Study of a Centrifugal Blood Pump in a Mock Loop System

被引:6
作者
Uebelhart, Beatriz [1 ]
da Silva, Bruno Utiyama [1 ]
Fonseca, Jeison [1 ]
Bock, Eduardo [2 ]
Leme, Juliana [1 ]
da Silva, Cibele [1 ]
Leao, Tarcisio [1 ]
Andrade, Aron [1 ]
机构
[1] Dante Pazzanese Inst Cardiol, BR-04012180 Sao Paulo, Brazil
[2] Fed Inst Educ Sci & Technol Sao Paulo, Sao Paulo, Brazil
关键词
Ventricular assist device; Centrifugal blood pump; Cardiovascular simulator;
D O I
10.1111/aor.12228
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An implantable centrifugal blood pump (ICBP) is being developed to be used as a ventricular assist device (VAD) in patients with severe cardiovascular diseases. The ICBP system is composed of a centrifugal pump, a motor, a controller, and a power supply. The electricity source provides power to the controller and to a motor that moves the pump's rotor through magnetic coupling. The centrifugal pump is composed of four parts: external conical house, external base, impeller, and impeller base. The rotor is supported by a pivot bearing system, and its impeller base is responsible for sheltering four permanent magnets. A hybrid cardiovascular simulator (HCS) was used to evaluate the ICBP's performance. A heart failure (HF) (when the heart increases beat frequency to compensate for decrease in blood flow) was simulated in the HCS. The main objective of this work is to analyze changes in physiological parameters such as cardiac output, blood pressure, and heart rate in three situations: healthy heart, HF, and HF with left circulatory assistance by ICBP. The results showed that parameters such as aortic pressure and cardiac output affected by the HF situation returned to normal values when the ICBP was connected to the HCS. In conclusion, the test results showed satisfactory performance for the ICBP as a VAD.
引用
收藏
页码:946 / 949
页数:4
相关论文
共 10 条
[1]  
Bock E, 2011, 18 C ARG BIOING 2 TA
[2]   Implantable Centrifugal Blood Pump With Dual Impeller and Double Pivot Bearing System: Electromechanical Actuator, Prototyping, and Anatomical Studies [J].
Bock, Eduardo ;
Antunes, Pedro ;
Leao, Tarcisio ;
Uebelhart, Beatriz ;
Fonseca, Jeison ;
Leme, Juliana ;
Utiyama, Bruno ;
da Silva, Cibele ;
Cavalheiro, Andre ;
Santos Filho, Diolino ;
Dinkhuysen, Jarbas ;
Biscegli, Jose ;
Andrade, Aron ;
Arruda, Celso .
ARTIFICIAL ORGANS, 2011, 35 (05) :437-442
[3]   Cardiovascular Simulator Improvement: Pressure Versus Volume Loop Assessment [J].
Fonseca, Jeison ;
Andrade, Aron ;
Nicolosi, Denys E. C. ;
Biscegli, Jose F. ;
Leme, Juliana ;
Legendre, Daniel ;
Bock, Eduardo ;
Lucchi, Julio Cesar .
ARTIFICIAL ORGANS, 2011, 35 (05) :454-458
[4]  
GUYTON A.C., 1976, TXB MED PHYSL, V5th
[5]  
Libby P., 2008, Braunwald's Heart Disease, V8th
[6]  
Mcgee WT, 2009, GUIDE CARDIOPULMONAR
[7]  
Mesquita ET, 2004, BRAZ ARCH CARDIOL, V82, P494
[8]  
Uebelhart B, 2012, 7 TECHN MED SCI INT
[9]  
Uebelhart B, 2011, 21 BRAZ C MECH ENG C
[10]  
World Health Organization, 2013, CARD DIS