The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis
被引:21
作者:
Pecoraro, Valentina
[1
]
Petri, Davide
[2
]
Costantino, Giorgio
[3
]
Squizzato, Alessandro
[4
]
Moja, Lorenzo
[5
]
Virgili, Gianni
[6
]
Lucenteforte, Ersilia
[2
]
机构:
[1] AUSL Modena, Dept Lab Med & Pathol, Osped Civile St Agostino Estense, Modena, Italy
[2] Univ Pisa, Dept Clin & Expt Med, Via Roma 10, I-56126 Pisa, Italy
[3] Univ Milan, Osped Maggiore Policlin, IRCCS Fdn Ca Granda, UOC Pronto Soccorso & Med DUrgenza, Milan, Italy
[4] Univ Insubria, Dept Med & Surg, Como, Italy
[5] Univ Milan, Dept Biomed Sci Hlth, Milan, Italy
[6] AOU Careggi, Dept Neurosci Psychol Drug Res & Child Hlth NEURO, Florence, Italy
关键词:
Body temperature;
Diagnostic tests;
Fever;
Systematic review;
Thermometers;
TEMPORAL ARTERY THERMOMETRY;
RECTAL THERMOMETRY;
TYMPANIC MEMBRANE;
EAR THERMOMETRY;
CLINICAL ACCURACY;
SKIN THERMOMETER;
CHILDREN YOUNGER;
AXILLARY;
FEVER;
RELIABILITY;
D O I:
10.1007/s11739-020-02556-0
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
Not much is known about how accurate and reproducible different thermometers are at diagnosing patients with suspected fever. The study aims at evaluating which peripheral thermometers are more accurate and reproducible. We searched Medline, Embase, Scopus, WOS, CENTRAL, and Cinahl to perform: (1) diagnostic accuracy meta-analysis (MA) using rectal mercury-in-glass or digital thermometry as reference, and bivariate models for pooling; (2) network MA to estimate differences in mean temperature between devices; (3) Bland-Altman method to estimate 95% coefficient of reproducibility. PROSPERO registration: CRD42020174996. We included 46 studies enrolling more than 12,000 patients. Using 38 degrees C (100.4 degrees F) as cut-off temperature, temporal infrared thermometry had a sensitivity of 0.76 (95% confidence interval, 0.65, 0.84; low certainty) and specificity of 0.96 (0.92, 0.98; moderate certainty); tympanic infrared thermometry had a sensitivity of 0.77 (0.60, 0.88; low certainty) and specificity of 0.98 (0.95, 0.99; moderate certainty). For all the other index devices, it was not possible to pool the estimates. Compared to the rectal mercury-in-glass thermometer, mean temperature differences were not statistically different from zero for temporal or tympanic infrared thermometry; the median coefficient of reproducibility ranged between 0.53 degrees C [0.95 degrees F] for infrared temporal and 1.2 degrees C [2.16 degrees F] for axillary digital thermometry. Several peripheral thermometers proved specific, but not sensitive for diagnosing fever with rectal thermometry as a reference standard, meaning that finding a temperature below 38 degrees C does not rule out fever. Fixed differences between temperatures together with random error means facing differences between measurements in the order of 2 degrees C [4.5 degrees F]. This study informs practitioners of the limitations associated with different thermometers; peripheral ones are specific but not sensitive.
引用
收藏
页码:1071 / 1083
页数:13