High resolution, high fill factor a-Si:H sensor arrays for optical imaging

被引:3
|
作者
Rahn, JT [1 ]
Lemmi, F [1 ]
Mei, P [1 ]
Lu, JP [1 ]
Boyce, JB [1 ]
Street, RA [1 ]
Apte, RB [1 ]
Ready, SE [1 ]
Van Schuylenbergh, KF [1 ]
Nylen, P [1 ]
Ho, J [1 ]
Fulks, RT [1 ]
Lau, R [1 ]
Weisfield, RL [1 ]
机构
[1] Xerox PARC, Palo Alto, CA 94304 USA
来源
AMORPHOUS AND HETEROGENEOUS SILICON THIN FILMS: FUNDAMENTALS TO DEVICES-1999 | 1999年 / 557卷
关键词
D O I
10.1557/PROC-557-809
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Amorphous silicon large area sensor arrays are in production for x-ray medical imaging. The most common pixel design works very well for many applications but is limited in spatial resolution because the available sensor area (the fill factor) vanishes in small pixels. One solution is a 3-dimensional structure in which the sensor is placed above the active matrix addressing. However, such high fill factor designs have previously introduce cross talk between pixels. We present data for a design in which the a-Si:H p-i-n photodiode sensor layer has a continuous i-layer and top p(+)-layer, and a patterned n(+)-layer contact to the pixel. Arrays of 64 mu m and 75 mu m pitch have been fabricated and are the highest resolution a-Si:H arrays reported to date. The resolution matches the pixel size, and sensitivity has been improved by the high fill factor. Comparison is made between arrays with standard TFTs and TFTs with self-aligned source and drain contacts. Data line capacitance is improved by use of the self-aligned contacts. Measurements are included on the contact to bias capacitance. The high fill factor design greatly suppresses lateral leakage currents, while retaining ease of processing. Provided illumination levels remain below saturation, the resolution matches expectation for the pixel size.
引用
收藏
页码:809 / 814
页数:6
相关论文
共 50 条
  • [1] High resolution, high fill factor a-Si:H sensor arrays for optical imaging
    Rahn, J.T.
    Lemmi, F.
    Mei, P.
    Lu, J.P.
    Boyce, J.B.
    Street, R.A.
    Apte, R.B.
    Ready, S.E.
    Van Schuylenbergh, K.F.
    Nylen, P.
    Ho, J.
    Fulks, R.T.
    Lau, R.
    Weisfield, R.L.
    Materials Research Society Symposium - Proceedings, 1999, 557 : 809 - 814
  • [2] High resolution, high fill factor a-Si:H sensor arrays for medical imaging
    Rahn, JT
    Lemmi, F
    Weisfield, RL
    Lujan, R
    Mei, P
    Lu, JP
    Ho, J
    Ready, SE
    Apte, RB
    Nylen, P
    Boyce, J
    Street, RA
    MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 : 510 - 517
  • [3] High Fill Factor a-Si:H Sensor Arrays with Reduced Pixel Crosstalk
    Vygranenko, Y.
    Sazonov, A.
    Striakhilev, D.
    Chang, J. H.
    Heiler, G.
    Lai, J.
    Tredwell, T.
    Nathan, A.
    AMORPHOUS AND POLYCRYSTALLINE THIN-FILM SILICON SCIENCE AND TECHNOLOGY-2008, 2008, 1066 : 463 - +
  • [4] High dynamic range active pixel sensor arrays for digital x-ray imaging using a-Si:H
    Lai, J
    Nathan, A
    Rowlands, J
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A, 2006, 24 (03): : 850 - 853
  • [5] HIGH-RESOLUTION OPTICAL-RECORDING ON A-SI FILMS
    GOTCHIYAEV, VZ
    KOROLKOV, VP
    SOKOLOV, AP
    CHERNUKHIN, VP
    JOURNAL OF NON-CRYSTALLINE SOLIDS, 1991, 137 : 1297 - 1300
  • [6] RADIATION IMAGING WITH 2D A-SI SENSOR ARRAYS
    FUJIEDA, I
    NELSON, S
    STREET, RA
    WEISFIELD, RL
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (04) : 1056 - 1062
  • [7] A High-resolution Optical-fiber Imaging Sensor
    Ran, Zhao
    You, Zhou
    Dong, Wei
    Yun, Wang
    Yang, Fu
    Wang Jiuwang
    OPTICAL FIBER SENSORS AND COMMUNICATION (AOPC 2019), 2019, 11340
  • [8] High performance A-Si:H(:Cl) TFT
    Kim, SK
    Lee, KS
    Choi, JH
    Kim, CS
    Jang, J
    PROCEEDINGS OF THE THIRD SYMPOSIUM ON THIN FILM TRANSISTOR TECHNOLOGIES, 1997, 96 (23): : 138 - 145
  • [9] Fabrication of Microlens Arrays with High Quality and High Fill Factor by Inkjet Printing
    Zhang, Qiaoshuang
    Schambach, Maximilian
    Schlisske, Stefan
    Jin, Qihao
    Mertens, Adrian
    Rainer, Christian
    Hernandez-Sosa, Gerardo
    Heizmann, Michael
    Lemmer, Uli
    ADVANCED OPTICAL MATERIALS, 2022, 10 (14)
  • [10] Spectrally resolved fill factor measurements on a-Si:H based solar cell structures
    Wagner, M
    Peros, D
    Guse, S
    Bohm, M
    THIN FILMS FOR PHOTOVOLTAIC AND RELATED DEVICE APPLICATIONS, 1996, 426 : 71 - 76