Comparison of the patient-activated event recording system vs. traditional 24 h Holter electrocardiography in individuals with paroxysmal palpitations or dizziness

被引:31
作者
de Asmundis, Carlo [1 ]
Conte, Giulio [1 ]
Sieira, Juan [1 ]
Chierchia, Gian-Battista [1 ]
Rodriguez-Manero, Moises [1 ]
Di Giovanni, Giacomo [1 ]
Ciconte, Giuseppe [1 ]
Levinstein, Moises [1 ]
Baltogiannis, Giannis [1 ]
Saitoh, Yukio [1 ]
Casado-Arroyo, Ruben [1 ]
Brugada, Pedro [1 ]
机构
[1] Vrije Univ Brussel, Univ Ziekenhuis Brussel, Heart Rhythm Management Ctr, B-1090 Brussels, Belgium
来源
EUROPACE | 2014年 / 16卷 / 08期
关键词
Palpitations; Patient-activated event recorder; 24 h Holter-ECG recording; HeartScan; DIAGNOSTIC YIELD; UTILITY; COST; ARRHYTHMIAS; SYNCOPE;
D O I
10.1093/europace/eut411
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Electrocardiographic documentation of symptomatic episodes of palpitations by means of traditional methods such as 24 h Holter monitoring (HM) or loop recorders is challenging. Patient-activated electrocardiography (ECG) recorders have been proved to be a useful tool in the diagnosis of arrhythmias in these patients. However, no comparison studies between the two techniques have been conducted. The aim of this study was to compare the diagnostic value of Holter ECG and a patient-activated event recorder (OMRON portable HeartScan ECG Monitor(A (R))) (HeartScan) in the detection of arrhythmias in patients with paroxysmal palpitations or dizziness suggestive of cardiac arrhythmias. Patients with paroxysmal palpitations or dizziness were eligible for this study. All patients underwent an HM for 24 h and a 15-day HeartScan after the HM. Six hundred and twenty-five patients (48% male, mean age: 37 +/- 11 years) were included in the study. All patients present with normal heart structure, normal baseline 12-lead ECG, and normal echocardiogram. Indications for ECG monitoring were palpitations in 577 patients (92.3%) and dizziness in 48 (7.7%). Holter monitoring offered a clinical diagnosis in 11 patients (1.8%). Conversely, HeartScan diagnosed the clinical arrhythmia in 558 individuals (89%). Detection of symptoms-related arrhythmias by means of HeartScan was significantly higher when compared with HM (P < 0.01). The studied system proved to be an efficient event recorder providing the diagnosis of the clinical arrhythmia in 89% of patients with paroxysmal palpitations or dizziness. Further studies are needed to confirm our results.
引用
收藏
页码:1231 / 1235
页数:5
相关论文
共 12 条
[1]   Norman J. "Jeff" Holter - "Father" of ambulatory ECG monitoring [J].
Barold, SS .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2005, 14 (02) :117-118
[2]   Utility and cost of event recorders in the diagnosis of palpitations, presyncope, and syncope [J].
Fogel, RI ;
Evans, JJ ;
Prystowsky, EN .
AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (02) :207-&
[3]   Prospective, multicentre validation of a simple, patient-operated electrocardiographic system for the detection of arrhythmias and electrocardiographic changes [J].
Kaleschke, Gerrit ;
Hoffmann, Boris ;
Drewitz, Imke ;
Steinbeck, Gerhard ;
Naebauer, Michael ;
Goette, Andreas ;
Breithardt, Guenter ;
Kirchhof, Paulus .
EUROPACE, 2009, 11 (10) :1362-1368
[4]   Cardiac event recorders yield more diagnoses and are more cost-effective than 48-hour holter monitoring in patients with palpitations - A controlled clinical trial [J].
Kinlay, S ;
Leitch, JW ;
Neil, A ;
Chapman, BL ;
Hardy, DB ;
Fletcher, PJ .
ANNALS OF INTERNAL MEDICINE, 1996, 124 (01) :16-20
[5]   INCREMENTAL DIAGNOSTIC YIELD OF LOOP ELECTROCARDIOGRAPHIC RECORDERS IN UNEXPLAINED SYNCOPE [J].
LINZER, M ;
PRITCHETT, ELC ;
PONTINEN, M ;
MCCARTHY, E ;
DIVINE, GW .
AMERICAN JOURNAL OF CARDIOLOGY, 1990, 66 (02) :214-219
[6]  
Mason J, 2000, CECIL TXB MED, P230
[7]   First experience of monitoring with cardiac event recorder electrocardiography Omron system in childhood population for sporadic, potentially arrhythmia-related symptoms [J].
Park, Michael Hoe Kwung ;
de Asmundis, Carlo ;
Chierchia, Gian Battista ;
Sarkozy, Andrea ;
Benatar, Abraham ;
Brugada, Pedro .
EUROPACE, 2011, 13 (09) :1335-1339
[8]   Prospective comparison of the diagnostic utility of a standard event monitor versus a "leadless" portable ECG monitor in the evaluation of patients with palpitations [J].
Scherr, Daniel ;
Dalal, Darshan ;
Henrikson, Charles A. ;
Spragg, David D. ;
Berger, Ronald D. ;
Calkins, Hugh ;
Cheng, Alan .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2008, 22 (01) :39-44
[9]  
Schuchert A, 2002, Z KARDIOL, V91, P62, DOI 10.1007/s392-002-8373-4
[10]   Utility of patient-activated cardiac event recorders in the detection of cardiac arrhythmias [J].
Wu, CC ;
Hsieh, MH ;
Tai, CT ;
Chiang, CE ;
Yu, WC ;
Lin, YK ;
Tsao, HM ;
Ding, PYA ;
Chen, SA .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2003, 8 (02) :117-120