Performance evaluation of the new hematology analyzer UniCel DxH 900

被引:18
作者
Martinez-Iribarren, Alicia [1 ]
Tejedor, Xavier [1 ]
Sala Sanjaume, Angels [1 ]
Leis, Alba [1 ]
Dolade Botias, Maria [1 ]
Morales-Indiano, Cristian [1 ]
机构
[1] Hosp Badalona Germans Trias & Pujol, Dept Lab Med, Lab Clin Metropolitana Nord LCMN, Badalona, Spain
关键词
cell blood count; DxH900; flag; hematology analyzer; leukocyte differential; performance evaluation; BLOOD-CELL COUNTS; SAPPHIRE;
D O I
10.1111/ijlh.13448
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction UniCel DxH900 (Beckman Coulter, Miami, Florida, USA) is a quantitative, multi-parameter, automated hematology analyzer for in vitro diagnostic use in clinical laboratories. The aim of this study was to evaluate the analytical performance of the new DxH900 analyzer to verify its diagnostic and clinical utility in the hematology laboratory of a tertiary care hospital in Spain. The most important and novel feature offered by DxH900 analyzer is providing MDW (monocyte distribution width), a new hematologic parameter which is being clinically validated as an early sepsis indicator with promising results. Methods We evaluated imprecision (including MDW), linearity, and carryover of DxH900. Method comparison for cell blood count (CBC) was performed in relation to DxH800 with 100 samples. We compared leukocyte differential (DIFF) from DxH900 with manual 400-cell differential. 390 samples were assessed for flag performance. Results Results obtained for between days and within-run imprecision were good. DxH900 showed excellent linearity (R = 1.00) over analytical range for white blood cells, red blood cells, hemoglobin, platelets, and reticulocyte count (RET) (R = 0.96) and no significant carryover effect. CBC and RET on the DxH900 correlated well with DxH800 (R >= 0.99). Comparison with manual differential showed excellent correlation (R >= 0.88), except for basophils. Flagging performance exhibited sensitivity over 90% for majority of alarm messages and very high negative predictive value (over 95%). Conclusion UniCel DxH900 Coulter analyzer provides reliable results and fully comparable to DxH800. DxH900 is an accurate, highly precise analyzer with good analytical performances to be used effectively in high-volume laboratories.
引用
收藏
页码:623 / 631
页数:9
相关论文
共 19 条
[1]   Is the Blood Basophil Count Sufficiently Precise, Accurate, and Specific? Three Automated Hematology Instruments and Flow Cytometry Compared [J].
Amundsen, Erik K. ;
Henriksson, Carola E. ;
Holthe, Mette R. ;
Urdal, Petter .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2012, 137 (01) :86-92
[2]  
[Anonymous], 2010, H26A2 CLIN LAB STAND
[3]  
[Anonymous], 2003, EP06A CLIN LAB STAND
[4]   ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting [J].
Briggs, C. ;
Culp, N. ;
Davis, B. ;
d'Onofrio, G. ;
Zini, G. ;
Machin, S. J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 (06) :613-627
[5]   Evaluation and comparison of automated hematology analyzer, flow cytometry, and digital morphology analyzer for monocyte counting [J].
Buoro, S. ;
Moioli, V. ;
Seghezzi, M. ;
Previtali, G. ;
Alessio, M. G. ;
Lopez, R. Simon ;
Ortolani, C. ;
Ottomano, C. ;
Lippi, G. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 (05) :577-585
[6]   White Blood Cell Counts Reference Methodology [J].
Chabot-Richards, Devon S. ;
George, Tracy I. .
CLINICS IN LABORATORY MEDICINE, 2015, 35 (01) :11-+
[7]  
Clinical and Laboratory Standards Institute, 2014, EP10A3 CLIN LAB STAN
[8]   Monocyte Distribution Width: A Novel Indicator of Sepsis-2 and Sepsis-3 in High-Risk Emergency Department Patients [J].
Crouser, Elliott D. ;
Parrillo, Joseph E. ;
Seymour, Christopher W. ;
Angus, Derek C. ;
Bicking, Keri ;
Esguerra, Vincent G. ;
Peck-Palmer, Octavia M. ;
Magari, Robert T. ;
Julian, Mark W. ;
Kleven, Jennifer M. ;
Raj, Paarth J. ;
Procopio, Gabrielle ;
Careaga, Diana ;
Tejidor, Liliana .
CRITICAL CARE MEDICINE, 2019, 47 (08) :1018-1025
[9]   Improved Early Detection of Sepsis in the ED With a Novel Monocyte Distribution Width Biomarker [J].
Crouser, Elliott D. ;
Parrillo, Joseph E. ;
Seymour, Christopher ;
Angus, Derek C. ;
Bicking, Keri ;
Tejidor, Liliana ;
Magari, Robert ;
Careaga, Diana ;
Williams, JoAnna ;
Closser, Douglas R. ;
Samoszuk, Michael ;
Herren, Luke ;
Robart, Emily ;
Chaves, Fernando .
CHEST, 2017, 152 (03) :518-526
[10]   Development, History, and Future of Automated Cell Counters [J].
Green, Ralph ;
Wachsmann-Hogiu, Sebastian .
CLINICS IN LABORATORY MEDICINE, 2015, 35 (01) :1-+