Effectiveness of Trevo stent retriever in acute ischemic stroke: Comparison with Solitaire stent

被引:18
作者
Yi, Ho Jun [1 ]
Lee, Dong Hoon [1 ]
Kim, Sang Uk [1 ]
机构
[1] Catholic Univ Korea, St Vincents Hosp, Dept Neurosurg, 93 Jungbu Daero Ji Dong, Suwon 16247, Gyeonggi Do, South Korea
基金
美国国家卫生研究院;
关键词
angiography; reperfusion; stent; stroke; thrombectomy; MECHANICAL THROMBECTOMY; ENDOVASCULAR TREATMENT; INTRAARTERIAL TREATMENT; RANDOMIZED-TRIAL; DEVICE; OUTCOMES;
D O I
10.1097/MD.0000000000010747
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Recently, mechanical thrombectomy with stent retriever has achieved faster and higher rates of recanalization for intracranial major vessel occlusion. However, comparative studies of the most widely used Solitaire and Trevo stents have been rarely published.The authors retrospectively reviewed a total of 200 patients who performed mechanical thrombectomy at our center during 4 years and divided patients into 2 groups: mechanical thrombectomy with Solitaire stent (Group 1: Solitaire) and mechanical thrombectomy with Trevo stent (Group 2: Trevo). All patients underwent a clinical assessment with National Institutes of Health Stroke Scale (NIHSS) score and underwent modified Rankin Scale (mRS) score. Radiologic results were evaluated using thrombolysis in cerebral infarction (TICI) score and number of stent passes. In addition, multiple time intervals were analyzed.There was no statistically significant difference in clinical outcome between the 2 groups. Trevo group revealed the shorter procedure time, less number of stent passage, and more one pass cases than Solitaire group with statistically significance (P=.009, P=.014, P=.030). In addition, Trevo group achieved higher successful recanalization (TICI 2b or 3) rate (89.7%) with statically significant than group1 (82.3%) (P=.018). In multivariate logistic regression analysis, the use of Trevo stent was a predictive for successful recanalization. (odds ratio 1.40, 95% confidence interval 1.250-1.550, P=.028).Our study suggests that the Trevo stent allows higher recanalization rate through the less number of stent passages and shorter procedure time than the Solitaire stent. More randomized control trials are needed to determine which stents are more effective.
引用
收藏
页数:7
相关论文
共 18 条
[1]   Guidelines for the early management of adults with ischemic stroke -: A guideline from the American Heart Association/American Stroke Association Stroke Council, Clinical Cardiology Council, Cardiovascular Radiology and Intervention Council, and the Atherosclerotic Peripheral Vascular Disease and Quality of Care Outcomes in Research Interdisciplinary Working Groups (Reprinted from Stroke, vol 38, pg 1655-1711, 2007) [J].
Adams, Harold P., Jr. ;
del Zoppo, Gregory ;
Alberts, Mark J. ;
Bhatt, Deepak L. ;
Brass, Lawrence ;
Furlan, Anthony ;
Grubb, Robert L. ;
Higashida, Randall T. ;
Jauch, Edward C. ;
Kidwell, Chelsea ;
Lyden, Patrick D. ;
Morgenstern, Lewis B. ;
Qureshi, Adnan I. ;
Rosenwasser, Robert H. ;
Scott, Phillip A. ;
Wijdicks, Eelco F. M. .
CIRCULATION, 2007, 115 (20) :E478-E534
[2]   Histological examination of vascular damage caused by stent retriever thrombectomy devices [J].
Arai, Daisuke ;
Ishii, Akira ;
Chihara, Hideo ;
Ikeda, Hiroyuki ;
Miyamoto, Susumu .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2016, 8 (10) :992-+
[3]   A Randomized Trial of Intraarterial Treatment for Acute Ischemic Stroke [J].
Berkhemer, O. A. ;
Fransen, P. S. S. ;
Beumer, D. ;
van den Berg, L. A. ;
Lingsma, H. F. ;
Yoo, A. J. ;
Schonewille, W. J. ;
Vos, J. A. ;
Nederkoorn, P. J. ;
Wermer, M. J. H. ;
van Walderveen, M. A. A. ;
Staals, J. ;
Hofmeijer, J. ;
van Oostayen, J. A. ;
Nijeholt, G. J. Lycklama A. ;
Boiten, J. ;
Brouwer, P. A. ;
Emmer, B. J. ;
de Bruijn, S. F. ;
van Dijk, L. C. ;
Kappelle, L. J. ;
Lo, R. H. ;
Van Dijk, E. J. ;
de Vries, J. ;
de Kort, P. L. M. ;
van Rooij, W. J. J. ;
van den Berg, J. S. P. ;
van Hasselt, B. A. A. M. ;
Aerden, L. A. M. ;
Dallinga, R. J. ;
Visser, M. C. ;
Bot, J. C. J. ;
Vroomen, P. C. ;
Eshghi, O. ;
Schreuder, T. H. C. M. L. ;
Heijboer, R. J. J. ;
Keizer, K. ;
Tielbeek, A. V. ;
den Hertog, H. M. ;
Gerrits, D. G. ;
van den Berg-Vos, R. M. ;
Karas, G. B. ;
Steyerberg, E. W. ;
Flach, H. Z. ;
Marquering, H. A. ;
Sprengers, M. E. S. ;
Jenniskens, S. F. M. ;
Beenen, L. F. M. ;
van den Berg, R. ;
Koudstaal, P. J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2015, 372 (01) :11-20
[4]   Endovascular Therapy for Ischemic Stroke with Perfusion-Imaging Selection [J].
