Markerless high-frequency prospective motion correction for neuroanatomical MRI

被引:52
作者
Frost, Robert [1 ,2 ]
Wighton, Paul [1 ]
Karahanoglu, F. Isik [1 ,2 ]
Robertson, Richard L. [3 ]
Grant, P. Ellen [3 ,4 ]
Fischl, Bruce [1 ,2 ,5 ]
Tisdall, M. Dylan [6 ]
van der Kouwe, Andre [1 ,2 ]
机构
[1] Massachusetts Gen Hosp, Athinoula A Martinos Ctr Biomed Imaging, Charlestown, MA USA
[2] Harvard Med Sch, Dept Radiol, Boston, MA 02115 USA
[3] Harvard Med Sch, Boston Childrens Hosp, Dept Radiol, Boston, MA 02115 USA
[4] Boston Childrens Hosp, Fetal Neonatal Neuroimaging & Dev Sci Ctr, Boston, MA USA
[5] MIT, Comp Sci & Artificial Intelligence Lab, 77 Massachusetts Ave, Cambridge, MA 02139 USA
[6] Univ Penn, Dept Radiol, Perelman Sch Med, Philadelphia, PA 19104 USA
基金
美国国家卫生研究院;
关键词
brain morphometry; high-frequency prospective motion correction; markerless motion tracking; real-time motion correction; SURFACE-BASED ANALYSIS; REAL-TIME MOTION; HEAD MOTION; VOLUMETRIC NAVIGATORS; OPTICAL TRACKING; SELECTIVE REACQUISITION; BRAIN MORPHOMETRY;
D O I
10.1002/mrm.27705
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To integrate markerless head motion tracking with prospectively corrected neuroanatomical MRI sequences and to investigate high-frequency motion correction during imaging echo trains. Methods: A commercial 3D surface tracking system, which estimates head motion by registering point cloud reconstructions of the face, was used to adapt the imaging FOV based on head movement during MPRAGE and T-2 SPACE (3D variable flip-angle turbo spin-echo) sequences. The FOV position and orientation were updated every 6 lines of k-space (< 50 ms) to enable "within-echo-train" prospective motion correction (PMC). Comparisons were made with scans using "before-echo-train" PMC, in which the FOV was updated only once per TR, before the start of each echo train (ET). Continuous-motion experiments with phantoms and in vivo were used to compare these high-frequency and low-frequency correction strategies. MPRAGE images were processed with FreeSurfer to compare estimates of brain structure volumes and cortical thickness in scans with different PMC. Results: The median absolute pose differences between markerless tracking and MR image registration were 0.07/0.26/0.15 mm for x/y/z translation and 0.06 degrees/0.02 degrees/0.12 degrees for rotation about x/y/z. The PMC with markerless tracking substantially reduced motion artifacts. The continuous-motion experiments showed that within-ET PMC, which minimizes FOV encoding errors during ETs that last over 1 second, reduces artifacts compared with before-ET PMC. T-2 SPACE was found to be more sensitive to motion during ETs than MPRAGE. FreeSurfer morphometry estimates from within-ET PMC MPRAGE images were the most accurate. Conclusion: Markerless head tracking can be used for PMC, and high-frequency within-ET PMC can reduce sensitivity to motion during long imaging ETs.
引用
收藏
页码:126 / 144
页数:19
相关论文
共 55 条
[1]   Prospective motion correction for 3D pseudo-continuous arterial spin labeling using an external optical tracking system [J].
Aksoy, Murat ;
Maclaren, Julian ;
Bammer, Roland .
MAGNETIC RESONANCE IMAGING, 2017, 39 :44-52
[2]   Real-Time Optical Motion Correction for Diffusion Tensor Imaging [J].
Aksoy, Murat ;
Forman, Christoph ;
Straka, Matus ;
Skare, Stefan ;
Holdsworth, Samantha ;
Hornegger, Joachim ;
Bammer, Roland .
MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (02) :366-378
[3]   Subtle In-Scanner Motion Biases Automated Measurement of Brain Anatomy From In Vivo MRI [J].
Alexander-Bloch, Aaron ;
Clasen, Liv ;
Stockman, Michael ;
Ronan, Lisa ;
Lalonde, Francois ;
Giedd, Jay ;
Raznahan, Armin .
HUMAN BRAIN MAPPING, 2016, 37 (07) :2385-2397
[4]   Toward Quantifying the Prevalence, Severity, and Cost Associated With Patient Motion During Clinical MR Examinations [J].
Andre, Jalal B. ;
Bresnahan, Brian W. ;
Mossa-Basha, Mahmud ;
Hoff, Michael N. ;
Smith, C. Patrick ;
Anzai, Yoshimi ;
Cohen, Wendy A. .
JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2015, 12 (07) :689-695
[5]  
[Anonymous], MEASURING TRANSFORMA
[6]   Prospective motion correction with NMR markers using only native sequence elements [J].
Aranovitch, Alexander ;
Haeberlin, Maximilian ;
Gross, Simon ;
Dietrich, Benjamin E. ;
Wilm, Bertram J. ;
Brunner, David O. ;
Schmid, Thomas ;
Luechinger, Roger ;
Pruessmann, Klaas P. .
MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (04) :2046-2056
[7]  
Benjaminsen C, 2016, P 24 ANN M ISMRM SIN
[8]   EFFECTS OF DIFFUSION ON FREE PRECESSION IN NUCLEAR MAGNETIC RESONANCE EXPERIMENTS [J].
CARR, HY ;
PURCELL, EM .
PHYSICAL REVIEW, 1954, 94 (03) :630-638
[9]   Cortical surface-based analysis - I. Segmentation and surface reconstruction [J].
Dale, AM ;
Fischl, B ;
Sereno, MI .
NEUROIMAGE, 1999, 9 (02) :179-194
[10]   ADAPTIVE TECHNIQUE FOR HIGH-DEFINITION MR IMAGING OF MOVING STRUCTURES [J].
EHMAN, RL ;
FELMLEE, JP .
RADIOLOGY, 1989, 173 (01) :255-263