): : 330 - 334