Arterial blood pressure monitoring in overweight critically ill patients: invasive or noninvasive?

被引:82
作者
Araghi, Ali [1 ]
Bander, Joseph J. [1 ]
Guzman, Jorge A. [1 ]
机构
[1] Wayne State Univ, Sch Med, Div Pulm Crit Care & Sleep Med, Detroit, MI 48202 USA
来源
CRITICAL CARE | 2006年 / 10卷 / 02期
关键词
D O I
10.1186/cc4896
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction Blood pressure measurements frequently guide management in critical care. Direct readings, commonly from a major artery, are considered to be the gold standard. Because arterial cannulation is associated with risks, alternative noninvasive blood pressure (NIBP) measurements are routinely used. However, the accuracy of NIBP determinations in overweight patients in the outpatient setting is variable, and little is known about critically ill patients. This prospective, observational study was performed to compare direct intra-arterial blood pressure ( IABP) with NIBP measurements obtained using auscultatory and oscillometric methods in overweight patients admitted to our medical intensive care unit. Method Adult critically ill patients with a body mass index (BMI) of 25 kg/m(2) or greater and a functional arterial line ( assessed using the rapid flush test) were enrolled in the study. IABP measurements were compared with those obtained noninvasively. A calibrated aneroid manometer ( auscultatory technique) with arm cuffs compatible with arm sizes and a NIBP monitor ( oscillometric technique) were used for NIBP measurements. Agreement between methods was assessed using Bland-Altman analysis. Results Fifty-four patients ( 23 males) with a mean ( +/- standard error) age of 57 +/- 3 years were studied. The mean BMI was 34.0 +/- 1.4 kg/m(2). Mean arm circumference was 32 +/- 0.6 cm. IABP readings were obtained from the radial artery in all patients. Only eight patients were receiving vasoactive medications. Mean overall biases for the auscultatory and oscillometric techniques were 4.1 +/- 1.9 and - 8.0 +/- 1.7 mmHg, respectively ( P < 0.0001), with wide limits of agreement. The overestimation of blood pressure using the auscultatory technique was more important in patients with a BMI of 30 kg/ m(2) or greater. In hypertensive patients both NIBP methods underestimated blood pressure as determined using direct IABP measurement. Conclusion Oscillometric blood pressure measurements underestimated IABP readings regardless of patient BMI. Auscultatory measurements were also inaccurate, tending to underestimate systolic blood pressure and overestimate mean arterial and diastolic blood pressure. NIBP can be inaccurate among overweight critically ill patients and lead to erroneous interpretations of blood pressure.
引用
收藏
页数:5
相关论文
共 17 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients [J].
Bur, A ;
Herkner, H ;
Vlcek, M ;
Woisetschläger, C ;
Derhaschnig, U ;
Delle Karth, G ;
Laggner, AN ;
Hirschl, MM .
CRITICAL CARE MEDICINE, 2003, 31 (03) :793-799
[3]   Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients [J].
Bur, A ;
Hirschl, MM ;
Herkner, H ;
Oschatz, E ;
Kofler, J ;
Woisetschläger, C ;
Laggner, AN .
CRITICAL CARE MEDICINE, 2000, 28 (02) :371-376
[4]   Clinical approach to the critically ill, morbidly obese patient [J].
El-Solh, AA .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2004, 169 (05) :557-561
[5]  
FINNIE KJC, 1984, CRIT CARE MED, V12, P965
[6]   DIRECT BLOOD-PRESSURE MEASUREMENT - DYNAMIC-RESPONSE REQUIREMENTS [J].
GARDNER, RM .
ANESTHESIOLOGY, 1981, 54 (03) :227-236
[7]   An evaluation of a new continuous blood pressure monitoring system in critically ill patients [J].
Heard, SO ;
Lisbon, A ;
Toth, I ;
Ramasubramanian, R .
JOURNAL OF CLINICAL ANESTHESIA, 2000, 12 (07) :509-518
[8]   Measuring blood pressure accurately - New and persistent challenges [J].
Jones, DW ;
Appel, LJ ;
Sheps, SG ;
Roccella, EJ ;
Lenfant, C .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2003, 289 (08) :1027-1030
[9]   The obese patient in the ICU [J].
Marik, P ;
Varon, J .
CHEST, 1998, 113 (02) :492-498
[10]   ACCURACY OF AUSCULTATORY BLOOD-PRESSURE MEASUREMENTS IN HYPERTENSIVE AND OBESE SUBJECTS [J].
NIELSEN, PE ;
LARSEN, B ;
HOLSTEIN, P ;
POULSEN, HL .
HYPERTENSION, 1983, 5 (01) :122-127