Diagnostic imaging capabilities of the Ocelot -Optical Coherence Tomography System, ex-vivo evaluation and clinical relevance

被引:4
|
作者
Dohad, Suhail [1 ,2 ]
Shao, John [3 ]
Cawich, Ian [4 ]
Kankaria, Manish [5 ]
Desai, Arjun [5 ]
机构
[1] Cardiovasc Med Grp Southern Calif, Beverly Hills, CA 90210 USA
[2] Cedars Sinai Med Ctr, Cedars Sinai Heart Inst, Los Angeles, CA 90048 USA
[3] Columbia Univ Hosp, New York, NY USA
[4] Arkansas Heart Hosp, Little Rock, AK USA
[5] Avinger Inc, Redwood City, CA USA
来源
BMC MEDICAL IMAGING | 2015年 / 15卷
关键词
INTRAVASCULAR ULTRASOUND; ATHEROSCLEROTIC PLAQUE; CORONARY ANGIOSCOPY; VALIDATION; MORPHOLOGY;
D O I
10.1186/s12880-015-0098-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Optical coherence tomography (OCT) is a high-resolution sub-surface imaging modality using near-infrared light to provide accurate and high contrast intra-vascular images. This enables accurate assessment of diseased arteries before and after intravascular intervention. This study was designed to corroborate diagnostic imaging equivalence between the Ocelot and the Dragonfly OCT systems with regards to the intravascular features that are most important in clinical management of patients with atherosclerotic vascular disease. These intravascular features were then corroborated in vivo during treatment of peripheral arterial disease (PAD) pathology using the Ocelot catheter. Methods: In order to compare the diagnostic information obtained by Ocelot (Avinger Inc., Redwood City, CA) and Dragonfly (St. Jude Medical, Minneapolis, MN) OCT systems, we utilized ex-vivo preparations of arterial segments. Ocelot and Dragonfly catheters were inserted into identical cadaveric femoral peripheral arteries for image acquisition and interpretation. Three independent physician interpreters assessed the images to establish accuracy and sensitivity of the diagnostic information. Histologic evaluation of the corresponding arterial segments provided the gold standard for image interpretation. In vivo clinical images were obtained during therapeutic interventions that included crossing of peripheral chronic total occlusions (CTOs) using the Ocelot catheter. Results: Strong concordance was demonstrated when matching image characteristics between both OCT systems and histology. The Dragonfly and Ocelot system's vessel features were interpreted with high sensitivity (91.1-100 %) and specificity (86.7-100 %). Inter-observer concordance was documented with excellent correlation across all vessel features. The clinical benefit that the Ocelot OCT system provided was demonstrated by comparable procedural images acquired at the point of therapy. Conclusions: The study demonstrates equivalence of image acquisition and consistent physician interpretation of images acquired by the Ocelot and the Dragonfly OCT systems in-spite of distinct image processing algorithms and catheter configurations. This represents a dramatic shift away from both fluoroscopic imaging and diagnostic-only OCT imaging during peripheral arterial intervention towards therapeutic devices that incorporate real time diagnostic OCT imaging. In the clinical practice, these diagnostic capabilities have translated to best-in-class safety and efficacy for CTO crossing using the Ocelot catheter.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Diagnostic imaging capabilities of the Ocelot -Optical Coherence Tomography System, ex-vivo evaluation and clinical relevance
    Suhail Dohad
    John Shao
    Ian Cawich
    Manish Kankaria
    Arjun Desai
    BMC Medical Imaging, 15
  • [2] Optical coherence tomography and Raman spectroscopy of the ex-vivo retina
    Evans, Julia W.
    Zawadzki, Robert J.
    Liu, Rui
    Chan, James W.
    Lane, Stephen M.
    Werner, John S.
    JOURNAL OF BIOPHOTONICS, 2009, 2 (6-7) : 398 - 406
  • [3] Optical coherence tomography attenuation imaging for lipid core detection: an ex-vivo validation study
    Gnanadesigan, Muthukaruppan
    Hussain, Ali S.
    White, Stephen
    Scoltock, Simon
    Baumbach, Andreas
    van der Steen, Antonius F. W.
    Regar, Evelyn
    Johnson, Thomas W.
    van Soest, Gijs
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2017, 33 (01): : 5 - 11
  • [4] Use of formalin-fixed tissues for ex-vivo imaging with optical coherence tomography (OCT)
    Shortkroff, Sonya
    Goodwin, Alicia
    Giattina, Susanne
    Liu, Bin
    Brezinski, Mark E.
    IMAGING, MANIPULATION, AND ANALYSIS OF BIOMOLECULES, CELLS, AND TISSUES IV, 2006, 6088
  • [5] Optical coherence tomography attenuation imaging for lipid core detection: an ex-vivo validation study
    Muthukaruppan Gnanadesigan
    Ali S. Hussain
    Stephen White
    Simon Scoltock
    Andreas Baumbach
    Antonius F. W. van der Steen
    Evelyn Regar
    Thomas W. Johnson
    Gijs van Soest
    The International Journal of Cardiovascular Imaging, 2017, 33 : 5 - 11
  • [6] Ex-vivo evaluation of miniaturized probes for endoscopic optical coherence tomography in urothelial cancer diagnostics
    Schoeb, Dominik Stefan
    Wollensak, Carolin
    Kretschmer, Simon
    Gonzalez-Cerdas, Gerardo
    Ataman, Caglar
    Kayser, Gian
    Dressler, Franz Friedrich
    Gratzke, Christian
    Zappe, Hans
    Miernik, Arkadiusz
    ANNALS OF MEDICINE AND SURGERY, 2022, 77
  • [7] Endometrium imaging using real-time rotational optical coherence tomography imaging system A pilot, prospective and ex-vivo study
    Law, Tracy Sze Man
    Wu, Fangrong
    Xu, Hui
    Wang, Chi Chiu
    Li, Tin Chiu
    MEDICINE, 2019, 98 (44) : e17738
  • [8] Feasibility and methodology of optical coherence tomography imaging of human intracranial aneurysms - ex-vivo pilot study
    Vuong, Barry
    Sun, Cuiru
    Khiel, Tim-Rasmus
    Gardecki, Joseph A.
    Standish, Beau A.
    da Costa, Leodante
    de Morais, Josaphat
    Tearney, Guillermo J.
    Yang, Victor X. D.
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS VIII, PTS 1 AND 2, 2012, 8207
  • [9] OPTICAL COHERENCE TOMOGRAPHY IMAGING OF EX VIVO HUMAN KIDNEY
    Luttmann, Amber
    Hariri, Lida
    Rice, Photini Faith
    Winkler, Amy
    Bracamontec, Erika
    Sokoloff, Mitchell
    Nguyen, Mike
    Barton, Jennifer
    LASERS IN SURGERY AND MEDICINE, 2010, : 86 - 87
  • [10] Transscleral optical coherence tomography - An experimental study in ex-vivo human eyes
    Hoerauf, H
    Winkler, J
    Scholz, C
    Wirbelauer, C
    Gordes, RS
    Koch, P
    Engelhardt, R
    Laqua, H
    Birngruber, R
    LASERS IN SURGERY AND MEDICINE, 2002, 30 (03) : 209 - 215