Comparison between human and model observer performance in low- contrast detection tasks in CT images: application to images reconstructed with filtered back projection and iterative algorithms

被引:18
作者
Hernandez-Giron, I. [1 ,2 ]
Calzado, A. [2 ]
Geleijns, J. [3 ]
Joemai, R. M. S. [3 ]
Veldkamp, W. J. H. [3 ]
机构
[1] Univ Rovira & Virgili, E-43007 Tarragona, Spain
[2] Univ Complutense Madrid, Dept Radiol, Madrid, Spain
[3] Leiden Univ, Med Ctr, Dept Radiol, Leiden, Netherlands
关键词
VISUAL-PERCEPTION; SYSTEMS; NOISE;
D O I
10.1259/bjr.20140014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To compare low-contrast detectability (LCDet) performance between a model [non-pre-whitening matched filter with an eye filter (NPWE)] and human observers in CT images reconstructed with filtered back projection (FBP) and iterative [adaptive iterative dose reduction three-dimensional (AIDR 3D; Toshiba Medical Systems, Zoetermeer, Netherlands)] algorithms. Methods: Images of the Catphan (R) phantom (Phantom Laboratories, New York, NY) were acquired with Aquilion ONE T 320-detector row CT (Toshiba Medical Systems, Tokyo, Japan) at five tube current levels (20-500mA range) and reconstructed with FBP and AIDR 3D. Samples containing either low-contrast objects (diameters, 2-15mm) or background were extracted and analysed by the NPWE model and four human observers in a two-alternative forced choice detection task study. Proportion correct (PC) values were obtained for each analysed object and used to compare human and model observer performances. An efficiency factor (h) was calculated to normalize NPWE to human results. Results: Human and NPWE model PC values (normalized by the efficiency, h50.44) were highly correlated for the whole dose range. The Pearson's product-moment correlation coefficients (95% confidence interval) between human and NPWE were 0.984 (0.972-0.991) for AIDR 3D and 0.984 (0.971-0.991) for FBP, respectively. Bland-Altman plots based on PC results showed excellent agreement between human and NPWE [mean absolute difference 0.560.4%; range of differences (24.7%, 5.6%)]. Conclusion: The NPWE model observer can predict human performance in LCDet tasks in phantom CT images reconstructed with FBP and AIDR 3D algorithms at different dose levels.
引用
收藏
页数:10
相关论文
共 30 条
[1]   Iterative reconstruction methods in X-ray CT [J].
Beister, Marcel ;
Kolditz, Daniel ;
Kalender, Willi A. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02) :94-108
[2]   Iterative reconstruction in image space (IRIS) in cardiac computed tomography: initial experience [J].
Bittencourt, Marcio Sommer ;
Schmidt, Bernhard ;
Seltmann, Martin ;
Muschiol, Gerd ;
Ropers, Dieter ;
Daniel, Werner Guenther ;
Achenbach, Stephan .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2011, 27 (07) :1081-1087
[3]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[4]   Prewhitening revisited [J].
Burgess, AE .
IMAGE PERCEPTION: MEDICAL IMAGING 1998, 1998, 3340 :55-64
[5]   Human observer detection experiments with mammograms and power-law noise [J].
Burgess, AE ;
Jacobson, FL ;
Judy, PF .
MEDICAL PHYSICS, 2001, 28 (04) :419-437
[6]   Visual Perception Studies and Observer Models in Medical Imaging [J].
Burgess, Arthur E. .
SEMINARS IN NUCLEAR MEDICINE, 2011, 41 (06) :419-436
[7]   HUMAN VISUAL-PERCEPTION AND ROC METHODOLOGY IN MEDICAL IMAGING [J].
CHESTERS, MS .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (07) :1433-1476
[8]   Projected Cancer Risks From Computed Tomographic Scans Performed in the United States in 2007 [J].
de Gonzalez, Amy Berrington ;
Mahesh, Mahadevappa ;
Kim, Kwang-Pyo ;
Bhargavan, Mythreyi ;
Lewis, Rebecca ;
Mettler, Fred ;
Land, Charles .
ARCHIVES OF INTERNAL MEDICINE, 2009, 169 (22) :2071-2077
[9]  
Eckstein M.P., 2000, HDB MED IMAGING, V1, P595
[10]  
Fan J, 2012, P SOC PHOTO-OPT INS, V8318