Pentacam and Orbscan II Measurements of Posterior Corneal Elevation Before and After Photorefractive Keratectomy

被引:39
作者
Ha, Byoung Jin [3 ]
Kim, Sun Woong [2 ]
Kim, Sang Woo [1 ]
Kim, Eung Kweon [1 ]
Kim, Tae-im [1 ]
机构
[1] Yonsei Univ, Coll Med, Dept Ophthalmol, Inst Vis Res, Seoul, South Korea
[2] Soonchunhyang Univ, Coll Med, Dept Ophthalmol, Puchon, South Korea
[3] Siloam Eye Hosp, Seoul, South Korea
关键词
IN-SITU KERATOMILEUSIS; THICKNESS MEASUREMENTS; POST-LASIK; LASER; ULTRASOUND; SURFACE; PACHYMETRY; ACCURACY; ECTASIA; MYOPIA;
D O I
10.3928/1081597X-20090301-09
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To compare the measurements of posterior corneal, elevation above the best-fit sphere before and after photorefractive, keratectomy (PRK) using two different technologies-Scheimpflug imaging using Pentacam (Oculus) and scanning slit combined with Placido imaging using Orbscan II (Bausch & Lomb). METHODS: In a retrospective observational case series study, 60 eyes measured using Orbscan II and Pentacam before or after PRK were evaluated. The data were analyzed as three groups: preoperative, early postoperative,(from 1 to 3 months after surgery), and late postoperative (>3 months postoperatively but not exceeding I year). Anterior chamber depth, posterior best-fit float sphere, posterior central elevation (PCE) above the best-fit sphere, and posterior maximum elevation (PME) above the best-fit sphere in the 6.0-mm zone were analyzed. RESULTS: Posterior central elevation and PME values measured using Orbscan II were greater than those measured using Pentacam (P<.001). Anterior chamber depth values measured using Orbscan II were smaller than those measured using Pentacam in all three groups (P<.001). Using Orbscan II, PCE and PME values for the two postoperative groups were greater than for the preoperative group. Using Pentacam, the PCE values for the three groups were similar, and the PME values for the preoperative and late postoperative groups were similar. CONCLUSIONS: These findings indicate that Orbscan II measurements of posterior corneal elevation before and after excimer laser refractive corneal surgery may not be valid. Pentacam measurement showed no significant change in posterior corneal elevation from before to after excimer laser corneal surgery. [J Refract Surg. 2009; 25:290-295.]
引用
收藏
页码:290 / 295
页数:6
相关论文
共 19 条
[1]  
[Anonymous], CORNEAL TOPOGRAPHY W
[2]   Factors affecting the forward shift of posterior corneal surface after laser in situ keratomileusis [J].
Baek, TM ;
Lee, KH ;
Kagaya, F ;
Tomidokoro, A ;
Amano, S ;
Oshika, T .
OPHTHALMOLOGY, 2001, 108 (02) :317-320
[3]   Assessing the accuracy of Orbscan II post-LASIK: apparent keratectasia is paradoxically associated with anterior chamber depth reduction in successful procedures [J].
Cairns, G ;
Ormonde, SE ;
Gray, T ;
Hadden, OB ;
Morris, T ;
Ring, P ;
McGhee, CNJ .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2005, 33 (02) :147-152
[4]   Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes [J].
Chakrabarti, HS ;
Craig, JP ;
Brahma, A ;
Malik, TY ;
McGhee, CNJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (11) :1823-1828
[5]   Changes in the posterior cornea after laser in situ keratomileusis and photorefractive keratectomy [J].
Ciolino, Joseph B. ;
Belin, Michael W. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (09) :1426-1431
[6]   Comparison of corneal pachymetry using ultrasound and Orbscan II [J].
Fakhry, MA ;
Artola, A ;
Belda, JL ;
Ayala, MJ ;
Alió, JL .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) :248-252
[7]   Corneal changes after laser refractive surgery for myopia: Comparison of Orbscan II and Pentacam findings [J].
Hashemi, Hassan ;
Mehravaran, Shiva .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (05) :841-847
[8]   Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures [J].
Iskander, NG ;
Penno, EA ;
Peters, NT ;
Gimbel, HV ;
Ferensowicz, M .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (05) :681-685
[9]  
Kim SW, 2007, J REFRACT SURG, V23, P888
[10]  
Krachmer JH, 2005, CORNEA, V2