Dependability of Posterior-Segment Spectral Domain Optical Coherence Tomography for Measuring Central Corneal Thickness

被引:4
作者
Correa-Perez, Maria E. [1 ,2 ]
Olmo, Natalia [2 ]
Lopez-Miguel, Alberto [1 ]
Fernandez, Itziar [1 ,3 ]
Coco-Martin, Mara B. [1 ]
Maldonado, Miguel J. [1 ]
机构
[1] Univ Valladolid, IOBA, E-47011 Valladolid, Spain
[2] Hosp Inmaculada, Dept Oftalmol, Almeria, Spain
[3] Ctr Invest Biomed Red Bioingn Biomat & Nanomed CI, Valladolid, Spain
关键词
spectral domain optical coherence tomography; central corneal thickness; reliability; accuracy; IN-SITU KERATOMILEUSIS; HIGH MYOPIA; PACHYMETRY; REPRODUCIBILITY; ECTASIA; REPEATABILITY; RISK;
D O I
10.1097/ICO.0000000000000253
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: The aim of this study was to assess the intersession and interobserver reliability of central corneal thickness (CCT) measurements using a posterior-segment spectral domain optical coherence tomography (SD-OCT). Methods: Sixty patients were examined with a posterior-segment SD-OCT system (Topcon 3D-OCT). To analyze intersession reproducibility, 1 examiner measured CCT in 60 eyes 2 times 1 week apart, and to study interobserver reproducibility, a second examiner obtained another measurement during the first session. Within-subject reproducibility, coefficient of reproducibility (CR), and intraclass correlation coefficients (ICC) were calculated. Agreement between Topcon 3D OCT and ultrasound pachymetry (USP) was also assessed. Results: Posterior-segment SD-OCT showed an intersession reproducibility of 12.7 mu m (CR = 2.4%) and an excellent ICC (0.96). No systematic difference between sessions was found (average difference, 1.7 mu m; P = 0.15). We found no significant bias between observers (mean difference, -1.5 mu m; P = 0.29). The interobserver reproducibility was 16.4 mu m (CR = 3.1%), and the ICC was 0.94. Topcon 3D OCT CCT values were lower than USP readings (mean difference, 7.6 +/- 8.1 mu m; 95% confidence interval, 9.9-5.3 mu m; P < 0.001). Conclusions: With a commercially available retinal SD-OCT system, reliable intersession and interobserver CCT measurements can be feasibly obtained. The criteria for a significant change, which would be the one exceeding the reproducibility, indicate that intersession and interobserver variations in CCT of more than 13 and 16 mm, respectively, may reflect true corneal change with this particular device. These estimates should help investigators and clinicians differentiate actual CCT modification from measurement random error. The slight CCT underestimation with respect to USP is within the range of other OCT devices.
引用
收藏
页码:1219 / 1224
页数:6
相关论文
共 26 条
[1]   Evaluation of a Risk Factor Scoring System for Corneal Ectasia After LASIK in Eyes With Normal Topography [J].
Binder, Perry S. ;
Trattler, William B. .
JOURNAL OF REFRACTIVE SURGERY, 2010, 26 (04) :241-250
[2]   Measurement error and correlation coefficients [J].
Bland, JM ;
Altman, DG .
BRITISH MEDICAL JOURNAL, 1996, 313 (7048) :41-42
[3]  
Bland M., 2000, An introduction to medical statistics, P269
[4]   Corneal oedema after cataract surgery: predisposing factors and corneal graft outcome [J].
Claesson, Margareta ;
Armitage, W. John ;
Stenevi, Ulf .
ACTA OPHTHALMOLOGICA, 2009, 87 (02) :154-159
[5]   Long-term results of laser in situ keratomileusis for high myopia: Risk for ectasia [J].
Condon, Patrick I. ;
O'Keefe, Michael ;
Binder, Perry S. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (04) :583-590
[6]   Precision of High Definition Spectral-Domain Optical Coherence Tomography for Measuring Central Corneal Thickness [J].
Correa-Perez, Maria E. ;
Lopez-Miguel, Alberto ;
Miranda-Anta, Silvia ;
Iglesias-Cortinas, Dario ;
Alio, Jorge L. ;
Maldonado, Miguel J. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2012, 53 (04) :1752-1757
[7]   Mapping Corneal Thickness Using Dual-Scheimpflug Imaging at Different Stages of Keratoconus [J].
Demir, Selim ;
Ortak, Huseyin ;
Yeter, Volkan ;
Alim, Sait ;
Sayin, Osman ;
Tas, Ufuk ;
Sonmez, Baris .
CORNEA, 2013, 32 (11) :1470-1474
[8]  
Domalpally Amitha, 2010, Ophthalmic Surg Lasers Imaging, V41 Suppl, pS34, DOI 10.3928/15428877-20100325-01
[9]   Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach [J].
Doughty, MJ ;
Zaman, ML .
SURVEY OF OPHTHALMOLOGY, 2000, 44 (05) :367-408
[10]   The Role of Axial Resolution of Optical Coherence Tomography on the Measurement of Corneal and Epithelial Thicknesses [J].
Ge, Lili ;
Yuan, Yimin ;
Shen, Meixiao ;
Tao, Aizhu ;
Wang, Jianhua ;
Lu, Fan .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (01) :746-755