共 50 条
Undersampled projection-reconstruction imaging for time-resolved contrast-enhanced imaging
被引:0
|作者:
Vigen, KK
Peters, DC
Grist, TM
Block, WF
Mistretta, CA
机构:
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53792 USA
[2] Univ Wisconsin, Dept Radiol, Madison, WI 53792 USA
关键词:
undersampled projection reconstruction;
contrast-enhanced MR angiography;
time-resolved MR imaging;
D O I:
10.1002/(SICI)1522-2594(200002)43:2<170::AID-MRM2>3.0.CO;2-P
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
In time-resolved contrast-enhanced 3D MR angiography, spatial resolution is traded for high temporal resolution. A hybrid method is presented that attempts to reduce this tradeoff in two of the spatial dimensions. It combines an undersampled projection acquisition in two dimensions with variable rate k-space sampling in the third. Spatial resolution in the projection plane is determined by readout resolution and is limited primarily by signal-to-noise ratio. Oversampling the center of k-space combined with temporal k-space interpolation provides time frames with minimal venous contamination, Results demonstrating improved resolution in phantoms and volunteers are presented using angular undersampling factors up to eight with acceptable projection reconstruction artifacts. (C) 2000 Wiley-Liss, Inc.
引用
收藏
页码:170 / 176
页数:7
相关论文