Fixed-flexion knee radiography using a new positioning device produced highly repeatable measurements of joint space width: ELSA-Brasil Musculoskeletal Study (ELSA-Brasil MSK)

被引:0
作者
Telles, Rosa Weiss [1 ]
Costa-Silva, Luciana [1 ]
Machado, Luciana A. C. [1 ]
Padilha dos Reis, Rodrigo Citton [1 ]
Barreto, Sandhi Maria [1 ]
机构
[1] Univ Fed Minas Gerais, Ctr Invest ELSA Brasil Minas Gerais, Fac Med, Belo Horizonte, MG, Brazil
关键词
Fixed-flexion radiography; Joint space width; Repeatability; Knee; Osteoarthritis; TEST-RETEST RELIABILITY; OSTEOARTHRITIS; SENSITIVITY; REPRODUCIBILITY; PERFORMANCE; PROGRESSION; AGREEMENT; ALIGNMENT; HEALTH; SCALE;
D O I
10.1016/j.rbr.2016.10.003
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To describe the performance of a non-fluoroscopic fixed-flexion PA radiographic protocol with a new positioning device, developed for the assessment of knee osteoarthritis (OA) in ELSA-Brasil MSK. Material and methods: A test-retest design including 19 adults (38 knee images) was conducted. Feasibility of the radiographic protocol was assessed by image quality parameters and presence of radioanatomic alignment according to intermargin distance (IMD) values. Repeatability was assessed for IMD and joint space width (JSW) measured at three different locations. Results: Approximately 90% of knee images presented excellent quality. Frequencies of nearly perfect radioanatomic alignment (IMD <1 mm) ranged from 29% to 50%, and satisfactory alignment was found in up to 71% and 76% of the images (IMD <= 1.5 mm and <= 1.7 mm, respectively). Repeatability analyses yielded the following results: IMD [SD of mean difference = 1.08; coefficient of variation (%CV) = 54.68%; intraclass correlation coefficient (ICC) (95%0) = 0.59 (0.34-0.77)1; JSW [SD of mean difference = 0.34-0.61; %CV = 4.48% - 9.80%; ICC (95%CI)= 0.74 (0.55-0.85) -0.94 (0.87-0.97)]. Adequately reproducible measurements of IMD and JSW were found in 68% and 87% of the images, respectively. Conclusions: Despite the difficulty in achieving consistent radioanatomic alignment between subsequent radiographs in terms of IMD, the protocol produced highly repeatable JSW measurements when these were taken at midpoint and 10 mm from the medial extremity of the medial tibial plateu. Therefore, measurements of JSW at these locations can be considered adequate for the assessment of knee OA in ELSA-Brasil MSK. (C) 2016 Published by Elsevier Editora Ltda.
引用
收藏
页码:154 / 161
页数:8
相关论文
共 35 条
  • [1] RADIOGRAPHIC ASSESSMENT OF PROGRESSION IN OSTEOARTHRITIS
    ALTMAN, RD
    FRIES, JF
    BLOCH, DA
    CARSTENS, J
    COOKE, D
    GENANT, H
    GOFTON, P
    GROTH, H
    MCSHANE, DJ
    MURPHY, WA
    SHARP, JT
    SPITZ, P
    WILLIAMS, CA
    WOLFE, F
    [J]. ARTHRITIS AND RHEUMATISM, 1987, 30 (11): : 1214 - 1225
  • [2] Aquino EM, 2012, AM J EPIDEMIOL, V175, P315
  • [3] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [4] Effect of medial tibial plateau alignment on serial radiographs on the capacity to predict progression of knee osteoarthritis
    Botha-Scheepers, S.
    Dougados, M.
    Ravaud, P.
    Le Graverand, M. -P. Hellio
    Watt, I.
    Breedveld, F. C.
    Kloppenburg, M.
    [J]. OSTEOARTHRITIS AND CARTILAGE, 2008, 16 (02) : 272 - 276
  • [5] Fixed-flexion knee radiography: the sensitivity to detect knee joint space narrowing in osteoarthritis
    Botha-Scheepers, S.
    Kloppenburg, M.
    Kroon, H. M.
    Le Graverand, M. -P. Hellio
    Breedveld, F. C.
    Ravaud, P.
    Dougados, M.
    [J]. OSTEOARTHRITIS AND CARTILAGE, 2007, 15 (03) : 350 - 353
  • [6] Brasil. Ministerio da Saude, 2009, REV SAUDE PUBL, V43
  • [7] Buckland-Wright JC, 2003, J RHEUMATOL, V30, P329
  • [8] Questionnaire development in ELSA-Brasil: challenges of a multidimensional instrument
    Chor, Dora
    de Mello Alves, Marcia Guimaraes
    Giatti, Luana
    Cade, Nagela Valadao
    Nunes, Maria Angelica
    Bisi Molina, Maria del Carmen
    Bensenor, Isabela M.
    Aquino, Estela M. L.
    Passos, Valeria
    Santos, Simone M.
    Mendes da Fonseca, Maria de Jesus
    de Oliveira, Leticia Cardoso
    [J]. REVISTA DE SAUDE PUBLICA, 2013, 47 : 27 - 36
  • [9] Comparison of fixed flexion, fluoroscopic semi-flexed and MTP radiographic methods for obtaining the minimum medial joint space width of the knee in longitudinal osteoarthritis trials
    Cline, G. A.
    Meyer, J. M.
    Stevens, R.
    Buckland-Wright, C.
    Peterfy, C.
    Beary, John F.
    [J]. OSTEOARTHRITIS AND CARTILAGE, 2006, 14 : A32 - A36
  • [10] Social capital in ELSA-Brasil: test-retest reliability of the Resource Generator scale
    Griep, Rosane Haerter
    Santos, Simone M.
    Cardoso, Leticia de Oliveira
    Mendes da Fonseca, Maria de Jesus
    de Mello Alves, Marcia Guimaraes
    Souto, Ester Paiva
    Chor, Dora
    [J]. REVISTA DE SAUDE PUBLICA, 2013, 47 : 131 - 139