ANALYTICAL AND CLINICAL EVALUATION OF SYSMEX UF1000I FOR AUTOMATED SCREENING OF CEREBROSPINAL FLUIDS

被引:10
作者
Buoro, Sabrina [1 ]
Ottomano, Cosimo [2 ]
Esposito, Sara Appassiti [1 ]
Gherardi, Patrizia [1 ]
Alessio, Maria Grazia [1 ]
Crippa, Alberto [1 ]
Farina, Claudio [3 ]
Raglio, Annibale [3 ]
Lippi, Giuseppe [2 ]
机构
[1] Azienda Osped Papa Giovanni XXIII, Lab Anal Chim Clin, Bergamo, Italy
[2] Azienda Osped Univ Parma, Unita Operat Diagnost Ematochim, I-43126 Parma, Italy
[3] Azienda Osped Papa Giovanni XXIII, Lab Microbiol & Virol, Bergamo, Italy
关键词
cerebrospinal fluid; meningitis; automation; cytofluorimeter; PERFORMANCE EVALUATION; HEMATOLOGY ANALYZERS; CELL ANALYSIS; BLOOD-CELLS; XE-5000; MODE; RELIABILITY;
D O I
10.2478/jomb-2013-0034
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Background: We evaluated the performance of Sysmex UF1000i for cell counting and differential cell count, as well as for assessment of bacteria load in cerebrospinal fluid (CSF), as a potential approach for the rapid screening of meningitis or bacterial encephalitis. Methods: We analyzed 77 consecutive CSF samples, 34 of which (44%) displayed leukocyte count > 5 white blood cell (WBC)/mu L with optical microscopy. Results on the UF-1000i were compared with those obtained by microscopic analysis. Imprecision was evaluated by testing three CSF samples with leukocyte values between 3.5 and 28.8 WBC/mu L in 10 replicates. Carry-over was evaluated with the Broughton formula on three CSF pools with leukocyte counts between 93.5 and 132.5 WBC/mu L. Linearity was assessed according to CLSI document EP6-A. In the presence of bacteria, identification and antibiogram were performed with Vitex (Biomerieux), except for Neisserie meningitidis (ApiNH, Biomerieux). Sensitivity tests were performed with Vitex and disc diffusion. Results: Optimal correlation was found between UF-1000i and optical microscopy, displaying Pearson's correlation of 0.99 and mean bias of -3.5 WBC/mu L (95% Cl, from -7.0 to 0.0 WBC/mu L). Imprecision varied between 12 and 16%. Linearity was excellent, 4-278 WBC/mu L. Carry-over was negligible. ROC analysis yielded AUC of 0.99 for both WBC and bacterial counts. The agreement at threshold > 4 WBC/mu L was 0.91, with sensitivity and specificity of 1.00 and 0.84. At <= bacteria/mu L cut-off, accuracy was 0.98, sensitivity 1.00 and specificity 0.97. Conclusions: According to these results, CSF screening with UF-1000i seems a reliable approach in terms of instrument performance, turnaround time and overall laboratory efficiency.
引用
收藏
页码:191 / 196
页数:6
相关论文
共 24 条
[1]  
[Anonymous], 2006, H56A CLSI
[2]  
[Anonymous], CLIN CHEM LAB MED
[3]   Automated flow cytometric analysis of blood cells in cerebrospinal fluid - Analytic performance [J].
Aune, MW ;
Becker, JL ;
Brugnara, C ;
Canfield, W ;
Dorfman, DM ;
Fiehn, W ;
Fischer, G ;
Fitzpatrick, P ;
Flaming, TH ;
Henriksen, HK ;
Kunicka, JE ;
Lackner, KJ ;
Minchello, E ;
Mullenix, PA ;
Myers, M ;
Petersen, A ;
Ternstrom, W ;
Wilson, S .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2004, 121 (05) :690-700
[4]   Evaluation of the XE-5000 for the automated analysis of blood cells in cerebrospinal fluid [J].
Boer, Klas ;
Deufel, Thomas ;
Reinhoefer, Mike .
CLINICAL BIOCHEMISTRY, 2009, 42 (7-8) :684-691
[5]   Screening for Urinary Tract Infection with the Sysmex UF-1000i Urine Flow Cytometer [J].
Broeren, Maarten A. C. ;
Bahceci, Semiha ;
Vader, Huib L. ;
Arents, Niek L. A. .
JOURNAL OF CLINICAL MICROBIOLOGY, 2011, 49 (03) :1025-1029
[6]   CARRY-OVER IN AUTOMATIC ANALYZERS [J].
BROUGHTON, PMG .
JOURNAL OF AUTOMATIC CHEMISTRY, 1984, 6 (02) :94-95
[7]   Analytical evaluation of Sysrnex UF-1000i for flow cytometric analysis of peritoneal fluid [J].
Buoro, Sabrina ;
Gustinetti, Rosanna ;
Dominoni, Paola ;
Alessio, Maria Grazia ;
Filisetti, Patrizia ;
Luca, Maria Grazia ;
Fagiuoli, Stefano ;
Ottomano, Cosimo ;
Lippi, Giuseppe .
CLINICAL BIOCHEMISTRY, 2012, 45 (15) :1263-1265
[8]   Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids [J].
de Jonge, Robert ;
Brouwer, Rob ;
de Graaf, Marieke T. ;
Luitwieler, Ronald L. ;
Fleming, Cherina ;
de Frankrijker-Merkestijn, Magda ;
Smitt, Peter A. E. Sillevis ;
Boonstra, Joke G. ;
Lindemans, Jan .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (05) :665-675
[9]   When and How to Investigate the Patient with Headache [J].
De Luca, G. C. ;
Bartleson, J. D. .
SEMINARS IN NEUROLOGY, 2010, 30 (02) :131-144
[10]   Emergency department management of meningitis and encephalitis [J].
Fitch, Michael T. ;
Abrahamian, Fredrick M. ;
Moran, Gregory J. ;
Talan, David A. .
INFECTIOUS DISEASE CLINICS OF NORTH AMERICA, 2008, 22 (01) :33-52