Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer

被引:55
作者
Ogbuehi, Kelechi C. [1 ]
机构
[1] Coll Appl Med Sci, Dept Optometry & Vis Sci, Riyadh 11433, Saudi Arabia
关键词
Goldmann tonometry; limits of agreement; non-contact tonometry; repeatability; Topcon CT80;
D O I
10.1111/j.1444-0938.2006.00068.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: The reliability of non-contact tonometers has been reported extensively in the literature. This study was designed to assess reliability of the new Topcon CT80 non-contact tonometer in normotensive subjects, using the Goldmann tonometer as the standard. Methods: The accuracy of the Topcon CT 80 non-contact torrometer was assessed by comparing its IOP assessments with those of the Goldmann applanation tonometer, on 60 right eyes of young healthy subjects with normal intraocular pressures. Each subject's intraocular pressure was assessed with each technique on two separate occasions, one week apart. The reliability of each technique was determined by the assessment of its inter-session repeatability using the Bland-Altman method. The 95 per cent limits of agreement for the two methods were also determined. Results: No statistically significant difference was found between the average intraocular pressures measured with the two techniques (p > 0.05). The inter-session repeatability indices for the two techniques did not differ significantly (p > 0.05). The mean difference in intraocular measurements between the two techniques was 0.2 +/- 1.5 mmHg (mean +/- SD) and the 95 per cent limits of agreement were -3.14 and +2.74 mmHg, with the non-contact tonometer returning higher readings than the Goldmann tonometer. Conclusion: In this sample of normotensive subjects, the Topcon CT80 non-contact torrometer proved to be accurate and as reliable as the Goldmann tonometer in the assessment of intraocular pressure. Thus, it can be used as an objective clinical method for the assessment of normal intraocular pressure.
引用
收藏
页码:310 / 314
页数:5
相关论文
共 31 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] EVALUATION OF KEELER PULSAIR NONCONTACT TONOMETRY - RELIABILITY AND REPRODUCIBILITY
    BONOMI, L
    BARAVELLI, S
    COBBE, C
    TOMAZZOLI, L
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1991, 229 (03) : 210 - 212
  • [3] Brencher H L, 1991, J Am Optom Assoc, V62, P395
  • [4] COCKRELL DJ, 1991, COLLECTION OF TECHNICAL PAPERS - 11TH AIAA AERODYNAMIC DECELERATOR SYSTEMS TECHNOLOGY CONFERENCE, P23
  • [5] NONCONTACT APPLANATION TONOMETER - DESCRIPTION AND CLINICAL EVALUATION
    FORBES, M
    PICO, G
    GROLMAN, B
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1974, 91 (02) : 134 - 140
  • [6] GROLMAN B, 1972, AMER J OPT ARCH AM A, V49, P646
  • [7] Grolman B, 1990, J Am Optom Assoc, V61, P857
  • [8] HANSEN MK, 1995, ACTA OPHTHALMOL SCAN, V73, P176
  • [9] Clinical performance of non-contact tonometry by Reichert AT550® in glaucomatous patients
    Jorge, J
    González-Méijome, JM
    Díaz-Rey, JA
    Almeida, JB
    Ribeiro, P
    Parafita, MA
    [J]. OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2003, 23 (06) : 503 - 506
  • [10] Clinical performance of the Reichert AT550:: a new non-contact tonometer
    Jorge, J
    Díaz-Rey, JA
    González-Méijome, JM
    Almeida, JB
    Parafita, MA
    [J]. OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2002, 22 (06) : 560 - 564