An automated microfluidic-based immunoassay cartridge for allergen screening and other multiplexed assays

被引:25
|
作者
Tai, Lung-Wen [1 ]
Tseng, Ko-Yuan [1 ]
Wang, Shaw-Tsu [1 ]
Chiu, Chao-Cheng [1 ]
Kow, Chak Hone [1 ]
Chang, Ping [1 ]
Chen, Chichen [1 ]
Wang, Jiu-Yao [2 ,3 ]
Webster, James R. [1 ]
机构
[1] Agnitio Sci & Technol, Hsinchu 30075, Taiwan
[2] Natl Cheng Kung Univ, Coll Med, Dept Pediat, Tainan 70101, Taiwan
[3] Natl Cheng Kung Univ, Ctr Micronano Sci & Technol, Tainan 70101, Taiwan
关键词
Microfluidic; Allergen; Immunoassay; Microarray; CROSS-REACTIVITY; IGE ANTIBODIES; SERUM; MICROARRAYS; AMPLIFICATION; PERFORMANCE; SYSTEM; DEVICE;
D O I
10.1016/j.ab.2009.05.009
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A microfluidic cartridge and system for multiplexed immunoassays is described. The passive microfluidic cartridge was composed of three layers of injection molded plastic sealed together using a thermal staking technique. Using this platform technology, a specific immunoglobulin E (IgE) panel assay was constructed. Allergen extract targets, positive and negative controls, and IgE calibration standards were immobilized within the cartridge as a microarray. A computer-controlled solenoid array provided the necessary actuation force for Pumping reagents within the cartridge to perform an automated, chemiluminescent indirect immunoassay. A 20-target allergen extract panel was demonstrated on the device with a total analysis time of 27 min. Allergen screening results showed 84% agreement for 3 house dust mites (N = 300) compared with a commercial test and 80% agreement overall (N = 978). Average coefficients of variation (N = 80) were measured as 20.5% for low/medium levels and 20.4% for medium/high levels. The average limit of detection (N = 160) was measured at 0.535 AU, and cutoff levels of 1.0 AU were estimated at less than 1 IU/ml (2.4 ng/ml). Such a system has potential applications in decentralized allergen screening as well as in other near-patient diagnostic immunoassays where multiplexed analysis, ease of use, and short analysis time are critical. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:98 / 105
页数:8
相关论文
共 50 条
  • [1] FAST AUTOMATED MICROFLUIDIC-BASED MULTIPLEXED IMMUNOFLUORESCENCE FOR TUMOR MICROENVIRONMENT ANALYSIS
    Kehren, A.
    Procopio, M. G.
    Pelz, B.
    Siddiqui, Z.
    Roman, K.
    Adnane, S.
    Brajkovic, S.
    Hoyt, C.
    Dupouy, D. G.
    Soltermann, A.
    JOURNAL FOR IMMUNOTHERAPY OF CANCER, 2020, 8 : A25 - A25
  • [2] Fully Automated Quantification of Insulin Concentration Using a Microfluidic-Based Chemiluminescence Immunoassay
    Yao, Ping
    Liu, Zhu
    Tung, Steve
    Dong, Zaili
    Liu, Lianqing
    JALA, 2016, 21 (03): : 387 - 393
  • [3] Microfluidic-based automated multiplex immunophenotyping and imaging
    Pelz, B.
    Migliozzi, D.
    Cappi, G.
    Dupouy, D.
    Gijs, M.
    VIRCHOWS ARCHIV, 2018, 473 : S1 - S1
  • [4] Determination of multiple allergen-specific IgE by microfluidic immunoassay cartridge in clinical settings
    Shyur, Shyh-Dar
    Jan, Ren-Long
    Webster, James R.
    Chang, Ping
    Lu, Yu-Jung
    Wang, Jiu-Yao
    PEDIATRIC ALLERGY AND IMMUNOLOGY, 2010, 21 (04) : 623 - 633
  • [5] Streptavidin hydrogels for microfluidic-based transcription factor assays
    Imamoglu, Husniye
    Chen, Hui
    Bindu, Nanduri
    Burgess, Shane C.
    Thomas, Gloria
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2006, 231
  • [6] Microfluidic-based automated chromogenic RNA in situ hybridization
    Procopio, M.
    Murlidhar, V.
    Doolittle, E.
    Zhang, B.
    Dupouy, D. G.
    Ma, X.
    VIRCHOWS ARCHIV, 2020, 477 : S146 - S146
  • [7] Different types of electrospun nanofibers and their effect on microfluidic-based immunoassay
    Mahmoudifard, Matin
    Vossoughi, Manouchehr
    Soleimani, Masoud
    POLYMERS FOR ADVANCED TECHNOLOGIES, 2019, 30 (04) : 973 - 982
  • [8] MULTIPLEXED, TEMPORAL IMMUNOPHENOTYPING WITH MICROFLUIDIC-BASED LOCALIZED SURFACE PLASMON RESONANCE
    Shanley, Thomas
    Cornell, Timothy
    McHugh, Walker
    Kurabayashi, Katsuo
    Oh, Bo-Ram
    Fu, Jianping
    CRITICAL CARE MEDICINE, 2016, 44 (12)
  • [9] A smart microfluidic-based fish farm for zebrafish screening
    Karthick Mani
    Chia-Yuan Chen
    Microfluidics and Nanofluidics, 2021, 25
  • [10] A smart microfluidic-based fish farm for zebrafish screening
    Mani, Karthick
    Chen, Chia-Yuan
    MICROFLUIDICS AND NANOFLUIDICS, 2021, 25 (03)