COMPARISON OF THE KING VISION VIDEO LARYNGOSCOPE WITH THE MACINTOSH LARYNGOSCOPE

被引:32
作者
Murphy, Laurel D. [1 ]
Kovacs, George J. [1 ]
Reardon, Peter M. [1 ]
Law, John Adam [2 ]
机构
[1] Dalhousie Univ, Dept Emergency Med, Halifax, NS B3H 3A7, Canada
[2] Dalhousie Univ, Dept Anesthesia, Halifax, NS B3H 3A7, Canada
关键词
airway management; video laryngoscope; manikin study; cadaver study; ENDOTRACHEAL INTUBATION; TRACHEAL INTUBATION; AIRWAY MANAGEMENT; SCOPE; COMPLICATIONS; METAANALYSIS; OROTRACHEAL; GLIDESCOPE; AIRTRAQ(R); EXPERIENCE;
D O I
10.1016/j.jemermed.2014.02.008
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Endotracheal intubation is a common procedure in the emergency department, and new devices may improve intubation time, success, or view. Objective: We compared the King Vision video laryngoscope (KVVL; King Systems, Noblesville, IN) to the Macintosh direct laryngoscope (DL) in simulated normal and difficult airways. Methods: Using manikins and clinical-grade cadavers, difficult airway scenarios were simulated using head movement restriction or a cervical spine collar. Four scenarios were studied using the KVVL and DL: normal manikin airway, difficult manikin airway, normal cadaver airway, and difficult cadaver airway. Primary outcomes were time to intubation and rate of successful intubation. Secondary outcomes were the percent of glottic opening and Cormack-Lehane grade visualized. Results: Thirty-two paramedics participated in the study. In the normal manikin airway scenario, time to intubation was 3.4 s (99% confidence interval [CI] 0.1-6.6) faster with the KVVL compared with DL. Time to intubation was 11.3 s (99% CI 2.4-20.2) faster with the KVVL in the difficult cadaver airway scenario. There was no difference in time to intubation in the other 2 scenarios. In the difficult cadaver airway, 10 of 32 participants failed to successfully intubate the trachea using DL, whereas all KVVL intubations were successful. All scenarios found a lower Cormack-Lehane grade and higher percentage of glottic opening with the KVVL compared to DL. Conclusion: The KVVL was slightly faster than Macintosh DL in two of four studied airway scenarios, and had a higher success rate in the difficult cadaver airway scenario. Further study is required in the clinical setting. (C) 2014 Elsevier Inc.
引用
收藏
页码:239 / 245
页数:7
相关论文
共 34 条
[11]   A META-ANALYSIS OF PREHOSPITAL AIRWAY CONTROL TECHNIQUES PART I: OROTRACHEAL AND NASOTRACHEAL INTUBATION SUCCESS RATES [J].
Hubble, Michael W. ;
Brown, Lawrence ;
Wilfong, Denise A. ;
Hertelendy, Attila ;
Benner, Randall W. ;
Richards, Michael E. .
PREHOSPITAL EMERGENCY CARE, 2010, 14 (03) :377-401
[12]   Airway research: the current status and future directions [J].
Isono, S. ;
Greif, R. ;
Mort, T. C. .
ANAESTHESIA, 2011, 66 :3-10
[13]   Morbidity related to emergency endotracheal intubation-A substudy of the KETAmine SEDation trial [J].
Jabre, Patricia ;
Avenel, Alexandre ;
Combes, Xavier ;
Kulstad, Erik ;
Mazariegos, Ixchel ;
Bertrand, Lionel ;
Lapostolle, Frederic ;
Adnet, Frederic .
RESUSCITATION, 2011, 82 (05) :517-522
[14]  
Konrad C, 1998, ANESTH ANALG, V86, P635
[15]   Airway management: "the times they are a-changin'' [J].
Kovacs, George .
CANADIAN JOURNAL OF EMERGENCY MEDICINE, 2013, 15 (06) :317-320
[16]   Limitations of difficult airway prediction in patients intubated in the emergency department [J].
Levitan, RM ;
Everett, WW ;
Ochroch, EA .
ANNALS OF EMERGENCY MEDICINE, 2004, 44 (04) :307-313
[17]   Assessment of airway visualization: Validation of the percentage of glottic opening (POGO) scale [J].
Levitan, RM ;
Ochroch, EA ;
Kush, S ;
Shofer, FS ;
Hollander, JE .
ACADEMIC EMERGENCY MEDICINE, 1998, 5 (09) :919-923
[18]   A paramedic study comparing the use of the Airtraq®, Airway Scope and Macintosh laryngoscopes in simulated prehospital airway scenarios [J].
Lewis, A. R. ;
Hodzovic, I. ;
Whelan, J. ;
Wilkes, A. R. ;
Bowler, I. ;
Whitfield, R. .
ANAESTHESIA, 2010, 65 (12) :1187-1193
[19]   Tracheal Intubation of a Difficult Airway Using Airway Scope, Airtraq, and Macintosh Laryngoscope: A Comparative Manikin Study of Inexperienced Personnel [J].
Liu, Liangji ;
Tanigawa, Koichi ;
Kusunoki, Shinji ;
Tamura, Tomoko ;
Ota, Kohei ;
Yamaga, Satoshi ;
Kida, Yoshiko ;
Otani, Tadatsugu ;
Sadamori, Takuma ;
Takeda, Taku ;
Iwasaki, Yasumasa ;
Hirohashi, Nobuyuki .
ANESTHESIA AND ANALGESIA, 2010, 110 (04) :1049-1055
[20]  
Maharaj ChrisenH., 2007, Anesthesiology, V107, P53