Surface characteristics and corrosion resistance properties of TiNi shape memory alloy coated with Ta

被引:54
作者
Cheng, Y [1 ]
Cai, W [1 ]
Li, HT [1 ]
Zheng, YF [1 ]
Zhao, LC [1 ]
机构
[1] Harbin Inst Technol, Sch Mat Sci & Engn, Harbin 150001, Peoples R China
关键词
TiNi alloy; ion plating; annealing; corrosion;
D O I
10.1016/j.surfcoat.2004.01.012
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Ti-50.6 at.% Ni alloy samples were coated with tantalum by the arc ion plating method with the aim to improve their radiopacity. Surface characteristics and corrosion resistance properties were investigated in the present study. The results of XRD show that the film as-deposited was composed of a metastable tetragonal p-phase tantalum which can transform into alpha-phase tantalum after annealing at 700degreesC, 800degreesC, 900degreesC for 1 h, respectively. TEM observations show that thin nano-structured Ta layers were obtained in the as-deposited samples. Mixture of microcrystallite (approx. 40 nm) and large-size crystal (approx. 150 nm) beta-phase tantalum was formed in the film after annealing at 700degreesC. When annealed at 900degrees'C, the film is composed of dominantly uniform large-size alpha-phase tantalum. The results of XPS survey and high resolution spectra show that a thin oxide film with Ta2O5 in the outmost layer and tantalum suboxides in the inner layer, are formed inside tantalum coating as a result of natural passivation of Ta in the atmosphere. The corrosion resistance was determined by electrochemical methods in 0.9% NaCl solution at 37degreesC. Both the as-deposited and annealed samples exhibit excellent corrosion resistance property. Compared to the untreated coating samples, the annealed Ta coating samples exhibit higher open-circuit corrosion potential (E-corr), breakdown potential (E-br), and lower passive current density (I-p). The results indicate that the annealing treatment can improve the corrosion resistance property. (C) 2004 Elsevier B.V All rights reserved.
引用
收藏
页码:346 / 352
页数:7
相关论文
共 16 条
  • [1] TANTALUM AND ITS ALLOYS
    CARDONNE, SM
    KUMAR, P
    MICHALUK, CA
    SCHWARTZ, HD
    [J]. INTERNATIONAL JOURNAL OF REFRACTORY METALS & HARD MATERIALS, 1995, 13 (04) : 187 - &
  • [2] Radiopacity of tantalum oxide nanoparticle filled resins
    Chan, DCN
    Titus, HW
    Chung, KH
    Dixon, H
    Wellinghoff, ST
    Rawls, HR
    [J]. DENTAL MATERIALS, 1999, 15 (03) : 219 - 222
  • [3] An overview of nitinol medical applications
    Duerig, T
    Pelton, A
    Stöckel, D
    [J]. MATERIALS SCIENCE AND ENGINEERING A-STRUCTURAL MATERIALS PROPERTIES MICROSTRUCTURE AND PROCESSING, 1999, 273 : 149 - 160
  • [4] An overview of superelastic stent design
    Duerig, TW
    Tolomeo, DE
    Wholey, M
    [J]. MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2000, 9 (3-4) : 235 - 246
  • [5] METABOLISM OF TANTALUM IN RAT
    FLESHMAN, DG
    SILVA, AJ
    SHORE, B
    [J]. HEALTH PHYSICS, 1971, 21 (03): : 385 - +
  • [6] HODGSON DE, 1997, SMST 97, P461
  • [7] Ma JL, 2000, J MATER SCI TECHNOL, V16, P534
  • [8] EFFECTIVENESS OF CORONARY STENTING FOR THE TREATMENT OF CHRONIC TOTAL OCCLUSION IN ANGINA-PECTORIS
    MEDINA, A
    MELIAN, F
    DELEZO, JS
    PAN, M
    ROMERO, M
    HERNANDEZ, E
    MARRERO, J
    ORTEGA, JR
    PAVLOVIC, D
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1994, 73 (16) : 1222 - 1224
  • [9] Moulder F., 1992, HDB XRAY PHOTOELECTR, P171
  • [10] ELECTROCHEMICAL CORROSION OF METALLIC BIOMATERIALS
    POURBAIX, M
    [J]. BIOMATERIALS, 1984, 5 (03) : 122 - 134