Dose reduction in mammography with photon counting imaging

被引:13
作者
Thunberg, S [1 ]
Adelöw, L [1 ]
Blom, O [1 ]
Cöster, A [1 ]
Egerström, J [1 ]
Eklund, M [1 ]
Egnell, P [1 ]
Francke, T [1 ]
Jordung, U [1 ]
Kristoffersson, T [1 ]
Lindman, K [1 ]
Lindqvist, L [1 ]
Marchal, D [1 ]
Olla, H [1 ]
Penton, E [1 ]
Peskov, V [1 ]
Rantanen, J [1 ]
Solokov, S [1 ]
Svedenhag, P [1 ]
Ullberg, C [1 ]
Weber, N [1 ]
机构
[1] XCounter AB, SE-18233 Danderyd, Sweden
来源
MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 2004年 / 5368卷
关键词
photon counting; digital mammography; dose reduction; CDMAM phantom;
D O I
10.1117/12.530649
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to investigate if the glandular dose to the breast in mammography can significantly be reduced without compromising image quality, when using photon counting technology, in a multi-slit scanning photon counting detector, compared to a conventional film mammography system and commercial available digital mammography systems with TFT-array detectors. A CDMAM phantom study, with two different thicknesses of additional PMMA absorber, 4 cm and 7 cm respectively, has shown that multi-slit scanning photon counting detector technology can reduce the dose, without reducing the image quality. This comparison was made to two commercial available digital mammography systems Senographe 2000D (from GEMS) and Selenia (from Lorad). The results show that dose can be reduced with 63% to 77%, depending on object thickness, when using XCT for mammography. This dose reduction has also been verified clinically through a small pilot study with patients and specimen, where the comparison was made between XCT and film.
引用
收藏
页码:457 / 465
页数:9
相关论文
共 6 条
[1]  
[Anonymous], MED PHYS
[2]  
*ART MED SYST BV, 2002, US MAN CDMAM 3 4
[3]   Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol [J].
Dance, DR ;
Skinner, CL ;
Young, KC ;
Beckett, JR ;
Kotre, CJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3225-3240
[4]   MONTE-CARLO CALCULATION OF CONVERSION FACTORS FOR THE ESTIMATION OF MEAN GLANDULAR BREAST DOSE [J].
DANCE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1211-1219
[5]  
THILANDERKLANG A, 1997, INVESTIGATION XRAY S
[6]   Evaluation of a photon counting mammography system [J].
Thunberg, S ;
Francke, T ;
Egerström, J ;
Eklund, M ;
Ericsson, L ;
Kristoffersson, TC ;
Peskov, V ;
Rantanen, J ;
Sokolov, S ;
Svedenhag, P ;
Ullberg, C ;
Weber, N .
MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 :202-208