External Fixator Clamp Reuse Degrades Clamp Mechanical Performance

被引:10
作者
Gilley, Robert S. [1 ]
Beason, David P.
Snyder, David M.
Boston, Raymond C.
Radin, Alex
Kapatkin, Amy S.
Smith, Gail K.
Soslowsky, Louis J.
机构
[1] Univ Penn, Sch Vet Med, Dept Clin Studies, Philadelphia, PA 19104 USA
关键词
SKELETAL FIXATION; PIN INSERTION; HOLDING POWER; BONE; STRENGTH; CONFIGURATIONS; BIOMECHANICS; STIFFNESS; DESIGNS;
D O I
10.1111/j.1532-950X.2009.00521.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
To determine the effects of clamp reuse for the Kirschner-Ehmer (KE); Securos; and the IMEX-SK clamp. Experimental bench test of mechanical properties. Specially designed fixtures were used to mechanically test 18 clamps of each type with respect to 6 mechanical variables: fixator pin slippage, connecting bar slippage, fixator pin rotation, connecting bar rotation, and clockwise and counterclockwise clamp-bolt axis pivot. Each clamp was tested 6 times for each variable at 7.68 Nm of clamp-bolt tightening torque. Results were compared using repeated measures ANOVA. For the IMEX-SK clamp, a significant degradation in the force required to cause slippage was found for connecting bar slippage and fixator pin rotation; however, this clamp also had a significant increase in the force to initiate slippage for clockwise clamp-bolt axis pivot with reuse. The Securos clamp had significant degradation in connecting bar slippage, connecting bar rotation, and fixator pin rotation whereas the KE clamp had significant degradation in connecting bar slippage only. All 3 external fixator clamp types degraded in 1 or more movement variables in their ability to resist motion with reuse. The IMEX and Securos clamps were more subject to degradation than the KE clamp and this may have clinical importance for fixator composite rigidity. Fracture-reduction stability is related to the ability to resist motion within a clamp. The unpredictable nature of degradation we found cautions against repeated use. Clinicians should consider reuse of external fixator clamps with the knowledge that repeated use degrades clamp mechanical performance.
引用
收藏
页码:530 / 536
页数:7
相关论文
共 15 条
[1]   A COMPARISON OF NONTHREADED, ENHANCED THREADED, AND ELLIS FIXATION PINS USED IN TYPE-I EXTERNAL SKELETAL FIXATORS IN DOGS [J].
ANDERSON, MA ;
MANN, FA ;
WAGNERMANN, C ;
HAHN, AW ;
JIANG, BL ;
TOMLINSON, JL .
VETERINARY SURGERY, 1993, 22 (06) :482-489
[2]  
Anderson MA, 1996, J AM VET MED ASSOC, V208, P883
[3]  
ARO HT, 1989, CLIN ORTHOP RELAT R, P246
[4]  
*ASTM, 1998, ANN BOOK ASTM STAND, P1142
[5]   COMPARISON OF THE STRENGTH AND HOLDING POWER OF 4 PIN DESIGNS FOR USE WITH HALF PIN (TYPE-1) EXTERNAL SKELETAL FIXATION [J].
BENNETT, RA ;
EGGER, EL ;
HISTAND, M ;
ELLIS, AB .
VETERINARY SURGERY, 1987, 16 (03) :207-211
[6]   EXVIVO BIOMECHANICS OF KIRSCHNER-EHMER EXTERNAL SKELETAL FIXATION APPLIED TO CANINE TIBIAE [J].
BOUVY, BM ;
MARKEL, MD ;
CHELIKANI, S ;
EGGER, EL ;
PIERMATTEI, DL ;
VANDERBY, R .
VETERINARY SURGERY, 1993, 22 (03) :194-207
[7]  
BRINKER WO, 1985, J AM ANIM HOSP ASSOC, V21, P801
[8]   A COMPARISON OF ACUTE PULL-OUT STRENGTH BETWEEN 2-WAY AND ONE-WAY TRANSFIXATION PIN INSERTION FOR EXTERNAL SKELETAL FIXATION IN CANINE BONE [J].
DERNELL, WS ;
HARARI, J ;
BLACKKETTER, DM .
VETERINARY SURGERY, 1993, 22 (02) :110-114
[9]   EFFECT OF FIXATION PIN INSERTION ON THE BONE-PIN INTERFACE [J].
EGGER, EL ;
HISTAND, MB ;
BLASS, CE ;
POWERS, BE .
VETERINARY SURGERY, 1986, 15 (03) :246-252
[10]   STATIC STRENGTH EVALUATION OF 6 EXTERNAL SKELETAL FIXATION CONFIGURATIONS [J].
EGGER, EL .
VETERINARY SURGERY, 1983, 12 (03) :130-136