Fluidics and heat generation of Alcon Infiniti and Legacy, Bausch&Lomb Millennium, and Advanced Medical Optics Sovereign phacoemulsification systems

被引:13
作者
Floyd, Michael S. [1 ]
Valentine, Jeremy R. [1 ]
Olson, Randall J. [1 ]
机构
[1] John A Moran Eye Ctr, Dept Ophthalmol & Visual Sci, Salt Lake City, UT 84132 USA
关键词
D O I
10.1016/j.ajo.2006.04.042
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
center dot PURPOSE: To study heat generation, vacuum, and flow characteristics of the Alcon Infiniti and Bausch & Lomb Millennium with results compared with the Alcon Legacy and advanced medical optics (AMO) Sovereign machines previously studied. center dot DESIGN: Experimental study. center dot METHODS: Heat generation with continuous ultrasound was determined with and without a 200,g weight. Flow and vacuum were determined from 12 to 40-ml/min in 2-ml/min steps. The impact of a STAAR Cruise Control was also tested. center dot RESULTS: Millennium created the most heat/20% of power (5.67 +/- 0.5 degrees C unweighted and 6.80 +/- 0.80 degrees C weighted), followed by Sovereign (4.59 +/- 0.70 degrees C unweighted and 5.65 +/- 0.72 degrees C weighted), Infiniti (2.79 +/- 0.62 degrees C unweighted and 3.96 +/- 0.31 degrees C weighted), and Legacy (1.99 +/- 0.49 degrees C unweighted and 4.27 +/- 0.76 degrees C weighted; P < .0001 for all comparisons between machines except Infiniti vs Legacy, both weighted). Flow studies revealed that Millennium Peristaltic was 17% less than indicated (P < .0001 to all other machines), and all other machines were within 3.5% of indicated. Cruise Control decreased flow by 4.1% (P < .0001 for same machine without it). Millennium Venturi had the great. est vacuum (81% more than the least Sovereign; P < .0001), and Cruise Control increased vacuum in a peristaltic machine 35% more than the Venturi system (P < .0001). center dot CONCLUSIONS: Percent power is not consistent in regard to heat generation, however, flow was accurate for all machines except Millennium Peristaltic. Restriction with Cruise Control elevates unoccluded vacuum to levels greater than the Venturi system tested.
引用
收藏
页码:387 / 392
页数:6
相关论文
共 7 条
[1]  
ADAMS W, IN PRESS AM J OPHTHA
[2]   A survey about phacoemulsification incision thermal contraction incidence and causal relationships [J].
Bradley, MJ ;
Olson, RJ .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 141 (01) :222-224
[3]   Comparison of thermal features associated with 2 phacoemulsification machines [J].
Brinton, JP ;
Adams, W ;
Kumar, R ;
Olson, RJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (02) :288-293
[4]   Thermal comparison of the AdvanTec Legacy, Sovereign WhiteStar, and Millennium phacoemulsification systems [J].
Mackool, RJ ;
Sirota, MA .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (04) :812-817
[5]   In-air thermal imaging comparison of Legacy AdvanTec, Millennium, and Sovereign WhiteStar phacoemulsification systems [J].
Olson, MD ;
Miller, KM .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (08) :1640-1647
[6]  
PAYNE M, IN PRESS J CATARACT
[7]   Volume-based characterization of postocclusion surge [J].
Zacharias, J ;
Zacharias, S .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (10) :1976-1982