Forces and displacements in colon surgery

被引:56
作者
de Visser, H [1 ]
Heijnsdijk, EAM [1 ]
Herder, JL [1 ]
Pistecky, PV [1 ]
机构
[1] Delft Univ Technol, MISIT, NL-2628 CD Delft, Netherlands
来源
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES | 2002年 / 16卷 / 10期
关键词
colon surgery; laparoscopy; pull forces; safety; tissue manipulation; trocar placement;
D O I
10.1007/s00464-002-9003-7
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: This study aims to find criteria for designing safer laparoscopic graspers for manipulation of delicate tissues. Since the experiment was performed in an open-surgery setup without any movement restrictions, the results are relevant to general surgery as well as laparoscopic. Method: Four surgeons were asked to apply via two clamps enough force to a pig's colon to stretch the mesocolon for dissection. Directions, magnitudes, and application points of the forces were recorded. Results: The surgeons applied per clamp on average 2.5 N and maximally 5 N to the colon. These forces were at an angle of about 40degrees to 70degrees relative to the horizontal plane and approximately 60degrees relative to each other. Conclusions: Safe laparoscopic graspers should be able to transmit at least a 5 N pull force to the tissue without damaging it. To minimize the forces used, the angles mentioned should be taken into account when placing the trocars.
引用
收藏
页码:1426 / 1430
页数:5
相关论文
共 7 条
  • [1] Bannenberg JIG, 1994, MINIM INVAS THER S1, V3, P37
  • [2] High pressures are generated at the tip of laparoscopic graspers
    Cartmill, JA
    Shakeshaft, AJ
    Walsh, WR
    Martin, CJ
    [J]. AUSTRALIAN AND NEW ZEALAND JOURNAL OF SURGERY, 1999, 69 (02): : 127 - 130
  • [3] Sensitivity of laparoscopic dissectors - What can you feel?
    den Boer, KT
    Herder, JL
    Sjoerdsma, W
    Meijer, DW
    Gouma, DJ
    Stassen, HG
    [J]. SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1999, 13 (09): : 869 - 873
  • [4] Frank T, 1995, Proc Inst Mech Eng H, V209, P111, DOI 10.1243/PIME_PROC_1995_209_327_02
  • [5] Patterns of failure at the instrument-tissue interface
    Marucci, DD
    Cartmill, JA
    Walsh, WR
    Martin, CJ
    [J]. JOURNAL OF SURGICAL RESEARCH, 2000, 93 (01) : 16 - 20
  • [6] THE PALPATOR - AN INSTRUMENT FOR MEASURING THE POSITIONS OF BONES IN 3 DIMENSIONS
    PRONK, GM
    VANDERHELM, FCT
    [J]. JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1991, 15 (01) : 15 - 20
  • [7] Toledo L, 1999, ANN CHIR, V53, P587