Test-retest variability in glaucoma patients tested with C-20-1 screening-mode frequency doubling technology perimetry

被引:9
作者
Brush, MB [1 ]
Chen, PP [1 ]
机构
[1] Univ Washington, Dept Ophthalmol, Seattle, WA 98195 USA
关键词
glaucoma; screening; FDT; perimetry; fluctuation; learning effect;
D O I
10.1097/00061198-200408000-00003
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate test-retest variability in glaucoma patients tested with C-20-1 (screening) Frequency Doubling Technology (FDT) perimetry. Methods: Thirty-one glaucoma patients with a range of visual field defects (17 with mild, 11 with moderate, and 3 with severe defects by AGIS classification) and experienced in standard automated perimetry but not FDT perimetry were prospectively recruited. The C-20-1 screening mode of FDT was consecutively administered three times. Concordance between tests for each patient was measured by presence and severity of defects for each test location. Learning and/or fatigue effects were also evaluated using FDT test scores in which number and severity of defects was graded. Results: Agreement between FDT test locations averaged 88% for any defect, but only 51% for severity of defect. Although the mean test score increased slightly over the three tests, subjects did not demonstrate a statistically significant change in test score across the three testings (P = 0.975). Among 17 patients with early glaucoma (AGIS score 1-5), two returned a normal initial FDT test, and four returned a normal second FDT test. Conclusions: Using C-20-1 FDT, concordance between tests at a given test location was high for the presence of any defect, which would be important in glaucoma screening, but was only fair for severity of defect. Six of 17 patients with early glaucoma returned normal tests during testing. This group of glaucoma patients with perimetric experience did not demonstrate a significant learning or fatigue effect with FDT screening perimetry.
引用
收藏
页码:273 / 277
页数:5
相关论文
共 38 条
[1]   Normal aging effects for frequency doubling technology perimetry [J].
Adams, CW ;
Bullimore, MA ;
Wall, M ;
Fingeret, M ;
Johnson, CA .
OPTOMETRY AND VISION SCIENCE, 1999, 76 (08) :582-587
[2]   Effect of dichoptic adaptation on frequency-doubling perimetry [J].
Anderson, AJ ;
Johnson, CA .
OPTOMETRY AND VISION SCIENCE, 2002, 79 (02) :88-92
[3]  
Anderson DR, 1992, AUTOMATED STATIC PER, P91
[4]  
BRUSH MB, IN PRESS AM J OPHTHA
[5]   Clinical comparison of frequency doubling technology perimetry and Humphrey perimetry [J].
Casson, R ;
James, B ;
Rubinstein, A ;
Ali, H .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2001, 85 (03) :360-362
[6]   Frequency doubling technology perimetry for detection of glaucomatous visual field loss [J].
Cello, KE ;
Nelson-Quigg, JM ;
Johnson, CA .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2000, 129 (03) :314-322
[7]  
Chauhan BC, 1999, INVEST OPHTH VIS SCI, V40, P648
[8]   Correlation of visual field progression between eyes in patients with open-angle glaucoma [J].
Chen, PP .
OPHTHALMOLOGY, 2002, 109 (11) :2093-2099
[9]   Visual field progression in patients with initially unilateral visual field loss from chronic open-angle glaucoma [J].
Chen, PP ;
Park, RJ .
OPHTHALMOLOGY, 2000, 107 (09) :1688-1692
[10]  
CHEN PP, 2003, OPHTHALMOLOGY, V110, P176