Low Power Integrated Circuits for Wireless Neural Recording Applications

被引:1
作者
Zhang, Xu [1 ]
Pei, Weihua [1 ]
Gui, Qiang [1 ]
Chen, Hongda [1 ]
机构
[1] Chinese Acad Sci, Inst Semicond, State Key Lab Integrated Optoelect, Beijing 100083, Peoples R China
来源
2008 IEEE ASIA PACIFIC CONFERENCE ON CIRCUITS AND SYSTEMS (APCCAS 2008), VOLS 1-4 | 2008年
关键词
D O I
10.1109/APCCAS.2008.4746107
中图分类号
TP3 [计算技术、计算机技术];
学科分类号
0812 ;
摘要
A group of prototype integrated circuits are presented for a wireless neural recording micro-system. An inductive link was built for transcutaneous wireless power transfer and data transmission. Power and data were transmitted by a pair of coils on a same carrier frequency. The integrated receiver circuitry was composed of a full-wave bridge rectifier, a voltage regulator, a date recovery circuit, a clock recovery circuit and a power detector. The amplifiers were designed with a limited bandwidth for neural signals acquisition. An integrated FM transmitter was used to transmit the extracted neural signals to external equipments. 16.5 mW power and 50 bps - 2.5 Kbps command data can be received over 1 MHz carrier within 10 mm. The total gain of 60 dB was obtained by the preamplifier and a main amplifier at 0.95Hz - 13.41 KHz with 0.215 mW power dissipation. The power consumption of the 100 MHz ASK transmitter is 0.374 mW. All the integrated circuits operated under a 3.3 V power supply except the voltage regulator.
引用
收藏
页码:650 / 653
页数:4
相关论文
共 16 条
  • [1] A wireless implantable multichannel digital neural recording system for a micromachined sieve electrode
    Akin, T
    Najafi, K
    Bradley, RM
    [J]. IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1998, 33 (01) : 109 - 118
  • [2] Single-unit neural recording with active microelectrode arrays
    Bai, Q
    Wise, KD
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2001, 48 (08) : 911 - 920
  • [3] Power, clock, and data recovery in a wireless neural recording device
    Black, Daniel J.
    Harrison, Reid R.
    [J]. 2006 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS, VOLS 1-11, PROCEEDINGS, 2006, : 5083 - +
  • [4] Chen D, 2004, P ANN INT IEEE EMBS, V26, P4071
  • [5] GHOVANLOO M, 2002, IEEE EMBS SPEC TOP C
  • [6] A 100-channel system for real time detection and storage of extracellular spike waveforms
    Guillory, KS
    Normann, RA
    [J]. JOURNAL OF NEUROSCIENCE METHODS, 1999, 91 (1-2) : 21 - 29
  • [7] A low-power integrated circuit for a wireless 100-electrode neural recording system
    Harrison, Reid R.
    Watkins, Paul T.
    Kier, Ryan J.
    Lovejoy, Robert O.
    Black, Daniel J.
    Greger, Bradley
    Solzbacher, Florian
    [J]. IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2007, 42 (01) : 123 - 133
  • [8] Harrison RR, 2002, 2002 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS, VOL V, PROCEEDINGS, P197
  • [9] Neuronal ensemble control of prosthetic devices by a human with tetraplegia
    Hochberg, Leigh R.
    Serruya, Mijail D.
    Friehs, Gerhard M.
    Mukand, Jon A.
    Saleh, Maryam
    Caplan, Abraham H.
    Branner, Almut
    Chen, David
    Penn, Richard D.
    Donoghue, John P.
    [J]. NATURE, 2006, 442 (7099) : 164 - 171
  • [10] Brain-machine interfaces: past, present and future
    Lebedev, Mikhail A.
    Nicolelis, Miguel A. L.
    [J]. TRENDS IN NEUROSCIENCES, 2006, 29 (09) : 536 - 546