Fifteen-year experience with Toyobo paracorporeal left ventricular assist system

被引:34
作者
Saito, Shunsuke [1 ]
Matsumiya, Goro [1 ]
Sakaguchi, Taichi [1 ]
Fujita, Tomoyuki [1 ]
Kuratani, Toru [1 ]
Ichikawa, Hajime [1 ]
Sawa, Yoshiki [1 ]
机构
[1] Osaka Univ, Div Cardiovasc Surg, Dept Surg, Grad Sch Med, Suita, Osaka 5650871, Japan
关键词
Toyobo left ventricular assist system; Paracorporeal; Bridge to transplantation; Bridge to recovery; HEART-FAILURE; CLINICAL-EXPERIENCE; DEVICE; TRANSPLANTATION; PATIENT; SUPPORT;
D O I
10.1007/s10047-009-0448-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Our 15-year experience of the Toyobo paracorporeal left ventricular assist system (LVAS) at Osaka University Hospital was reviewed. In total, 61 patients underwent Toyobo LVAS implantation from January 1992 to August 2007. Their mean age was 38.1 +/- 16.9 years. The etiologies of heart failure were idiopathic dilated cardiomyopathy in 35 patients, ischemic cardiomyopathy in 15, myocarditis in 5, secondary cardiomyopathy in 4, and others in 2. Preoperatively, intubation was required in 41. patients (67.3%), an intra-aortic balloon pump was required in 38 (62.3%), and extracorporeal membrane oxygenation was required in 30 (49.2%). Four patients underwent heart transplantation and 11 underwent LVAS removal. Of those I I patients, 4 were subjected to emergent removal because of device complications and all of them died. Of the 7 patients that underwent scheduled LVAS removal, heart failure recurred in 2 patients and reimplantation was required. In terms of major device-related complications, cerebral hemorrhage occurred in 16 patients (26.2%), cerebral infarction in 19 (31.1%), mediastinitis in 10 (16.4%), and inflow/outflow cannula exit site infection in 19 patients (31.1.%). The actuarial survival rate of the patients operated on in the last 5 years of this study was 66.3% at 6 months and 45.9% at 1 year. A] though the survival rate of patients supported by the Toyobo LVAS has recently improved, the morbidity rate is significant. Considering the current severe shortage of heart donors in Japan, it is important to introduce more durable devices with fewer complications and to establish the strategies for using the LVAS as a bridge to recovery.
引用
收藏
页码:27 / 34
页数:8
相关论文
共 23 条
[1]  
Dang NC, 2006, J HEART LUNG TRANSPL, V25, P1, DOI 10.1016/j.healun.2005.07.008
[2]   Mechanical Circulatory Support device database of the International Society for Heart and Lung Transplantation: Second Annual Report-2004 [J].
Deng, MC ;
Edwards, LB ;
Hertz, MI ;
Rowe, AW ;
Keck, BM ;
Kormos, R ;
Naftel, DC ;
Kirklin, JK .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2004, 23 (09) :1027-1034
[3]   MULTICENTER CLINICAL-EVALUATION OF THE HEARTMATE 1000-IP LEFT-VENTRICULAR ASSIST DEVICE [J].
FRAZIER, OH ;
ROSE, EA ;
MACMANUS, Q ;
BURTON, NA ;
LEFRAK, EA ;
POIRIER, VL ;
DASSE, KA .
ANNALS OF THORACIC SURGERY, 1992, 53 (06) :1080-1090
[4]   Medical progress - Implantable left ventricular assist devices [J].
Goldstein, DJ ;
Oz, MC ;
Rose, EA .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (21) :1522-1533
[5]   Grafted skeletal myoblast sheets attenuate myocardial remodeling in pacing-induced canine heart failure model [J].
Hata, Hiroki ;
Matsumiya, Goro ;
Miyagawa, Shigeru ;
Kondoh, Haruhiko ;
Kawaguchi, Naomasa ;
Matsuura, Nariaki ;
Shimizu, Tatsuya ;
Okano, Teruo ;
Matsuda, Hikaru ;
Sawa, Yoshiki .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2006, 132 (04) :918-924
[6]   Evidence of improved right ventricular structure after LVAD support in patients with end-stage cardiomyopathy [J].
Küçüker, SA ;
Stetson, SJ ;
Becker, KA ;
Akgül, A ;
Loebe, M ;
Lafuente, JA ;
Noon, GP ;
Koerner, MM ;
Entman, ML ;
Torre-Amione, G .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2004, 23 (01) :28-35
[7]  
MASAI T, 1995, ASAIO J, V41, pM522, DOI 10.1097/00002480-199507000-00065
[8]  
MATSUDA T, 1984, T AM SOC ART INT ORG, V30, P353
[9]   OVERSEAS TRANSPORT OF A PATIENT WITH AN EXTRACORPOREAL LEFT-VENTRICULAR ASSIST DEVICE [J].
MATSUWAKA, R ;
MATSUDA, H ;
KANEKO, M ;
MIYAMOTO, Y ;
MASAI, T ;
NAKATANI, T ;
SASAKI, E ;
FRAZIER, OH .
ANNALS OF THORACIC SURGERY, 1995, 59 (02) :522-523
[10]   Clinical experience with 111 Thoratec ventricular assist devices [J].
McBride, LR ;
Naunheim, KS ;
Fiore, AC ;
Moroney, DA ;
Swartz, MT .
ANNALS OF THORACIC SURGERY, 1999, 67 (05) :1233-1238