Campbell, B. C. V. ;
Mitchell, P. J. ;
Kleinig, T. J. ;
Dewey, H. M. ;
Churilov, L. ;
Yassi, N. ;
Yan, B. ;
Dowling, R. J. ;
Parsons, M. W. ;
Oxley, T. J. ;
Wu, T. Y. ;
Brooks, M. ;
Simpson, M. A. ;
Miteff, F. ;
Levi, C. R. ;
Krause, M. ;
Harrington, T. J. ;
Faulder, K. C. ;
Steinfort, B. S. ;
Priglinger, M. ;
Ang, T. ;
Scroop, R. ;
Barber, P. A. ;
McGuinness, B. ;
Wijeratne, T. ;
Phan, T. G. ;
Chong, W. ;
Chandra, R. V. ;
Bladin, C. F. ;
Badve, M. ;
Rice, H. ;
de Villiers, L. ;
Ma, H. ;
Desmond, P. M. ;
Donnan, G. A. ;
Davis, S. M. .
NEW ENGLAND JOURNAL OF MEDICINE, 2015, 372 (11) :1009-1018
[5]   Influence of Device Choice on the Effect of Intra-Arterial Treatment for Acute Ischemic Stroke in MR CLEAN (Multicenter Randomized Clinical Trial of Endovascular Treatment for Acute Ischemic Stroke in the Netherlands) [J].
Dippel, Diederik W. ;
Majoie, Charles B. ;
Roos, Yvo B. ;
van der Lugt, Aad ;
van Oostenbrugge, Robert J. ;
van Zwam, Wim H. ;
Lingsma, Hester F. ;
Koudstaal, Peter J. ;
Treurniet, Kilian M. ;
van den Berg, Lucie A. ;
Beumer, Debbie ;
Fransen, Puck S. ;
Berkhemer, Olvert A. .
STROKE, 2016, 47 (10) :2574-2581
[6]   Randomized Assessment of Rapid Endovascular Treatment of Ischemic Stroke [J].
Goyal, M. ;
Demchuk, A. M. ;
Menon, B. K. ;
Eesa, M. ;
Rempel, J. L. ;
Thornton, J. ;
Roy, D. ;
Jovin, T. G. ;
Willinsky, R. A. ;
Sapkota, B. L. ;
Dowlatshahi, D. ;
Frei, D. F. ;
Kamal, N. R. ;
Montanera, W. J. ;
Poppe, A. Y. ;
Ryckborst, K. J. ;
Silver, F. L. ;
Shuaib, A. ;
Tampieri, D. ;
Williams, D. ;
Bang, O. Y. ;
Baxter, B. W. ;
Burns, P. A. ;
Choe, H. ;
Heo, J. -H. ;
Holmstedt, C. A. ;
Jankowitz, B. ;
Kelly, M. ;
Linares, G. ;
Mandzia, J. L. ;
Shankar, J. ;
Sohn, S. -I. ;
Swartz, R. H. ;
Barber, P. A. ;
Coutts, S. B. ;
Smith, E. E. ;
Morrish, W. F. ;
Weill, A. ;
Subramaniam, S. ;
Mitha, A. P. ;
Wong, J. H. ;
Lowerison, M. W. ;
Sajobi, T. T. ;
Hill, M. D. .
NEW ENGLAND JOURNAL OF MEDICINE, 2015, 372 (11) :1019-1030
[7]   An efficacy and safety comparison between different stentriever designs in acute ischaemic stroke: a systematic review and meta-analysis [J].
Grech, R. ;
Pullicino, R. ;
Thornton, J. ;
Downer, J. .
CLINICAL RADIOLOGY, 2016, 71 (01) :48-57
[8]   A novel microcatheter-delivered, highly-flexible and fully-retrievable stent, specifically designed for intracranial use - Technical note [J].
Henkes, H ;
Flesser, A ;
Brew, S ;
Miloslavski, E ;
Doerfler, A ;
Felber, S ;
Monstadt, H ;
Kuehne, D .
INTERVENTIONAL NEURORADIOLOGY, 2003, 9 (04) :391-393
[9]   Comparison of non-stent retriever and stent retriever mechanical thrombectomy devices for the endovascular treatment of acute ischemic stroke [J].
Hentschel, Kate A. ;
Daou, Badih ;
Chalouhi, Nohra ;
Starke, Robert M. ;
Clark, Shannon ;
Gandhe, Ashish ;
Jabbour, Pascal ;
Rosenwasser, Robert ;
Tjoumakaris, Stavropoula .
JOURNAL OF NEUROSURGERY, 2017, 126 (04) :1123-1130
[10]   Thrombectomy within 8 Hours after Symptom Onset in Ischemic Stroke [J].
Jovin, T. G. ;
Chamorro, A. ;
Cobo, E. ;
de Miquel, M. A. ;
Molina, C. A. ;
Rovira, A. ;
San Roman, L. ;
Serena, J. ;
Abilleira, S. ;
Ribo, M. ;
Millan, M. ;
Urra, X. ;
Cardona, P. ;
Lopez-Cancio, E. ;
Tomasello, A. ;
Castano, C. ;
Blasco, J. ;
Aja, L. ;
Dorado, L. ;
Quesada, H. ;
Rubiera, M. ;
Hernandez-Perez, M. ;
Goyal, M. ;
Demchuk, A. M. ;
von Kummer, R. ;
Gallofre, M. ;
Davalos, A. .
NEW ENGLAND JOURNAL OF MEDICINE, 2015, 372 (24) :2296-2306