3-D optimal shape design of ferromagnetic pole in MRI magnet of open permanent-magnet type

被引:16
作者
Kim, DH [1 ]
Kim, BS [1 ]
Lee, JH [1 ]
Nah, WS [1 ]
Park, H [1 ]
机构
[1] Sungkyunkwan Univ, Sch Elect & Comp Engn, Suwon 440746, South Korea
关键词
homogeneous magnetic field; magnetic resonance imaging (MRI) devices; open permanent magnet type; optimal shape design;
D O I
10.1109/TASC.2002.1018679
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents 3-D optimal shape design method of ferromagnetic pole pieces for magnetic resonance imaging (MRI) devices of open permanent magnet type. The design method is based on, a continuum sensitivity analysis in conjunction with the boundary integral equation method (BIEM). The magnetic poles of an open magnet assembly are designed by the proposed method and the conventional shimming method. Magnetic field strength, ppm order homogeneity and working space volume are compared between the proposed method and conventional one.
引用
收藏
页码:1467 / 1470
页数:4
相关论文
共 6 条
[1]   THE DESIGN AND CONSTRUCTION OF HIGH FIELD-UNIFORMITY PERMANENT-MAGNET SYSTEM FOR MRI [J].
FENG, ZX ;
JIANG, XH ;
HAN, S .
IEEE TRANSACTIONS ON MAGNETICS, 1992, 28 (01) :641-643
[2]   A CAD-based design parameterization for shape optimization of elastic solids [J].
Hardee, E ;
Chang, KH ;
Tu, J ;
Choi, KK ;
Grindeanu, I ;
Yu, XM .
ADVANCES IN ENGINEERING SOFTWARE, 1999, 30 (03) :185-199
[3]   Optimization method for permanent-magnet structures [J].
Jensen, JH .
IEEE TRANSACTIONS ON MAGNETICS, 1999, 35 (06) :4465-4472
[4]  
KIM DH, 2000, P CEFC 2000 JUN, P132
[5]   DESIGN OF AN AXISYMMETRIC PERMANENT-MAGNET STRUCTURE FOR MAGNETIC-RESONANCE TOMOGRAPHY [J].
MOLFINO, P ;
REPETTO, M ;
BIXIO, A ;
MUT, GD ;
MARABOTTO, R .
IEEE TRANSACTIONS ON MAGNETICS, 1988, 24 (02) :994-997
[6]   Design procedure for permanent magnet assemblies with uniform magnetic fields for MRI devices [J].
Podol'skii, A .