Temperature in the anterior chamber during phacoemulsification

被引:13
作者
Suzuki, Hisaharu
Oki, Kotaro
Igarashi, Tsutornu [1 ]
Shiwa, Toshihiko [1 ]
Takahashi, Hiroshi [1 ]
机构
[1] Musashikosugi Hosp, Nippon Med Sch, Dept Ophthalmol, Kawasaki, Kanagawa 2118533, Japan
关键词
SOVEREIGN-WHITESTAR; ADVANTEC; SYSTEMS; LEGACY; MILLENNIUM; TIPS;
D O I
10.1016/j.jcrs.2013.08.063
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate changes in the aqueous humor temperature using 2 phacoemulsification units (Stellaris 28.5 kHz device and Whitestar Signature 40 kHz device). SETTING: Nippon Medical School, Musashikosugi Hospital, Kawasaki City, Kanagawa, Japan. DESIGN: Experimental study. METHODS: Aqueous humor temperatures were measured with a temperature probe set in the anterior chamber during ultrasound (US) oscillation in porcine eyes under 5 conditions. RESULTS: Continuous longitudinal oscillation caused a rapid rise in aqueous humor temperature, while the pulse and elliptical modes suppressed temperature elevation. Reducing the number of US tip vibrations did not reduce the temperature in the anterior chamber. However, raising the vacuum setting allowed the aspirations to rise to the set value, thereby lowering the temperature in the anterior chamber. CONCLUSION: Because differences in the phacoemulsification settings can lead to temperature elevations in the anterior chamber, surgeons must carefully monitor these settings to prevent corneal tissue damage. (C) 2014 ASCRS AND ESCRS.
引用
收藏
页码:805 / 810
页数:6
相关论文
共 13 条
[1]   Thermal effect on corneal incisions with different phacoemulsification ultrasonic tips [J].
Bissen-Miyajima, H ;
Shimmura, S ;
Tsubota, K .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1999, 25 (01) :60-64
[2]   Comparison of thermal features associated with 2 phacoemulsification machines [J].
Brinton, JP ;
Adams, W ;
Kumar, R ;
Olson, RJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (02) :288-293
[3]   Phacoemulsification conditions resulting in thermal wound injury [J].
Ernest, P ;
Rhem, M ;
McDermott, M ;
Lavery, K ;
Sensoli, A .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (11) :1829-1839
[4]   Heat production: Longitudinal versus torsional phacoemulsification [J].
Han, Young Keun ;
Miller, Kevin M. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (10) :1799-1805
[5]  
Innocenti B, 2008, J APPL BIOMATER BIOM, V6, P151
[7]   Thermal comparison of the AdvanTec Legacy, Sovereign WhiteStar, and Millennium phacoemulsification systems [J].
Mackool, RJ ;
Sirota, MA .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (04) :812-817
[8]   In-air thermal imaging comparison of Legacy AdvanTec, Millennium, and Sovereign WhiteStar phacoemulsification systems [J].
Olson, MD ;
Miller, KM .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (08) :1640-1647
[9]   Thermal study of bare tips with various system parameters and incision sizes [J].
Osher, RH ;
Injev, VP .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (05) :867-872
[10]   Thermal inertia associated with ultrapulse technology in phacoemulsification [J].
Payne, Marielle ;
Waite, Aaron ;
Olson, Randall J. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (06) :1032-1034