Monte Carlo evaluation of the potential benefits of flattening filter free beams from the Oncor® clinical linear accelerator

被引:4
作者
Asadi, Amin [1 ]
Razavi-Ratki, Seid Kazem [2 ,4 ]
Jabbari, Keyvan [3 ]
Najafzadeh, Milad [1 ]
Nickfarjam, Abolfazl [1 ,2 ]
机构
[1] Shahid Sadoughi Univ Med Sci, Fac Med, Med Phys Dept, Prof Hesabi Blvd,Alem Sq, Yazd, Iran
[2] Shahid Sadoughi Univ Med Sci, Radiotherapy Res Ctr, Yazd, Iran
[3] Isfahan Univ Med Sci, Med Phys Dept, Fac Med, Esfahan, Iran
[4] Shahid Sadoughi Univ Med Sci, Radiol Dept, Fac Med, Yazd, Iran
关键词
Radiation therapy; medical linear accelerator; flattening filter; Monte Carlo calculations; BEAMnrc; Monte Carlo code; PHOTON-BEAM; RADIOTHERAPY;
D O I
10.3233/XST-17315
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
OBJECTIVES: To evaluate the potential privileges of flattening filter-free (FFF) photon beams from Oncor (R) linac for 6MV and 18MV energies. METHODS: A Monte Carlo (MC) model of Oncor (R) linac was built using BEAMnrc MCCode and verified by the measured data using 6MV and 18MV energies. A comprehensive set of data was also characterized for MC model of Oncor (R) machine running with and without flattening filter (FF) for 6MV and 18MV beams in six field sizes. The investigated characteristics included mean energy, energy spectrum, photon spatial fluence, superficial dose, percent depth dose (PDD), dose output, and out-of-field dose with two indexes of lateral dose profile and isodose curve at three depths. RESULTS: Using FFF enhanced the energy uniformity 3.4 +/- 0.11% (6 MV) and 2.05 +/- 0.09% (18 MV) times and improved dose output by factor of 2.91 (6 MV) and 4.2 (18 MV) on the central axis, respectively. Using FFF also reduced the PDD dependencies by 9.1% (6 MV) and 5.57% (18 MV). In addition, using FFF had a lower out-of-field dose due to the reduced head scatter and softer spectra. CONCLUSIONS: The findings in this study suggested that using FFF, Oncor (R) machine could achieve better treatment results with lower dose toxicity and a shorter beam-on time.
引用
收藏
页码:281 / 302
页数:22
相关论文
共 18 条
  • [1] Monte Carlo study of in-field and out-of-field dose distributions from a linear accelerator operating with and without a flattening-filter
    Almberg, S. S.
    Frengen, J.
    Lindmo, T.
    [J]. MEDICAL PHYSICS, 2012, 39 (08) : 5194 - 5203
  • [2] Comparison of Head Scatter Factor for 6MV and 10MV flattened (FB) and Unflattened (FFF) Photon Beam using indigenously Designed Columnar Mini Phantom
    Ashokkumar, Sigamani
    Raj, Arunai Nambi N.
    Sinha, Sujit Nath
    Yadav, Girigesh
    Thiyagarajan, Rajesh
    Raman, Kothanda
    Mishra, Manindra Bhushan
    [J]. JOURNAL OF MEDICAL PHYSICS, 2014, 39 (03) : 184 - 191
  • [3] A Monte Carlo study of a flattening filter-free linear accelerator verified with measurements
    Dalaryd, Marten
    Kragl, Gabriele
    Ceberg, Crister
    Georg, Dietmar
    McClean, Brendan
    af Wetterstedt, Sacha
    Wieslander, Elinore
    Knoos, Tommy
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23) : 7333 - 7344
  • [4] Monte Carlo comparison of superficial dose between flattening filter free and flattened beams
    Javedan, Khosrow
    Feygelman, Vladimir
    Zhang, Ray R.
    Moros, Eduardo G.
    Correa, Candace R.
    Trotti, Andy
    Li, Weiqi
    Zhang, Geoffrey G.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (04): : 503 - 508
  • [5] Kawrakow I., 2000, PIRS877 NRCC
  • [6] Kawrakow I, 2000, PIRS701 NRC
  • [7] Comparison between Dual Arc VMAT and 7F-IMRT in the protection of hippocampus for patients during whole brain radiotherapy
    Li, Jun
    Tang, Xiao-Bin
    Wang, Bu-Hai
    Chen, Xue-Mei
    Chen, Da
    Chai, Lei
    [J]. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2016, 24 (03) : 457 - 466
  • [8] Ma C, 1995, PIRS0509D NRC
  • [9] Monte Carlo study on a flattening filter-free 18-MV photon beam of a medical linear accelerator
    Mesbahi, Asghar
    Nejad, Farshad Seyed
    [J]. RADIATION MEDICINE, 2008, 26 (06): : 331 - 336
  • [10] The physical characteristics of the 15 MV Varian Clinac 2100C unflattened beam
    Najem, M. A.
    Spyrou, N. M.
    Podolyak, Z.
    Abolaban, F. A.