Image quality and absorbed dose comparison of single- and dual-source cone-beam computed tomography

被引:2
作者
Miura, Hideharu [1 ,2 ]
Ozawa, Shuichi [1 ,2 ]
Okazue, Toshiya [1 ]
Kawakubo, Atsushi [1 ]
Yamada, Kiyoshi [1 ]
Nagata, Yasushi [1 ,2 ]
机构
[1] Hiroshima High Precis Radiotherapy Canc Ctr, Hiroshima, Japan
[2] Hiroshima Univ, Inst Biomed & Hlth Sci, Dept Radiat Oncol, Hiroshima, Japan
关键词
absorbed dose; dual-source cone-beam computed tomography; image quality; image-guided radiation therapy; RADIATION-THERAPY; GUIDED RADIOTHERAPY; CT; ASSURANCE; SYSTEM;
D O I
10.1002/acm2.12328
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Dual-source cone-beam computed tomography (DCBCT) is currently available in the Vero4DRT image-guided radiotherapy system. We evaluated the image quality and absorbed dose for DCBCT and compared the values with those for single-source CBCT (SCBCT). Methods: Image uniformity, Hounsfield unit (HU) linearity, image contrast, and spatial resolution were evaluated using a Catphan phantom. The rotation angle for acquiring SCBCT and DCBCT images is 215 degrees and 115 degrees, respectively. The image uniformity was calculated using measurements obtained at the center and four peripheral positions. The HUs of seven materials inserted into the phantom were measured to evaluate HU linearity and image contrast. The Catphan phantom was scanned with a conventional CT scanner to measure the reference HU for each material. The spatial resolution was calculated using high-resolution pattern modules. Image quality was analyzed using ImageJ software ver. 1.49. The absorbed dose was measured using a 0.6-cm(3) ionization chamber with a 16-cm-diameter cylindrical phantom, at the center and four peripheral positions of the phantom, and calculated using weighted cone-beam CT dose index (CBCTDIw). Results: Compared with that of SCBCT, the image uniformity of DCBCT was slightly reduced. A strong linear correlation existed between the measured HU for DCBCT and the reference HU, although the linear regression slope was different from that of the reference HU. DCBCT had poorer image contrast than did SCBCT, particularly with a high-contrast material. There was no significant difference between the spatial resolutions of SCBCT and DCBCT. The absorbed dose for DCBCT was higher than that for SCBCT, because in DCBCT, the two x-ray projections overlap between 45 degrees and 70 degrees. Conclusions: We found that the image quality was poorer and the absorbed dose was higher for DCBCT than for SCBCT in the Vero4DRT.
引用
收藏
页码:360 / 366
页数:7
相关论文
共 21 条
[1]  
[Anonymous], 2010, Comprehensive methodology for the evaluation of radiation dose in xray computed tomography (111)
[2]   A quality assurance program for image quality of cone-beam CT guidance in radiation therapy [J].
Bissonnette, Jean-Pierre ;
Moseley, Douglas J. ;
Jaffray, David A. .
MEDICAL PHYSICS, 2008, 35 (05) :1807-1815
[3]  
Chan M, 2011, BIOMED IMAGING INTER, V7, P7
[4]   A PRACTICAL METHOD TO MEASURE THE MTF OF CT SCANNERS [J].
DROEGE, RT ;
MORIN, RL .
MEDICAL PHYSICS, 1982, 9 (05) :758-759
[5]   X-ray scattering in single- and dual-source CT [J].
Engel, Klaus J. ;
Herrmann, Christoph ;
Zeitler, Guenter .
MEDICAL PHYSICS, 2008, 35 (01) :318-332
[6]   Interleaved acquisition for cross scatter avoidance in dual cone-beam CT [J].
Giles, William ;
Bowsher, James ;
Li, Hao ;
Yin, Fang-Fang .
MEDICAL PHYSICS, 2012, 39 (12) :7719-7728
[7]   An image quality comparison study between XVI and OBI CBCT systems [J].
Kamath, Srijit ;
Song, William ;
Chvetsov, Alexei ;
Ozawa, Shuichi ;
Lu, Haibin ;
Samant, Sanjiv ;
Liu, Chihray ;
Li, Jonathan G. ;
Palta, Jatinder R. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02) :376-390
[8]   Development of a four-dimensional image-guided radiotherapy system with a gimbaled X-ray head [J].
Kamino, Yuichiro ;
Takayama, Kenji ;
Kokubo, Masaki ;
Narita, Yuichiro ;
Hirai, Etsuro ;
Kawawda, Noriyuki ;
Mizowaki, Takashi ;
Nagata, Yasushi ;
Nishidai, Takehiro ;
Hiraoka, Masahiro .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (01) :271-278
[9]   Task Group 142 report: Quality assurance of medical accelerators [J].
Klein, Eric E. ;
Hanley, Joseph ;
Bayouth, John ;
Yin, Fang-Fang ;
Simon, William ;
Dresser, Sean ;
Serago, Christopher ;
Aguirre, Francisco ;
Ma, Lijun ;
Arjomandy, Bijan ;
Liu, Chihray ;
Sandin, Carlos ;
Holmes, Todd .
MEDICAL PHYSICS, 2009, 36 (09) :4197-4212
[10]   Cone-beam-CT guided radiation therapy:: technical implementation [J].
Létourneau, D ;
Wong, JW ;
Oldham, M ;
Gulam, M ;
Watt, L ;
Jaffray, DA ;
Siewerdsen, JH ;
Martinez, AA .
RADIOTHERAPY AND ONCOLOGY, 2005, 75 (03) :279-286