Visible colour displacement sensing system for manganese(II)

被引:20
作者
Dai, Zhaohua [1 ,2 ]
Khosla, Nikhil [2 ]
Canary, James W. [1 ]
机构
[1] NYU, Dept Chem, New York, NY USA
[2] Pace Univ, Dept Chem & Phys Sci, New York, NY 10038 USA
基金
美国国家卫生研究院;
关键词
manganese(II); colorimetric; sensor; displacement assay; ratiometric; INDICATOR; ASSAYS; NEUROTOXICITY; ZN; MN;
D O I
10.1080/10610270802438853
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A visible colour displacement system has been developed for the colorimetric and ratiometric detection of Mn(II) ions. The supramolecular displacement system is composed of 2-(5-bromo-2-pyridylazo)-5-[N-propyl-N-(3-sulfopropyl)amino]phenol (5-Br-PAPS), which binds Zn2+ with a change in colour and binds Mn2+ without changing colour, ethylene glycol-bis(2-aminoethylether)-N, N, N', N'-tetraacetic acid (EGTA) and Zn2+ at neutral pH. The sensing system exhibits the colour of 5-Br-PAPS itself since EGTA binds Zn2+ more strongly than 5-Br-PAPS does. However, upon presentation of Mn2+, Zn2+ is displaced from EGTA to bind 5-Br-PAPS to produce a pronounced colour change from yellow to purple. The absorbance decreases at 449nm and increases at 552nm, both linearly to Mn2+ concentration at low micromolar levels. This should be very advantageous since no suitable colorimetric reagent is available for the direct declination of Mn2+ at physiological pH.
引用
收藏
页码:296 / 300
页数:5
相关论文
共 26 条
[1]   Supramolecular analytical chemistry [J].
Anslyn, Eric V. .
JOURNAL OF ORGANIC CHEMISTRY, 2007, 72 (03) :687-699
[2]   AN ULTRASENSITIVE COLORIMETRIC ASSAY FOR MANGANESE [J].
BEYER, WF ;
FRIDOVICH, I .
ANALYTICAL BIOCHEMISTRY, 1988, 170 (02) :512-519
[3]   Neurotoxicity of inhaled manganese: Public health danger in the shower? [J].
Elsner, RJF ;
Spangler, JG .
MEDICAL HYPOTHESES, 2005, 65 (03) :607-616
[4]   USE OF COMPLEX DISPLACEMENT REACTIONS IN PHOTOMETRIC ANALYSIS [J].
INCZEDY, J .
TALANTA, 1970, 17 (12) :1212-&
[5]   Searching for a relationship between manganese and welding and Parkinson's disease [J].
Jankovic, J .
NEUROLOGY, 2005, 64 (12) :2021-2028
[6]   Transient evoked otoacoustic emissions pattern as a prognostic indicator for hearing preservation in acoustic neuroma surgery [J].
Kim, AH ;
Edwards, BM ;
Telian, SA ;
Kileny, PR ;
Arts, HA .
OTOLOGY & NEUROTOLOGY, 2006, 27 (03) :372-379
[7]   Essential tremor - Occupational exposures to manganese and organic solvents [J].
Louis, ED ;
Applegate, LM ;
Factor-Litvak, P ;
Parides, MK ;
Andrews, L .
NEUROLOGY, 2004, 63 (11) :2162-2164
[8]   KINETICS AND EQUILIBRIA OF THE ZN-EDTA-PAR POSTCOLUMN REACTION DETECTION SYSTEM FOR THE DETERMINATION OF ALKALINE-EARTH METALS [J].
LUCY, CA ;
DINH, HN .
ANALYTICAL CHEMISTRY, 1994, 66 (06) :793-797
[9]   Manganese neurotoxicity: Connecting the dots along the continuum of dysfunction [J].
Martin, CJ .
NEUROTOXICOLOGY, 2006, 27 (03) :347-349
[10]   Colorimetric and fluorimetric assays to quantitate micromolar concentrations of transition metals [J].
McCall, KA ;
Fierke, CA .
ANALYTICAL BIOCHEMISTRY, 2000, 284 (02) :307-315