Site Contributions for Radiated Emission Measurement Uncertainties above 1 GHz

被引:0
作者
Chen, Zhong [1 ]
Xiong, Zubiao [1 ]
机构
[1] ETS Lindgren Inc, 1301 Arrow Point Dr, Cedar Pk, TX 78613 USA
来源
2017 IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY & SIGNAL/POWER INTEGRITY (EMCSI) | 2017年
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Guidance on accounting for site imperfections in the uncertainty evaluation of emissions measurement result is provided by CISPR 16-4-2. Site Voltage Standing Wave Ratio (SVSWR) results from the chamber are used as the basis for estimating the uncertainty contributions due to the site. In this paper, we investigate two aspects of the uncertainty evaluation, and provide a method to estimate the uncertainties based on measured SVSWR data. Firstly, considering that the SVSWR results are under-sampled and under-reported using the current method, is it valid to assume 6 dB as the maximum SVSWR of the chamber? Secondly, SVSWR measurements are only performed at the front, left, right, and center (optionally) of the quiet zone (QZ). We study if those locations are adequate representations of the entire QZ performance.
引用
收藏
页码:504 / 509
页数:6