Bone mineral density measurement with dental quantitative CT prior to dental implant placement in cadaver mandibles: Pilot study

被引:84
作者
Homolka, P
Beer, A
Birkfellner, W
Nowotny, R
Gahleitner, A
Tschabitscher, M
Bergmann, H
机构
[1] Univ Vienna, Gen Hosp, Dept Biomed Engn & Phys, A-1090 Vienna, Austria
[2] Univ Vienna, Gen Hosp, Dept Diagnost Radiol, A-1090 Vienna, Austria
[3] Univ Vienna, Sch Dent, Dept Oral Surg, Vienna, Austria
[4] Univ Vienna, Dept Anat, A-1010 Vienna, Austria
[5] Evangelisch Krankenhaus, Dept Maxillofacial Surg, Vienna, Austria
关键词
bones; mineralization; computer tomography (CT); three-dimensional; experimental study; jaws; CT; teeth;
D O I
10.1148/radiol.2241010948
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To correlate torque forces during insertion of screw-type dental implants with bone mineral density (BMD) values determined preoperatively. MATERIALS AND METHODS: Dental quantitative computed tomography (CT) was performed with simultaneous imaging of five postmortem mandibles and a calibration standard containing defined concentrations of calcium hydroxyapatite. CT numbers were converted to local BMD values by assuming a linear relationship (BMD = a x HU + b), where a and b are calibration coefficients. The a, b, P, and t values, correlation coefficients, and standard errors were calculated. Dental implants (n = 25) were set, and insertion torques were recorded. BMD was determined at the implantation site and correlated with torque forces recorded during implant insertion. Calibration coefficients derived for specimens were compared with those derived for actual patients. RESULTS: Calibration coefficients (at 120 W) for the postmortem specimens were a = 0.760 +/- 0.03 (mean +/- SD) and b = 2.8 +/- 3.7 and for the patients were a = 0.804 +/- 0.06 and b = 5.2 +/- 4.2. Calibrated BMD values at the location of dental implants exhibit a significant correlation (R-2 = 0.83, P <.001) with insertion torques on the basis of a second-order model, which yields torque = (0.0055 X BMD + 0.73)(2) for the implants used and the surgical technique applied. CONCLUSION: Correlation exists between BMD measured with dental quantitative CT and the insertion torque of dental implants. (C) RSNA, 2002.
引用
收藏
页码:247 / 252
页数:6
相关论文
共 26 条
[1]  
BEER A, 2000, THESIS U VIENNA AUST
[2]  
Branemark P I, 1999, Clin Implant Dent Relat Res, V1, P2, DOI 10.1111/j.1708-8208.1999.tb00086.x
[3]  
Friberg B, 1999, INT J ORAL MAX SURG, V28, P297
[4]   Stability measurements of one-stage Branemark implants during healing in mandibles -: A clinical resonance frequency analysis study [J].
Friberg, B ;
Sennerby, L ;
Linden, B ;
Gröndahl, K ;
Lekholm, U .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1999, 28 (04) :266-272
[5]   IDENTIFICATION OF BONE QUALITY IN CONJUNCTION WITH INSERTION OF TITANIUM IMPLANTS - A PILOT-STUDY IN JAW AUTOPSY SPECIMENS [J].
FRIBERG, B ;
SENNERBY, L ;
ROOS, J ;
LEKHOLM, U .
CLINICAL ORAL IMPLANTS RESEARCH, 1995, 6 (04) :213-219
[6]   EVALUATION OF BONE-DENSITY USING CUTTING RESISTANCE MEASUREMENTS AND MICRORADIOGRAPHY - AN IN-VITRO STUDY IN PIG RIBS [J].
FRIBERG, B ;
SENNERBY, L ;
ROOS, J ;
JOHANSSON, P ;
STRID, CG ;
LEKHOLM, U .
CLINICAL ORAL IMPLANTS RESEARCH, 1995, 6 (03) :164-171
[7]   Basic study of the measurement of bone mineral content of cortical and cancellous bone of the mandible by computed tomography [J].
Iwashita, Y .
DENTOMAXILLOFACIAL RADIOLOGY, 2000, 29 (04) :209-215
[8]  
Jacobs R, 1998, RADIOGRAPHIC PLANNIN
[9]   THE EXCESSIVE LOSS OF BRANEMARK FIXTURES IN TYPE-IV BONE - A 5-YEAR ANALYSIS [J].
JAFFIN, RA ;
BERMAN, CL .
JOURNAL OF PERIODONTOLOGY, 1991, 62 (01) :2-4
[10]  
Johansson P., 1994, Int J Oral Maxillofac Implants, V9, P279