Agreement between left ventricular ejection fraction assessed in patients with gated IQ-SPECT and conventional imaging

被引:3
作者
Leva, Lucia [1 ]
Matheoud, Roberta [2 ]
Sacchetti, Gianmauro [1 ]
Carriero, Alessandro [3 ]
Brambilla, Marco [2 ]
机构
[1] Univ Hosp Maggiore della Carita, Dept Nucl Med, Novara, Italy
[2] Univ Hosp Maggiore della Carita, Dept Med Phys, Novara, Italy
[3] Univ Hosp Maggiore della Carita, Dept Radiol, Novara, Italy
关键词
CAD; gated SPECT; instrumentation; perfusion agents; statistical analysis; ITERATIVE RECONSTRUCTION ALGORITHMS; MYOCARDIAL-PERFUSION SPECT; RESOLUTION RECOVERY; VOLUME MEASUREMENTS; VALIDATION; SYSTEM; QUANTIFICATION; ASSOCIATION; ACQUISITION; PARAMETERS;
D O I
10.1007/s12350-018-1457-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The aim of the study was to assess the agreement between the left ventricular ejection fraction (LVEF) values obtained with IQ-SPECT and those obtained with a conventional gamma camera equipped with low-energy high-resolution (LEHR), considered as the method of reference. Methods Gated-stress MPI using 99mTc-tetrofosmin was performed in 55 consecutive patients. The patients underwent two sequential acquisitions (Method A and B) performed on Symbia-IQ SPECT with different acquisition times and one (Method C) on a Ecam SPECT equipped with LEHR collimators. The values of the different datasets were compared using the Bland-Altman analysis method: the bias and the limits of agreement (LA) were estimated in a head-to-head comparison of the three protocols. Results In the (Method A-Method C) comparison for LVEF, the bias was 3.8% and the LAs ranged from - 9.3% to 16.8%. The agreement was still lower between Method B and C, whilst only slightly improved when Methods A and B were compared. Conclusions The wide amplitude in LA intervals of about 30% indicates that IQ and LEHR GSPECT are not interchangeable. The values obtained with IQ-SPECT should only be used with caution when evaluating the functional state of the heart.
引用
收藏
页码:1714 / 1724
页数:11
相关论文
共 50 条
  • [31] Prognostic value of absent left ventricular ejection fraction reserve with regadenoson SPECT MPI
    Phillip Smith
    Ayman Farag
    Pradeep Bhambhvani
    Ami Iskandrian
    Fadi G. Hage
    Journal of Nuclear Cardiology, 2022, 29 : 978 - 986
  • [32] Gated Myocardial Perfusion SPECT: Algorithm-Specific Influence of Reorientation on Calculation of Left Ventricular Volumes and Ejection Fraction
    Knollmann, Daniela
    Winz, Oliver H.
    Meyer, Philipp T.
    Raptis, Mardjan
    Krohn, Thomas
    Koch, Karl-Christian
    Schaefer, Wolfgang M.
    JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (10) : 1636 - 1642
  • [33] Impact of Wiener filter in determining the left ventricular volume and ejection fraction using thallium-201 gated SPECT
    Manrique, A
    Hitzel, A
    Gardin, I
    Dacher, JN
    Véra, P
    NUCLEAR MEDICINE COMMUNICATIONS, 2003, 24 (08) : 907 - 914
  • [34] Sex-specific reference limits of left ventricular ejection fraction and volumes estimated by gated myocardial perfusion imaging for low-risk patients in China: a comparison between three quantitative algorithms
    Li, Jiajun
    Guo, Lin
    Liu, Jingzhe
    Fang, Wei
    Sun, Xiaoxin
    He, Zuoxiang
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2022, 12 (01) : 144 - 158
  • [35] Prognostic impacts of changes in left ventricular ejection fraction in heart failure patients with preserved left ventricular ejection fraction
    Yoshihisa, Akiomi
    Sato, Yu
    Kanno, Yuki
    Takiguchi, Mai
    Yokokawa, Tetsuro
    Abe, Satoshi
    Misaka, Tomofumi
    Sato, Takamasa
    Oikawa, Masayoshi
    Kobayashi, Atsushi
    Yamaki, Takayoshi
    Kunii, Hiroyuki
    Takeishi, Yasuchika
    OPEN HEART, 2020, 7 (01):
  • [36] Stress-related variations in left ventricular function as assessed with gated myocardial perfusion SPECT
    Verberne, HJ
    Dijkgraaf, MGW
    Somsen, GA
    van Eck-Smit, BLF
    JOURNAL OF NUCLEAR CARDIOLOGY, 2003, 10 (05) : 456 - 463
  • [37] Left Ventricular Mechanical Dyssynchrony as Assessed by Phase Analysis of ECG-Gated SPECT Myocardial Perfusion Imaging
    Chen, Ji
    Garcia, Ernest V.
    Lerakis, Stamatios
    Henneman, Maureen M.
    Bax, Jeroen J.
    Trimble, Mark A.
    Borges-Neto, Salvador
    Velazquez, Eric J.
    Iskandrian, Ami E.
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2008, 25 (10): : 1186 - 1194
  • [38] Potential Impact of Hybrid CZT SPECT/CT Imaging on Estimation Accuracy of Left Ventricular Volumes and Ejection Fraction: A Phantom Study
    Liu, Yi-Hwa
    Sandoval, Veronica
    Sinusas, Albert J.
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [39] Potential Impact of SPECT Resolution on Quantification of Left Ventricular Volumes and Ejection Fraction: A Phantom Study
    Liu, Hui
    Aslan, Mehmet
    Sandoval, Veronica
    Liu, Yi-Hwa
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2022, 42 (05) : 734 - 743
  • [40] Left ventricular dyssynchrony assessed by gated SPECT phase analysis is an independent predictor of death in patients with advanced coronary artery disease and reduced left ventricular function not undergoing cardiac resynchronization therapy
    Uebleis, Christopher
    Hellweger, Stefan
    Laubender, Ruediger Paul
    Becker, Alexander
    Sohn, Hae-Young
    Lehner, Sebastian
    Haug, Alexander
    Bartenstein, Peter
    Cumming, Paul
    Van Kriekinge, Serge D.
    Slomka, Piotr J.
    Hacker, Marcus
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 (10) : 1561 - 1569