The influence of stress conditions on the wear of UHMWPE for total joint replacements

被引:56
作者
Barbour, PSM
Barton, DC
Fisher, J
机构
[1] Department of Mechanical Engineering, University of Leeds, Leeds
基金
英国工程与自然科学研究理事会;
关键词
D O I
10.1023/A:1018515318630
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In vitro studies of the effect of contact stress on the wear of ultra high molecular weight polyethylene (UHMWPE) in orthopaedic applications have produced contradictory results which predict both increased and decreased wear with increasing contact stress. In vivo studies of functioning hip prostheses have reported that 22 mm femoral heads generate lower linear and volumetric wear rates than 32 mm femoral heads. The effect of decreasing the head size will increase the contact stress but decrease the sliding distance per motion cycle. The present study consists of wear experiments under a range of contact stress magnitudes and application conditions in order to simulate the wear processes occurring in vivo. The results from these tests indicated that the wear factor actually decreases with increasing contact stress if the stress was not varied with time. If a time dependent or spatially varying stress was applied, the wear factor can increase greatly when compared to similar magnitude constant contact stress. This effect may be due to the complex relationship between the rate of wear particle generation and the rate at which the particles are released from the interface. The results of these wear experiments are discussed in terms of the influence of the stress conditions upon potential wear processes in total hip and knee prostheses.
引用
收藏
页码:603 / 611
页数:9
相关论文
共 25 条
  • [1] THE INFLUENCE OF CONTACT STRESS ON THE WEAR OF UHMWPE FOR TOTAL REPLACEMENT HIP PROSTHESES
    BARBOUR, PSM
    BARTON, DC
    FISHER, J
    [J]. WEAR, 1995, 181 : 250 - 257
  • [2] THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT
    BARTEL, DL
    BICKNELL, VL
    WRIGHT, TM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) : 1041 - 1051
  • [3] BLUNN GW, 1991, CLIN ORTHOPAEDICS, V273, P253
  • [4] CLARKE IC, 1989, PROGR BIOENGINEERING, pCH18
  • [5] MACROSCOPIC AND MICROSCOPIC WEAR MECHANISMS IN ULTRA-HIGH-MOLECULAR-WEIGHT POLYETHYLENE
    COOPER, JR
    DOWSON, D
    FISHER, J
    [J]. WEAR, 1993, 162 : 378 - 384
  • [6] ANALYSIS OF SUBSURFACE CRACK-PROPAGATION AND IMPLICATIONS FOR WEAR OF ELASTICALLY DEFORMING MATERIALS
    DALLY, JW
    CHEN, YM
    JAHANMIR, S
    [J]. WEAR, 1990, 141 (01) : 95 - 114
  • [7] Hailey J L, 1996, Proc Inst Mech Eng H, V210, P3, DOI 10.1243/PIME_PROC_1996_210_385_02
  • [8] *HIBB KARLSS SOR I, 1994, ABAQUS STAND US MAN
  • [9] A PARAMETRIC ANALYSIS OF THE CONTACT STRESS IN ULTRA-HIGH-MOLECULAR-WEIGHT POLYETHYLENE ACETABULAR CUPS
    JIN, ZM
    DOWSON, D
    FISHER, J
    [J]. MEDICAL ENGINEERING & PHYSICS, 1994, 16 (05) : 398 - 405
  • [10] Polyethylene wear in Scanhip(R) arthroplasty with a 22 or 32 mm head - 62 matched patients followed for 7-9 years
    Kesteris, U
    Ilchmann, T
    Wingstrand, H
    Onnerfalt, R
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1996, 67 (02): : 125 - 127