Single-beat real-time three-dimensional echocardiographic automated contour detection for quantification of left ventricular volumes and systolic function

被引:14
作者
Ren, Ben [1 ]
Vletter, Wim B. [1 ]
McGhie, Jackie [1 ]
Soliman, Osama I. I. [1 ]
Geleijnse, Marcel L. [1 ]
机构
[1] Erasmus Univ, Med Ctr, Dept Cardiol, Thoraxctr, NL-3015 CE Rotterdam, Netherlands
关键词
Left ventricle; Three-dimensional; Echocardiography; CARDIAC MAGNETIC-RESONANCE; BORDER-DETECTION; CONTRAST ECHOCARDIOGRAPHY; EJECTION FRACTION; REPRODUCIBILITY; ACCURACY; QUANTITATION; ALGORITHM; FEASIBILITY; VALIDATION;
D O I
10.1007/s10554-013-0327-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To assess the feasibility and accuracy in measuring left ventricular (LV) end-diastolic volume (EDV), end-systolic volume (ESV) and ejection fraction (EF) with Siemens single-beat real-time 3D transthoracic echocardiography. The LV volumes and EF were measured in 3D datasets acquired by six imaging modes (time-1-harmonic (T1H), time-1-fundamental, time-2-harmonic, time-2-fundamental, space-1-harmonic (S1H), and space-1-fundamental) in 41 patients using the automated contouring algorithm and compared with manually corrected 3DE QLAB measurements. The main determinates of the temporal and spatial resolutions of 3D datasets acquired were the fundamental and harmonic modes. Consequently, the S1H mode had the lowest volume rate and highest spatial resolution. Compared with the 3DE QLAB analysis, the S1H mode resulted in the best LV volumes and EF estimates in all patients (0 +/- A 10 % for EF, -7 +/- A 44 ml for EDV, -7 +/- A 39 ml for ESV) and in the 10 patients with correct LV contour tracking according to a visual assessment from the multiplanar reconstruction views in all six modes (0 +/- A 9 % for EF, -3 +/- A 23 ml for EDV, -2 +/- A 14 ml for ESV). The T1H mode was the best alternative. Overall 28 patients (68 %) could be analysed automatically and satisfyingly with the S1H and T1H modes: 0 +/- A 8 % (EF), 0 +/- A 27 ml (EDV) and -1 +/- A 16 ml (ESV). The accuracy of the Siemens automated RT-3D algorithm in measuring LV volumes and EF is significantly influenced by the different imaging modes. The S1H mode may be the preferred 3D acquisition mode, supplemented by the T1H mode in enlarged LVs that do not fit in the S1H acquisition sector.
引用
收藏
页码:287 / 294
页数:8
相关论文
共 15 条
[1]   Feasibility of Single-Beat Full-Volume Capture Real-Time Three-Dimensional Echocardiography and Auto-Contouring Algorithm for Quantification of Left Ventricular Volume: Validation with Cardiac Magnetic Resonance Imaging [J].
Chang, Sung-A ;
Lee, Sang-Chol ;
Kim, Eun-Young ;
Hahm, Seung-Hee ;
Jang, Shin Yi ;
Park, Sung-Ji ;
Choi, Jin-Oh ;
Park, Seung Woo ;
Choe, Yeon Hyeon ;
Oh, Jae K. .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2011, 24 (08) :853-859
[2]   Performance of 3-Dimensional Echocardiography in Measuring Left Ventricular Volumes and Ejection Fraction [J].
Dorosz, Jennifer L. ;
Lezotte, Dennis C. ;
Weitzenkamp, David A. ;
Allen, Larry A. ;
Salcedo, Ernesto E. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 59 (20) :1799-1808
[3]   Reproducibility and accuracy of echocardiographic measurements of left ventricular parameters using real-time three-dimensional echocardiography [J].
Jenkins, C ;
Bricknell, K ;
Hanekom, L ;
Marwick, TH .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 44 (04) :878-886
[4]   Comparison of contrast agent-enhanced versus non-contrast agent-enhanced real-time three-dimensional echocardiography for analysis of left ventricular Systolic function [J].
Krenning, Boudewijn J. ;
Kirschbaum, Sharon W. ;
Soliman, Osama I. I. ;
Nemes, Attila ;
van Geuns, Robert-Jan ;
Vletter, Wim B. ;
Veltman, Caroline E. ;
ten Cate, Folkert J. ;
Roelandt, Jos R. T. C. ;
Geleijnse, Marcel L. .
AMERICAN JOURNAL OF CARDIOLOGY, 2007, 100 (09) :1485-1489
[5]   High-resolution transthoracic real-time three-dimensional echocardiography -: Quantitation of cardiac volumes and function using semi-automatic border detection and comparison with cardiac magnetic resonance imaging [J].
Kuhl, HP ;
Schreckenberg, M ;
Rulands, D ;
Katoh, M ;
Schäfer, W ;
Schummers, G ;
Bücker, A ;
Hanrath, P ;
Franke, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (11) :2083-2090
[6]   Accurate and reproducible measurement of left ventricular volume and ejection fraction by contrast echocardiography - A comparison with magnetic resonance imaging [J].
Malm, S ;
Frigstad, S ;
Sagberg, E ;
Larsson, H ;
Skjaerpe, T .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 44 (05) :1030-1035
[7]   Real-Time 3-Dimensional Echocardiographic Quantification of Left Ventricular Volumes [J].
Mor-Avi, Victor ;
Jenkins, Carly ;
Kuehl, Harald P. ;
Nesser, Hans-Joachim ;
Marwick, Thomas ;
Franke, Andreas ;
Ebner, Christian ;
Freed, Benjamin H. ;
Steringer-Mascherbauer, Regina ;
Pollard, Heidi ;
Weinert, Lynn ;
Niel, Johannes ;
Sugeng, Lissa ;
Lang, Roberto M. .
JACC-CARDIOVASCULAR IMAGING, 2008, 1 (04) :413-423
[8]   Validation of a novel automated border-detection algorithm for rapid and accurate quantitation of left ventricular volumes based on three-dimensional echocardiography [J].
Muraru, Denisa ;
Badano, Luigi P. ;
Piccoli, Gianluca ;
Gianfagna, Pasquale ;
Del Mestre, Lorenzo ;
Ermacora, Davide ;
Proclemer, Alessandro .
EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY, 2010, 11 (04) :359-368
[9]   Accuracy and reproducibility of quantitation of left ventricular function by real-time three-dimensional echocardiography versus cardiac magnetic resonance [J].
Soliman, Osama I. I. ;
Kirschbaum, Sharon W. ;
van Dalen, Bas M. ;
van der Zwaan, Heleen B. ;
Delavary, Babak Mahdavian ;
Vletter, Wim B. ;
van Geuns, Robert-Jan M. ;
Ten Cate, Folkert J. ;
Geleijnse, Marcel L. .
AMERICAN JOURNAL OF CARDIOLOGY, 2008, 102 (06) :778-783
[10]   A comparison between QLAB and TomTec full volume reconstruction for real time three-dimensional echocardiographic quantification of left ventricular volumes [J].
Soliman, Osama I. I. ;
Krenning, Boudewijn J. ;
Geleiinse, Marcel L. ;
Nemes, Attila ;
van Geuns, Robert-Jan ;
Baks, Timo ;
Anwar, Ashraf M. ;
Galema, Tjebbe W. ;
Vletter, Wim B. ;
Ten Cate, Folkert J. .